Page 281
Frame Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01) 15 Document no. M1044470-01 14. Content of service log Select Service Log from the menu. Check the contents of the Service Log for possible problems. If the monitor contains the recorder unit, record the Service Log on the recorder by selecting Record Log. 15. Voltages Check the power supply unit output voltages through the service menu: Monitor Setup Install/Service (Password 16-4-34) Service (password...
Page 282
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 16 Document no. M1044470-01 Change the module into the other slot. When the module is recognized, check that the values do not still increase faster than by 5 per second. 17. Battery voltage Enter the Battery service menu: Monitor Setup Install/Service (password 16-4-34) Service (password 26-23-8) Frame Power Supply Battery Check that the shown Battery...
Page 283
Frame Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01) 17 Document no. M1044470-01 22. Recording Press the key Record Wave and check that the module starts recording the selected waveforms. Press the key Stop on the module to stop recording. NOTE: InvBP waveforms are not recorded without a patient simulator. NOTE: If no recording appears, check that the paper roll is installed in a correct way. Only one side of the paper is printable. 23. Paper speed...
Page 284
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 18 Document no. M1044470-01 Replace the batteries, if the time was less than 45 minutes. 27. Battery charging Reconnect the power cord, switch the monitor on and enter the Battery service menu: Monitor Setup Install/Service (Password 16-4-34) Service (Password 26-23-8) Frame Power Supply Battery Check that the monitor starts charging one of the batteries and...
Page 285
Frame Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01) 19 Document no. M1044470-01 3.3 Disassembly and reassembly WARNINGThe S/5 Compact Monitor is always energized by an internal battery. A short circuit may cause internal damage. Do not touch any exposed wiring or conductive surface inside, as this may cause an electric shock. See the exploded view of the S/5 Compact Monitor Frame Unit in the ‘Spare Parts’ slot to get quick unerstaning of the structure....
Page 286
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 20 Document no. M1044470-01 3.3.5 To detach/remove the Module Mother Board 1. Detach/remove the PMB Board. 2. Unscrew the spacers to remove the Module Mother Board. NOTE: Be carefull not to damage the service reset switch. 3.3.6 To remove the Recorder Unit Disassemble the Side Frame Unit. Loosen the screws and extract the recorder. 3.3.7 Adjustment and...
Page 287
Frame Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01) 21 Document no. M0144470-01 3.4 Internal battery WARNINGThe S/5 Compact Monitor is always energized by an internal battery. A short circuit may cause internal damage. Do not touch any exposed wiring or conductive surface inside, as this may cause an electric shock. The monitor can be run either on mains power or on internal battery. Battery operation is initiated when the power cord is disconnected or...
Page 288
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 22 Document no. M0144470-01 3.4.3 Replacing the batteries 1. Open the Bottom Plate of the Frame Unit by opening the eight screws. 2. Detach the battery cables and remove the batteries. 3. Insert new batteries and connect the battery cables. NOTE: Remember the order of the batteries connected to the PMB board.
Page 289
Frame Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01) 23 Document no. M1044470-01 4 Troubleshooting For troubleshooting, see also section Display Unit: Troubleshooting in\ this manual. 4.1 General Trouble CauseWhat to do Monitor is not starting 1. On/Stby key pressed too long. 2. The batteries are empty. 3. If power cord connected. If the On/Stby LED is lit or blinks. AC/DC power unit may be faulty. 4. If the Frame unit fan is running. If AC/DC power...
Page 290
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 24 Document no. M1044470-01 Keyboard not working, and module communication not working.UPI section of the Central Processing Unit not functi oning normally. Restart the monitor. Replace the Central Processing Board. Keyboard partly not working. Keyboa rd faulty. Replace the keyboard foil. Connector Board faulty Re place the Connector Board. Display...