Page 231
Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01) 19 Document no. M1031520-02 2.10.2 Serial data interface The serial interface is an asynchronous RS-232 connection, which uses hardware flow control. The communication speed is 115.2 kbps. The communication uses the following parameters: Parity None, Data bits 8 and Stop bits 1. 2.10.3 Digital and analog outputs Digital outputs There are separate digital outputs. Both signals use TTL level. The outputs...
Page 232
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 20 Document no. M1031520-02 3 Service procedures 3.1 General service information The field service of the S/5 Compact Anesthesia Monitor and S/5 Compact Critical Care Monitor is limited to replacing faulty PC boards and mechanical parts. Return the faulty PC boards to GE Healthcare for repair. GE Healthcare is always available for service advice. Please...
Page 233
Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01) 21 Document no. M1031520-02 3.2.2 Recommended parts 3.2.3 Before beginning • Make sure that the monitor is switched to standby. •Disconnect the power cord and press the service reset switch until the LED next to it turns off. •If the monitor is connected to the Datex-Ohmeda Network, disconnect the Mon-Net cable from the Frame Unit. Take out any memory cards, if installed. •Detach the Display Unit from...
Page 234
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 22 Document no. M1031520-02 •screws are tightened properly •the connector pins are intact 5. Front Unit Check that: •the Connector Board screws are tightened properly and the cable connectors are clean and intact •the Inverter Board screws are tightened properly and the cable connectors are clean and intact •the Backlight cables are intact and attached...
Page 235
Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01) 23 Document no. M1031520-02 3.2.5 Functional inspection 9. Stand-by LED Connect the power cord. Check that the standby LED starts blinking or is on continuously, if the batteries do not need charging. 10. Start-up Switch the monitor on. Check that the monitor starts up properly, i.e. the alarm LEDs turn blank, the start-up sound is heard from the loudspeaker and the normal monitoring screen appears. No...
Page 236
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 24 Document no. M1031520-02 The FACTORY RESET should be performed after the battery replacement, see “Performing Factory Reset” . 14. Loudspeaker Check the loudspeaker by setting the alarm sound: Alarms Setup Alarm Volume Test the whole volume scale from 1 to 10 by turning the ComWheel and check that the alarm volume changes correspondingly. The alarm sound...
Page 237
Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01) 25 Document no. M1031520-02 19. Watchdog circuitry Test the watchdog circuitry: Monitor Setup Install Service (Password 16-4-34) Service (Password 26-23-8) Set/Test Perform the tests WATCHDOG and WD BY OVERLOAD. Check that the monitor restarts in each case. NOTE: When selecting WD BY OVERLOAD, restarting should take place approximately after 15 seconds. Preset monitor settings: Record/Print Record...
Page 238
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 26 Document no. M1031520-02 Check also the functioning of the side panel keys. 22. ComWheel check Turn the ComWheel clockwise and counter clockwise and check that each step generates a sound from the loudspeaker and the corresponding values at the bottom of the menu increase. Select Dummy Press. Press the ComWheel and check that the press generates a sound...
Page 239
Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01) 27 Document no. M1031520-02 26. Battery voltage Check that the shown Battery Voltage value is within 10 – 17V. 27. Monitor temperature Check that the shown Monitor Temperature value is within 20 - 40. 28. Battery operation Disconnect the power cord (without switching the monitor to standby). Check that the monitoring continues normally. The battery indicator should appear in the upper right hand corner...
Page 240
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 28 Document no. M1031520-02 Record/Print Record Waveforms Paper Speed --> 1 mm/s Press the key RECORD WAVE and wait until the recording stops. Check that the length of the scale is now 3.0 cm (± 0.5 cm). 33. Quality of recording Check that the quality of the recording is acceptable. MemCard option NOTE: The following steps are applicable only when the...