Page 61

Functional check
49
                                          Document no. M1031517-02
NOTE: * Ensure that the calibration gas and regulator are functioning properly before calibration. 
Perform annual maintenance on the regulator as required. For more information see section 
“Adjustments and calibrations” in Compact Airway modules slot.
For details on recommended accessories see “Supplies and Accessories” catalog.
5.1.1 Hemodynamic patient simulators
The following tables present the patient simulators’...

Page 62

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
50
Document no. M1031517-02                                                                           
Table 12 Adapter cables for hemodynamic patient simulators
5.2 Visual inspection
•Make sure that the monitor is switched to standby.
Disconnect the mains power cord from the monitor. 
If the monitor is connected to the Datex-Ohmeda Network, disconnect the Mon-Net cable from 
the monitor. Remove any memory cards, if installed.
1....

Page 63

Functional check
51
                                          Document no. M1031517-02
5.3 Functional inspection
5.3.1 General
1. Connect the mains power cord.
Check that the LED above the main switch is lit continuously (full charged batteries) or 
periodically (charging batteries).
2. Switch the monitor on.
Check that the monitor starts up properly, i.e. a normal start-up sound is heard from the 
loudspeaker, the alarm LEDs turn on and off, and the monitoring screen appears.
No error messages should...

Page 64

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
52
Document no. M1031517-02                                                                           
5.3.2 Display
1. Check that the picture on the screen is correct. 

5.3.3 Keyboard(s)
1. Tests with all the connected keyboards:
−Press the 
Monitor Setup key. Turn the ComWheel in both directions and check that the 
cursor in the menu moves correspondingly. Select Normal Screen and check that the 
menu disappears from the screen....

Page 65

Functional check
53
                                          Document no. M1031517-02
5.3.6 Compact Airway Module, E-CXXXXX/ M-CXXXXX
Wait until the message ‘Calibrating gas sensor’ disappears from the screen.
1. Check that the fan is running.
2. If the module contains membrane keys on the front panel, press each of the keys for at least 
one second and check that they are identified.

Compact Airway Modules with the Patient Spirometry option
3. Connect a clean spirometry tube and D-lite to the module....

Page 66

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
54
Document no. M1031517-02                                                                           
5.3.9 Multiparameter Hemodynamic Modules
ECG and RESP measurements
1. Check the function of the front panel membrane key ECG, Lead, if available.
2. Connect an ECG cable to the module. Connect the cable leads to a patient simulator. Check 
that all ECG and impedance respiration information is shown on the monitor screen as...

Page 67

Functional check
55
                                          Document no. M1031517-02
2. Check the InvBP channel with a patient simulator. 
Zero the InvBP channel. Then check that the values and waveforms correspond to the simulator 
settings.

Temperature measurement
3. Check the temperature channels with a patient simulator.
Check that the temperature measurement information is shown on the monitor screen as 
configured on the simulator.

5.3.11 Dual pressure Module, E-PP/ M-PP
1. Check the function...

Page 68

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
56
Document no. M1031517-02                                                                           
5.3.14 Nellcor Compatible Saturation module, E-NSAT/ M-NSAT
1. Connect a Nellcor SpO2 finger probe to the module. Check that the message ‘Pulse search ’ is 
shown. Check also that the shown message changes to ‘Check probe ’ within 30 seconds.
2. Attach the SpO2 probe on your finger. Check that a reading of 95-100 and a proper SpO2...

Page 69

Functional check
57
                                          Document no. M1031517-02
5.3.19 Recorder 
1. Press the Record Wave module key and check that the module starts recording the selected 
waveforms. Press the 
Stop module key to stop recording.
2. Check that the quality of the recordings is acceptable.

5.3.20 Network connection
1. Check that the Mon-Net cable connector and the Identification plug are clean and intact, then 
connect them to the frame unit.
Check that the monitor connects to the...

Page 70

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
58
Document no. M1031517-02                                                                           
5.3.24 General
•
Switch the monitor to standby
•Perform final cleaning
•Fill in all necessary documents
 