Page 251

Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
39
                                          Document no. M1031520-02
6. Open the four screws holding the metal EMC cover in the Display Unit front cover.
7. Open the two screws covering the video flat cable and ferrite in the LCD Display element. 
8. Lift the EMC cover carefully up.
9. Open the four screws holding the LCD element in the Display Unit front cover.
10. Lift the LCD Display element carefully up.
NOTE: Do not use excessive force when...

Page 252

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
40
Document no. M1031520-02                                                                           
3.3.13 Changing the LCD backlight, F-CM(C)1 rev.03 and rev.04
Perform the steps 1-8 as in section 3.3.12. Changing the LCD backlights, F-CM(C)(REC)1 rev.01, 
rev.02.
1. Detach the screw holding the backlight tube set in place.
2. Detach the backlight tube set by pulling it to the left.
NOTE: Do not bend the backlight tube.
NOTE: Do...

Page 253

Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
41
                                          Document no. M1031520-02
4. Wait for approximately 80 seconds. After the start-up screen appears, enter the Service View 
and make sure that the information regarding the monitor software has been updated. 
Memorize the serial number of the new software.
5. Remove the software card.
6. Replace the original device plate for monitor software by the new one supplied with the software 
card.
7. Perform...

Page 254

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
42
Document no. M1031520-02                                                                           
4 Troubleshooting
4.1 General (Display Unit)
Trouble Cause Treatment
Monitor is not starting 1. On/Stby key pressed too long.
2. The batteries are empty.
3. If power cord connected, AC/DC power unit 
may be faulty.
4. If AC/DC power unit is working, the PMB 
board in the Frame Unit may be faulty.
5. On/Stby key may be faulty.1....

Page 255

Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
43
                                          Document no. M1031520-02
4.2 Software License (service software)
NOTE: You can download the software only onto one monitor. During the downloading of software the 
serial number of the monitor is written on the software card, and if the loading for some reason would 
fail, the software can be downloaded again onto the same monitor, but not onto any other monitor.
4.3 Error messages
Battery temperature...

Page 256

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
44
Document no. M1031520-02                                                                           
4.4 Memory Option
4.5 Troubleshooting for the NET section
Trouble Cause Treatment
‘Two Data Cards in mem. module’ message Two Data cards detected. Remove MemCard from the left hand side slot 
of the module.
‘Two Menu Cards in mem. module’ message Two Menu cards detected. Remove MemCard from the right hand side slot 
of the module....

Page 257

Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
45
                                          Document no. M1031520-02
‘Network board error’ message shows on 
the monitor screenNET section on the CPB The NET section is defective. The board cannot be used. See 
network service page for details.
NET section memory on the CPB The SRAM memory of the NET section is defective or 
uninitialized. The board cannot be used. See network service 
page for details.
Other Site View shows no waveforms  No...

Page 258

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
46
Document no. M1031520-02                                                                           
5 Earlier revisions
This manual fully supports the earlier revisions of the S/5 Compact Anesthesia and Compact Critical 
Care Monitors. 

Page 259

Appendix A, Service check form, S/5 Compact Anesthesia Monitor, S/5 Compact Critical Care Monitor
A-1
Document no. M1031520-02
APPENDIX A Service check form, S/5 Compact Anesthesia Monitor, 
S/5 Compact Critical Care Monitor
OK = Test OK            N.A. = Test not applicable           Fail = Test failed Customer
 
Service Module type S/N
Service engineer
 Date
Monitor Installation
 
F- E- E-
S- E- M-
L- M- N-
      OK          N.A.         Fail     OK          N.A.         Fail
1. DC/DC power supply...

Page 260

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
A-2
Document no. M1031520-02
18. Power supply output voltages
+12 V  11.50...12.50 V
+3.3 V   3.00....3.60 V
Vbus 10.00...17.00
      OK          N.A.         Fail
19. Watchdog circuitry
                           20. Alarm LEDs
                           
21. Membrane keys check
                           22. ComWheel check
                           
23. K-ARK connector
                           24. Module communication...
Start reading GE S5 Service Manual