GE S5 Service Manual
Here you can view all the pages of manual GE S5 Service Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.
Page 71
General troubleshooting 59 Document no. M1031517-02 6 General troubleshooting Figure 17 S/5 Compact Monitor general troubleshooting flowchart Monitor not functioning Power cord connected? Connect power cord Monitor can be started now?Batteries were empty On/Standby LED is ON or blinks? PMB failure. Replace PMB. Monitor can be started now? Display Unit failure. No Yes Yes Yes No Possible AC/DC unit failure. Replace AC/DC unit. No No Frame Unit fan is...
Page 72
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 60 Document no. M1031517-02
Page 73
Appendix A, Functional check form, Datex-Ohmeda S/5 CAM, CCCM A-1 Document no. M1031517-02 APPENDIX A Functional check form, Datex-Ohmeda S/5 CAM, CCCM OK = Test OK N.A. = Test not applicable Fail = Test failed Customer Service Service engineer Date Monitor Installation Frame: F-CM________ System options: E/M- Software: N- E/M- K- N- E/M- K- N- E/M- Visual Inspection OK N.A. Fail 1. Check all units visually Functional Inspection...
Page 74
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors A-2 Document no. M1031517-02 Notes 5.3.8. Tonometry Module, E-TONO/ M-TONO S/N 5.3.9. Multiparameter Hemodynamic Modules S/N . ECG and RESP measurements . Temperature measurement . Invasive blood pressure measurement . SpO2 measurement . Non invasive blood pressure...
Page 75
Appendix A, Functional check form, Datex-Ohmeda S/5 CAM, CCCM A-3 Document no. M1031517-02 5.3.15. Datex-Ohmeda Oxygen Saturation module, M-OSATS/N 5.3.16. BIS Module, E-BIS/ M-BIS S/N 5.3.17. Entropy Module, E-ENTROPY/ M-ENTROPY S/N 5.3.18. Data Card and Menu Card function S/N 5.3.19. Recorder S/N 5.3.20. Network connection S/N 5.3.21....
Page 76
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors A-4 Document no. M1031517-02 For your notes:
Page 77
Appendix B, ElectroMagnetic Compatibility B-1 Document no. M1031517-02 APPENDIX B ElectroMagnetic Compatibility Table 1 Guidance and manufacturer’s declaration – electromagnetic emissions (1S/5™ CAM/ CCCM equipped with N-CMW option Guidance and manufacturer’s declaration – electromagnetic emissions The S/5™ CAM/ CCCM is intended for use in the electromagnetic environment specified below. The customer or the user of the S/5™ CAM/ CCCM should assure that it is used in such an environment. Emissions...
Page 78
B-2 Document no. M1031517-02 Table 2 Guidance and manufacturer’s declaration – electromagnetic immunity Guidance and manufacturer’s declaration – electromagnetic immunity The S/5™ CAM/ CCCM is intended for use in the electromagnetic environment specified below. The customer or the user of the S/5™ CAM/ CCCM should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance Electrostatic discharge (ESD) IEC 61000-4-2±6 kV...
Page 79
Appendix B, ElectroMagnetic Compatibility B-3 Document no. M1031517-02 Table 3 Guidance and manufacturer’s declaration – electromagnetic immunity Guidance and manufacturer’s declaration – electromagnetic immunity The S/5™ CAM/ CCCM is intended for use in the electromagnetic environment specified below. The customer or the user of the S/5™ CAM/ CCCM should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance levelElectromagnetic environment - guidance Portable and...
Page 80
B-4 Document no. M1031517-02 (1 For impedance RESP measurement. For invasive pressure measurement in M-series modules except M-PRE(S)TN. The invasive pressure measurement compliance level is dependent on the transducer used, and only the lower level is guaranteed for all transducers. (2 For BIS measurement in E-BIS/M-BIS, RESP measurement in M-(P)RE(S)TN and invasive pressure measurement in M-COPSv/E-COPSv and M-COP/E-COP modules. The invasive pressure measurement compliance level is...