Page 281

Frame Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
15
                                          Document no. M1044470-01
14. Content of service log
Select Service Log from the menu. Check the contents of the Service Log for possible problems. 
If the monitor contains the recorder unit, record the Service Log on the recorder by selecting 
Record Log.

15. Voltages
Check the power supply unit output voltages through the service menu: 
Monitor Setup
Install/Service (Password 16-4-34)
Service (password...

Page 282

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
16
Document no. M1044470-01                                                                           
Change the module into the other slot. When the module is recognized, check that the values do 
not still increase faster than by 5 per second.

17. Battery voltage
Enter the Battery service menu:
Monitor Setup
Install/Service (password 16-4-34)
Service (password 26-23-8)
Frame
Power Supply
Battery
Check that the shown Battery...

Page 283

Frame Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
17
                                          Document no. M1044470-01
22. Recording
Press the key 
Record Wave and check that the module starts recording the selected 
waveforms. Press the key
 Stop on the module to stop recording.
NOTE: InvBP waveforms are not recorded without a patient simulator.
NOTE: If no recording appears, check that the paper roll is installed in a correct way. Only one side of 
the paper is printable.

23. Paper speed...

Page 284

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
18
Document no. M1044470-01                                                                           
Replace the batteries, if the time was less than 45 minutes.

27. Battery charging
Reconnect the power cord, switch the monitor on and enter the Battery service menu:
Monitor Setup
Install/Service (Password 16-4-34)
Service (Password 26-23-8)
Frame
Power Supply
Battery
Check that the monitor starts charging one of the batteries and...

Page 285

Frame Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
19
                                          Document no. M1044470-01
3.3 Disassembly and reassembly
WARNINGThe S/5 Compact Monitor is always energized by an internal battery. A short 
circuit may cause internal damage. Do not touch any exposed wiring or 
conductive surface inside, as this may cause an electric shock.
See the exploded view of the  S/5 Compact Monitor Frame Unit in the ‘Spare Parts’ slot to get quick 
unerstaning of the structure....

Page 286

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
20
Document no. M1044470-01                                                                           
3.3.5 To detach/remove the Module Mother Board
1. Detach/remove the PMB Board.
2. Unscrew the spacers to remove the Module Mother Board.
NOTE: Be carefull not to damage the service reset switch.
3.3.6 To remove the Recorder Unit
Disassemble the Side Frame Unit.
Loosen the screws and extract the recorder. 
3.3.7 Adjustment and...

Page 287

Frame Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
21
                                          Document no. M0144470-01
3.4 Internal battery
WARNINGThe S/5 Compact Monitor is always energized by an internal battery. A short 
circuit may cause internal damage. Do not touch any exposed wiring or 
conductive surface inside, as this may cause an electric shock.
The monitor can be run either on mains power or on internal battery. Battery operation is initiated 
when the power cord is disconnected or...

Page 288

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
22
Document no. M0144470-01                                                                           
3.4.3 Replacing the batteries
1. Open the Bottom Plate of the Frame Unit by opening the eight screws.
2. Detach the battery cables and remove the batteries.
3. Insert new batteries and connect the battery cables.
NOTE: Remember the order of the batteries connected to the PMB board. 

Page 289

Frame Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
23
                                          Document no. M1044470-01
4 Troubleshooting
For troubleshooting, see also section Display Unit: Troubleshooting in\
  this manual.
4.1 General 
Trouble CauseWhat to do
Monitor is not starting 1. 
On/Stby key pressed too long.
2. The batteries are empty.
3. If power cord connected. If the 
On/Stby LED is lit or blinks. AC/DC 
power unit may be faulty.
4. If the Frame unit fan is running. If 
AC/DC power...

Page 290

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
24
Document no. M1044470-01                                                                           
Keyboard not working, and 
module communication not 
working.UPI section of the Central Processing 
Unit not functi
oning normally. Restart the monitor.
Replace the Central Processing Board.
Keyboard partly not working.  Keyboa rd faulty.  Replace the keyboard foil.
Connector Board faulty  Re place the Connector Board.
Display...