Page 61
Functional check 49 Document no. M1031517-02 NOTE: * Ensure that the calibration gas and regulator are functioning properly before calibration. Perform annual maintenance on the regulator as required. For more information see section “Adjustments and calibrations” in Compact Airway modules slot. For details on recommended accessories see “Supplies and Accessories” catalog. 5.1.1 Hemodynamic patient simulators The following tables present the patient simulators’...
Page 62
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 50 Document no. M1031517-02 Table 12 Adapter cables for hemodynamic patient simulators 5.2 Visual inspection •Make sure that the monitor is switched to standby. Disconnect the mains power cord from the monitor. If the monitor is connected to the Datex-Ohmeda Network, disconnect the Mon-Net cable from the monitor. Remove any memory cards, if installed. 1....
Page 63
Functional check 51 Document no. M1031517-02 5.3 Functional inspection 5.3.1 General 1. Connect the mains power cord. Check that the LED above the main switch is lit continuously (full charged batteries) or periodically (charging batteries). 2. Switch the monitor on. Check that the monitor starts up properly, i.e. a normal start-up sound is heard from the loudspeaker, the alarm LEDs turn on and off, and the monitoring screen appears. No error messages should...
Page 64
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 52 Document no. M1031517-02 5.3.2 Display 1. Check that the picture on the screen is correct. 5.3.3 Keyboard(s) 1. Tests with all the connected keyboards: −Press the Monitor Setup key. Turn the ComWheel in both directions and check that the cursor in the menu moves correspondingly. Select Normal Screen and check that the menu disappears from the screen....
Page 65
Functional check 53 Document no. M1031517-02 5.3.6 Compact Airway Module, E-CXXXXX/ M-CXXXXX Wait until the message ‘Calibrating gas sensor’ disappears from the screen. 1. Check that the fan is running. 2. If the module contains membrane keys on the front panel, press each of the keys for at least one second and check that they are identified. Compact Airway Modules with the Patient Spirometry option 3. Connect a clean spirometry tube and D-lite to the module....
Page 66
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 54 Document no. M1031517-02 5.3.9 Multiparameter Hemodynamic Modules ECG and RESP measurements 1. Check the function of the front panel membrane key ECG, Lead, if available. 2. Connect an ECG cable to the module. Connect the cable leads to a patient simulator. Check that all ECG and impedance respiration information is shown on the monitor screen as...
Page 67
Functional check 55 Document no. M1031517-02 2. Check the InvBP channel with a patient simulator. Zero the InvBP channel. Then check that the values and waveforms correspond to the simulator settings. Temperature measurement 3. Check the temperature channels with a patient simulator. Check that the temperature measurement information is shown on the monitor screen as configured on the simulator. 5.3.11 Dual pressure Module, E-PP/ M-PP 1. Check the function...
Page 68
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 56 Document no. M1031517-02 5.3.14 Nellcor Compatible Saturation module, E-NSAT/ M-NSAT 1. Connect a Nellcor SpO2 finger probe to the module. Check that the message ‘Pulse search ’ is shown. Check also that the shown message changes to ‘Check probe ’ within 30 seconds. 2. Attach the SpO2 probe on your finger. Check that a reading of 95-100 and a proper SpO2...
Page 69
Functional check 57 Document no. M1031517-02 5.3.19 Recorder 1. Press the Record Wave module key and check that the module starts recording the selected waveforms. Press the Stop module key to stop recording. 2. Check that the quality of the recordings is acceptable. 5.3.20 Network connection 1. Check that the Mon-Net cable connector and the Identification plug are clean and intact, then connect them to the frame unit. Check that the monitor connects to the...
Page 70
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 58 Document no. M1031517-02 5.3.24 General • Switch the monitor to standby •Perform final cleaning •Fill in all necessary documents