Service Menu 31 Document no. M1031519-01 4.6 COP SvO2 Meas. state: Measurement status shows: No optical module (No OM) connected, initializing the optical module (Init OM), normal measurement state (Normal) and failed module (OM fail). Value is a measured SvO 2 value. Gain is the gain of the remote red and infrared signals (0, 1, 2 or 3) Red int: Reflected red intensity Ired int: Reflected infrared intensity Loc red: Local red intensity Loc ired: Local infrared...
Page 152
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 32 Document no. M1031519-01 Button: The front panel Zero P4 and Start C.O. key functions can be confirmed by pressing the key and checking that the relevant OFF message turns to ON. Timeouts is a cumulative number that indicates how many times the module has not responded to the monitors inquiry. Bad checksums is a cumulative number that indicates how many...
Page 153
Service Menu 33 Document no. M1031519-01 4.7 NIBP Module Service menu for Non-invasive Blood Pressure Modules. Service Data Pressure shows the measured pressure multiplied by 10. This value is automatically zero-drift compensated. Zero shows the difference between the zeroing value in the permanent memory (stored when the module is calibrated) and the current automatic zero-drift compensation multiplied by 10. The value can change between +20 and -20 mmHg....
Page 154
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 34 Document no. M1031519-01 4.7.1 NIBP Demo A service menu for demonstrating the oscillometric method of NIBP measurement. The menu shows the real-time pressure signals that are measured from the NIBP cuff. The measurement result is shown in the adjoining digit field. Wave Recording is for selecting the recording option. If ON is selected, the pressure...
Page 155
Service Menu 35 Document no. M1031519-01 4.7.2 NIBP Calibration How to Calibrate Calibration selection is available only when protection is OFF. NOTE: Both channels B1 and B2 must be calibrated simultaneously. NOTE: The module must be in the frame during the whole procedure. NIBP calibration can be performed in the NIBP Service menu as follows: 1. If Protection is ON, change it to OFF by pressing the NIBP module buttons Auto ON/OFF and Start Cancel...
Page 156
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 36 Document no. M1031519-01 message field. After this, a pressure bar will appear beside the menu and the text ‘Calibrating’ will appear in NIBP digit field. 3. Connect an external mercury manometer with a pump to the module through the both tubes of the hose. Pump up to about 200 mmHg pressure (range of 150 to 300 mmHg allowed) according to the manometer....
Page 157
Service Menu 37 Document no. M1031519-01 4.7.4 NIBP Pulse Valve How to check Pulse Valve Wrap an adult cuff around a pipe and connect the cuff to the module. Select Start test and push the ComWheel. The pressure rises beyond 240 mmHg and stops. The pulse valve opens. The module counts the time it takes for the pressure to go down from 240 mmHg to 50 mmHg and displays it on the screen. The test can be manually stopped by selecting Stop test. The valve can be...
Page 158
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 38 Document no. M1031519-01 4.7.5 NIBP Buttons/Leds The selections Auto ON/OFF, Manual ON/OFF, STAT ON/OFF, and Measur. ON/OFF have effect only on the NIBP module, M-NIBP. Buttons/Leds Data For information on general items Pressure, Zero, Protect handle, Calibr. prot., +15 V power, AD0 to AD7 as well as Timeouts etc., see section “4.7 NIBP Module”. Buttons...
Page 159
Service Menu 39 Document no. M1031519-01 4.7.6 NIBP Pneumatics How to check Interval 20 mmHg -> 185 mmHg Select Start pump at different combinations of the valves open/closed and push the ComWheel. The module counts the time it takes for the pressure to go up from 20 mmHg to 185 mmHg and displays it. When all the valves are closed, the pump should be able to pump the pressure in about 1 to 4 seconds into an adult cuff wrapped around a pipe. The pump does not...
Page 160
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 40 Document no. M1031519-01 4.7.7 NIBP Watchdog How to test Adult watchdog time Select Test ADULT and press the ComWheel. Watchdog interval starts counting up seconds and keeps on counting as long as the +15 Vdirty is on. The time should be 120 to 140 seconds. How to test Infant watchdog time Select Test INFANT and press the ComWheel. Watchdog interval starts...