Service Menu
1
                                          Document no. M1031519-01
Introduction
The monitor has a Service Menu, which is a useful tool to examine monitor functions and to 
troubleshoot in case a fault occurs.
Service Menu structure
     Service Menu
Frame
Subnet  St at us
St at us
Power SupplyBatt ery
Set / Tes t
Key board
Key board Ty pe
Param et ersGas  Unit
EC G Module
ESTP Module
PP
C alibrat ions
Saf ety  valve
Serv ic e Log
Record Data
Maint enanc e
Planned Ma in t .
Repair
U pgrade...

Page 122

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
2
Document no. M1031519-01                                                                           
Service Menu
NOTE: The Service Menu pictures are for reference only. Details on the menu page can vary depending 
on the software version and the module type in use.
1. Press the 
Monitor Setup key.
2. Select Install/Service (password 
16-4-34).
3. Select Service (password 26-23-8). 

Page 123

Service Menu
3
                                          Document no. M1031519-01
1Frame
1.1 Memory
The frame menu includes service menus common 
for the frame.
A service menu to check the status of the memory 
used in the CPU board of the monitor.
Test Memory tests the condition of the 
EEPROM/Flash memory component of the CPU 
board. If the result of the test is Fail, see section 
“Error messages” in the Display unit slot.
Test SRAM tests the Static RAM memory of the CPU 
board in a similar way as the...

Page 124

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
4
Document no. M1031519-01                                                                           
1.2 Network
1.2.1 Network Status
The Network Status view shows the general status 
of the network
Location ID: Monitor’s location given at the 
setup. 
DRI level: Shows the selected level of network 
communication. The network communication is 
set according to the network software used (e.g. 
S-CNET01)
Interfaces: The field indicates...

Page 125

Service Menu
5
                                          Document no. M1031519-01
Subnet Status
The Subnet status view gives more accurate 
information of the different subnet id:s connected. All 
four Subnet status menus have a similar structure. 
The number of different packets transmitted and 
received by the monitor are shown in the columns 
below Tx and Rx. The packet types are:
Waveforms: Waveform data
Phys. data: Physiological numerical data
Alarms: Alarms, alarm profiles and alarm 
limits
Link...

Page 126

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
6
Document no. M1031519-01                                                                           
1.2.2 Network Config 
The DRI Level is for setting the monitors network 
communication. The network communication is set 
according to the network software used (e.g. 
S-CNET01).
Network software S-CNET01 ->DRI Level = 2001
Network software S-CNET02 ->DRI Level = 2001 or 
2002(WLAN)
Network software L-NET03 ->DRI Level = 2003
Virtual...

Page 127

Service Menu
7
                                          Document no. M1031519-01
Table 1 Ethernet service data
Value Usage Notes
Received packets (Statistics 
In/Packets)Total number of received packets since last 
cold start.
Transmitted packets (Statistics 
Out/Packets)Total number of transmitted packets since 
last cold start.
Received bytes (Statistics In/Bytes) Total number of received bytes since last 
cold start.
Transmitted bytes (Statistics 
In/Bytes)Total number of transmitted bytes since...

Page 128

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
8
Document no. M1031519-01                                                                           
1.2.4 WLAN
Driver: WLAN driver chip name.
Connection indicates the state of the WLAN 
connection.
Ethernet Addr: Ethernet address of the WLAN card 
used. 
AP MAC: Ethernet address of the access point where 
the monitor is currently connected.
Signal Quality indicates the quality of the radio signal 
between the monitor and the access...

Page 129

Service Menu
9
                                          Document no. M1031519-01
AP List
Access Points: shows the access points (max. eight) 
which are visible for the monitor. 
CH is the WLAN channel configured to this particular 
access point.
NL indicates the noise level of the channel used  
(0…100%) 
SL indicates the signal level of the channel used
(0…100%)
WLAN Config
Network ID is for setting the correct WLAN network 
ID name. 
WEP Algorithm is for setting the level of the 
encryption (width of...

Page 130

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
10
Document no. M1031519-01                                                                           
1.3 MemCards
1.3.1 Status
Module Present indicates whether the MemCard option 
is installed into the monitor. Possible values are YES and 
NO.
Module Active indicates whether the MemCard services 
are available. Possible values are YES and NO.
ROM indicates the status of the ROM memory of the 
module. Possible values are OK and ERR....