Home > GE > Medical Equipment > GE S5 Service Manual

GE S5 Service Manual

Here you can view all the pages of manual GE S5 Service Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.

Page 191

Service Menu
71
                                          Document no. M1031519-01
4.15.1 Sensor
Sensor Service data is changed to the sensor view.
Entropy refers to the respective module key. Is 
highlighted when the Entropy key is pressed on the 
module. (Opens the Entropy menu.)
Check Sensor is highlighted when the Check Sensor key 
is pressed on the module. Activates the sensor check.
Cable indicates whether the Entropy sensor cable is 
connected to the module or not.
Sensor indicates whether the...

Page 192

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
72
Document no. M1031519-01                                                                           
4.16 DIS Interfacing
4.16.1 Interfacing
How to interface
To select the parameter data source:
•Press the Monitor Setup and select Interfacing.
•Select the desired measurement parameter, for example Gases.
•Select the desired source by name, for example Aest. 
NOTE: The name of the device is visible on the list only if the device is...

Page 193

Service Menu
73
                                          Document no. M1031519-01
4.16.2 Status Page
4.16.3 DIS Interfacing
Access via Monitor setup - Interfacing.
The selection Next page is available, if more than 8 DIS 
modules are connected to the DIS bus simultaneously.
Interfaces
The menu displays a list of all connected DIS modules 
and the statuses of the corresponding external devices.
If the bus voltage is too low, you cannot add more 
devices. ‘DIS module bus voltage low. Do not add more...

Page 194

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
74
Document no. M1031519-01                                                                           
DIS Module specific page
Service Data
Product name: DIS module name.
Product type:  DIS module type.
Driver sw id:  DIS module software and its release date.
Short product name:  DIS module name.
Module serial number:  DIS module serial number.
HW card type:  PCB type.
HW id:  DIS module PCB identification number.
HW test date:  DIS...

Page 195

Service Menu
75
                                          Document no. M1031519-01
5 Set/Test
In the power supply unit, there is a safety element, which 
needs refreshing at least every 1.5 seconds. If the main 
CPUs software will not refresh, safety element gives 
reset to main CPU. In normal operation the main CPUs 
software refreshes safety element every 0.2 seconds.
The purpose of the safety element is to restart the 
monitor if there is a serious malfunction. This feature is 
useful in two cases:...

Page 196

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
76
Document no. M1031519-01                                                                           
5.1 Country Settings
Language
Select the software language.
National Reqs
Select the software features, which include national 
requirements.
Power Frequency
Set the power frequency (50 Hz/60 Hz). This setting is 
used to filter out possible power frequency interference 
from the parameter measurements.
Time format
Set the time...

Page 197

Service Menu
77
                                          Document no. M1031519-01
6 Service Log
Error, event, alarm and maintenance data is stored in Service Log.
The service log contains information about the occurred monitor errors, events and alarms since the last factory reset 
or service log reset. The service log is saved in the EEPROM memory of the main CPU board. The user can also store 
different maintenance events in the maintenance log.
Maintenance menu is for setting and viewing the...

Page 198

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
78
Document no. M1031519-01                                                                           
6.1 Maintenance
6.1.1 Planned Maintenance
The Maintenance History Log gives the user a possibility to store the maintenance history of the monitor. The user can 
store different planned maintenance (PM) events, repairs and upgrades in the maintenance history log.
Running hours: shows how many hours the monitor has 
been on. User...

Page 199

Service Menu
79
                                          Document no. M1031519-01
6.1.2 Repair
6.1.3 Upgrade
The repair menu gives the user a possibility to store re-
pair events in the maintenance history log. The different
repairs that can be saved are: Display, Power Supply,
Recorder, Frame (e.g. board in the frame), Parame-
ters (e.g. a board in the parameter module) or Other (a
miscellaneous repair not specified by previous op-
tions).
The upgrade menu gives the user a possibility to store
the...

Page 200

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
80
Document no. M1031519-01                                                                           
6.2 Error History
6.3 Event History
Last Errors: The section shows the last monitor errors 
and the time of their occurence.
Error counters: The section contains counters for each 
different (detected) monitor error. The time of 
occurrence of the last error is shown beside each 
counter.
Last log reset: The date and time of the last...