Page 71

General troubleshooting  
59
                                          Document no. M1031517-02
6 General troubleshooting     
Figure 17 S/5 Compact Monitor general troubleshooting flowchart
Monitor not
functioning
Power cord
connected?
Connect power
cord
Monitor can be
started now?Batteries were
empty
On/Standby LED
is ON or blinks?
PMB failure.
Replace PMB.
Monitor can be
started now?
Display Unit
failure.
No
Yes
Yes
Yes
No
Possible AC/DC
unit failure.
Replace AC/DC
unit.
No
No
Frame Unit fan is...

Page 72

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
60
Document no. M1031517-02                                                                            

Page 73

Appendix A, Functional check form, Datex-Ohmeda S/5 CAM, CCCM
A-1
Document no. M1031517-02
APPENDIX A Functional check form, 
Datex-Ohmeda S/5 CAM, CCCM
OK = Test OK            N.A. = Test not applicable           Fail = Test failed Customer
Service
Service engineer Date
Monitor Installation
Frame: F-CM________ System options: E/M-
Software: N- E/M-
K- N- E/M-
K- N- E/M-
Visual Inspection         OK          N.A.         Fail
1. Check all units visually
                           
Functional Inspection...

Page 74

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
A-2
Document no. M1031517-02
Notes 
 
5.3.8. Tonometry Module, E-TONO/ M-TONO S/N
                           
5.3.9. Multiparameter Hemodynamic Modules S/N
                           
. ECG and RESP measurements
                           
. Temperature measurement
                           
. Invasive blood pressure measurement
                           
. SpO2 measurement
                           
. Non invasive blood pressure...

Page 75

Appendix A, Functional check form, Datex-Ohmeda S/5 CAM, CCCM
A-3
Document no. M1031517-02
5.3.15. Datex-Ohmeda Oxygen Saturation module, 
M-OSATS/N
                           
5.3.16. BIS Module, E-BIS/ M-BIS S/N
                           
5.3.17. Entropy Module, E-ENTROPY/ M-ENTROPY S/N
                           
5.3.18. Data Card and Menu Card function S/N
                           
5.3.19. Recorder  S/N
                           
5.3.20. Network connection S/N
                           
5.3.21....

Page 76

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
A-4
Document no. M1031517-02
For your notes: 

Page 77

Appendix B, ElectroMagnetic Compatibility
B-1
Document no. M1031517-02
APPENDIX B ElectroMagnetic Compatibility
Table 1 Guidance and manufacturer’s declaration – electromagnetic emissions
 
(1S/5™ CAM/ CCCM equipped with N-CMW option 
Guidance and manufacturer’s declaration – electromagnetic emissions
The S/5™ CAM/ CCCM is intended for use in the electromagnetic environment specified below. The customer or 
the user of the S/5™ CAM/ CCCM should assure that it is used in such an environment.
Emissions...

Page 78

 
B-2
Document no.  M1031517-02
Table 2 Guidance and manufacturer’s declaration – electromagnetic immunity
Guidance and manufacturer’s declaration – electromagnetic immunity
The S/5™ CAM/ CCCM is intended for use in the electromagnetic environment specified below. The customer or the 
user of the S/5™ CAM/ CCCM should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic discharge 
(ESD)
IEC 61000-4-2±6 kV...

Page 79

Appendix B, ElectroMagnetic Compatibility
B-3
Document no. M1031517-02
Table 3 Guidance and manufacturer’s declaration – electromagnetic immunity
Guidance and manufacturer’s declaration – electromagnetic immunity
The S/5™ CAM/ CCCM is intended for use in the electromagnetic environment specified below. The customer or the 
user of the S/5™ CAM/ CCCM should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance 
levelElectromagnetic environment - guidance
Portable and...

Page 80

 
B-4
Document no.  M1031517-02
(1       For impedance RESP measurement. For invasive pressure measurement in M-series modules 
except M-PRE(S)TN. The invasive pressure measurement compliance level is dependent on the 
transducer used, and only the lower level is guaranteed for all transducers. 
(2       For BIS measurement in E-BIS/M-BIS, RESP measurement in M-(P)RE(S)TN and invasive 
pressure measurement in M-COPSv/E-COPSv and M-COP/E-COP modules. The invasive 
pressure measurement compliance level is...
Start reading GE S5 Service Manual

Related Manuals for GE S5 Service Manual

All GE manuals