GE S5 Service Manual
Here you can view all the pages of manual GE S5 Service Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.
Page 121
Service Menu 1 Document no. M1031519-01 Introduction The monitor has a Service Menu, which is a useful tool to examine monitor functions and to troubleshoot in case a fault occurs. Service Menu structure Service Menu Frame Subnet St at us St at us Power SupplyBatt ery Set / Tes t Key board Key board Ty pe Param et ersGas Unit EC G Module ESTP Module PP C alibrat ions Saf ety valve Serv ic e Log Record Data Maint enanc e Planned Ma in t . Repair U pgrade...
Page 122
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 2 Document no. M1031519-01 Service Menu NOTE: The Service Menu pictures are for reference only. Details on the menu page can vary depending on the software version and the module type in use. 1. Press the Monitor Setup key. 2. Select Install/Service (password 16-4-34). 3. Select Service (password 26-23-8).
Page 123
Service Menu 3 Document no. M1031519-01 1Frame 1.1 Memory The frame menu includes service menus common for the frame. A service menu to check the status of the memory used in the CPU board of the monitor. Test Memory tests the condition of the EEPROM/Flash memory component of the CPU board. If the result of the test is Fail, see section “Error messages” in the Display unit slot. Test SRAM tests the Static RAM memory of the CPU board in a similar way as the...
Page 124
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 4 Document no. M1031519-01 1.2 Network 1.2.1 Network Status The Network Status view shows the general status of the network Location ID: Monitor’s location given at the setup. DRI level: Shows the selected level of network communication. The network communication is set according to the network software used (e.g. S-CNET01) Interfaces: The field indicates...
Page 125
Service Menu 5 Document no. M1031519-01 Subnet Status The Subnet status view gives more accurate information of the different subnet id:s connected. All four Subnet status menus have a similar structure. The number of different packets transmitted and received by the monitor are shown in the columns below Tx and Rx. The packet types are: Waveforms: Waveform data Phys. data: Physiological numerical data Alarms: Alarms, alarm profiles and alarm limits Link...
Page 126
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 6 Document no. M1031519-01 1.2.2 Network Config The DRI Level is for setting the monitors network communication. The network communication is set according to the network software used (e.g. S-CNET01). Network software S-CNET01 ->DRI Level = 2001 Network software S-CNET02 ->DRI Level = 2001 or 2002(WLAN) Network software L-NET03 ->DRI Level = 2003 Virtual...
Page 127
Service Menu 7 Document no. M1031519-01 Table 1 Ethernet service data Value Usage Notes Received packets (Statistics In/Packets)Total number of received packets since last cold start. Transmitted packets (Statistics Out/Packets)Total number of transmitted packets since last cold start. Received bytes (Statistics In/Bytes) Total number of received bytes since last cold start. Transmitted bytes (Statistics In/Bytes)Total number of transmitted bytes since...
Page 128
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 8 Document no. M1031519-01 1.2.4 WLAN Driver: WLAN driver chip name. Connection indicates the state of the WLAN connection. Ethernet Addr: Ethernet address of the WLAN card used. AP MAC: Ethernet address of the access point where the monitor is currently connected. Signal Quality indicates the quality of the radio signal between the monitor and the access...
Page 129
Service Menu 9 Document no. M1031519-01 AP List Access Points: shows the access points (max. eight) which are visible for the monitor. CH is the WLAN channel configured to this particular access point. NL indicates the noise level of the channel used (0…100%) SL indicates the signal level of the channel used (0…100%) WLAN Config Network ID is for setting the correct WLAN network ID name. WEP Algorithm is for setting the level of the encryption (width of...
Page 130
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 10 Document no. M1031519-01 1.3 MemCards 1.3.1 Status Module Present indicates whether the MemCard option is installed into the monitor. Possible values are YES and NO. Module Active indicates whether the MemCard services are available. Possible values are YES and NO. ROM indicates the status of the ROM memory of the module. Possible values are OK and ERR....