Page 271
Frame Unit for F-CM(C)1 (rev. 04), F-CM(C)REC1 (Rev. 01) 5 Document no. M1044470-01 2.3 PMB (Power Management Board) Figure 2 Location of the Power Management Board (side frame panel removed) and the PMB block diagram The Power Management Board (PMB) is responsible for distributing DC current for the rest of the frame. The power source can be the AC/DC unit or one of the internal batteries, and the chosen power line is referred to as ‘VBUS’. The voltage of...
Page 272
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 6 Document no. M1044470-01 PMB CPU unit The CPU unit controls all other functions of the board. The functions include: •Battery current monitoring •Battery capacity calculations •Battery temperature monitoring •Battery charger control •Frame fan voltage control •Level-change of recorder serial line signals •Module power supply measurements (voltages & input...
Page 273
Frame Unit for F-CM(C)1 (rev. 04), F-CM(C)REC1 (Rev. 01) 7 Document no. M1044470-01 2.5 AC/DC Power Supply Unit The universal AC/DC power supply provides an isolated 15.5 V supply for the S/5 Compact Monitor. The power unit is attached to the Frame Unit by four screws and a power-D25 connector. The cooling air for the AC/DC unit is from the Frame Unit fan through cooling holes in the Frame Unit back plate. The power unit generates a load signal, which is read...
Page 274
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 8 Document no. M1044470-01 2.9 Connectors and signals Module motherboard Table 1 Module connector pinout (X1-X4): 25 pin male D-connector Pin No. Signal 1 Reset_485 2-15V 315Vdirty 4+15V 5 Data_485_ 6 Data_485 7GND 8 Reset_485_ 9CTSB 10 RTSB 11 RXDB 12 TXDB 13 GND 14 NC 15 GND 16 CTSC 17 RTSC 18 RXDC 19 TXDC 20 NC 21 PWM_ECG 22 RXD_RS232 23 TXD_RS232 24 +5V 25...
Page 275
Frame Unit for F-CM(C)1 (rev. 04), F-CM(C)REC1 (Rev. 01) 9 Document no. M1044470-01 2.9.1 External Connectors Table 2 Network Connector, X5 Table 3 Coding element connector, X4 Table 4 RS232 Serial data connector X11 For information about S/5 Compact Monitor external connectors see this manual, Part II Slot I “Display Unit”.RJ45 connector Pin No. Signal 1 2 3 4 5 6 7 8Tx + Tx - Rx + N/C N/C Rx - N/C N/C 9 pin female D-connector Pin No. Signal 1 2 3 4 5 6 7 8...
Page 276
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 10 Document no. M1044470-01 2.9.2 Serial Data Interface The Serial Data Interface connector is a 9-pin male D-connector (X3) on the side of the AC/DC power unit. The serial interface is an asyncronous RS-232 connection which uses hardware flow control. The communication speed is 115.2 kbps. The communication uses the following parameters: Parity None, Data...
Page 277
Frame Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01) 11 Document no. M1044470-01 3 Service procedures 3.1 General service information The field service of the S/5 Compact Anesthesia Monitor and S/5 Compact Critical Care Monitor is limited to replacing faulty PC boards and mechanical parts. Return the faulty PC boards to GE Healthcare for repair. GE Healthcare is always available for service advice. Please provide the unit serial number, full type...
Page 278
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 12 Document no. M1044470-01 3.2.3 Before beginning • Make sure that the monitor is switched to standby. •Disconnect the power cord and press the service reset switch until the LED next to it turns off. •If the monitor is connected to the Datex-Ohmeda Network, disconnect the Mon-Net cable. Take out any memory cards, if installed. •Remove the AC/DC Power Unit....
Page 279
Frame Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01) 13 Document no. M1044470-01 •there are no loose objects inside the battery slot Reattach the bottom cover. 5. External parts Check external parts: •the four rubber pads under the monitor are all in place •all stickers are intact •the Display Unit is attached properly •all mechanical parts are functioning properly •the service reset switch is intact •the handle is attached properly •the monitor frame is...
Page 280
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 14 Document no. M1044470-01 Reconnect the fuse holder and check that it locks up properly. 3.2.5 Functional inspection 10. Stand-by LEDs Connect the power cord. Check that the standby LED starts blinking. Check that the fan starts running. 11. Start-up Switch the monitor on. Check that the monitor starts up properly, i.e. the alarm LEDs turn blank, the...