Page 241

Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
29
                                          Document no. M1031520-02
38. Card type recognition
Wait until the information regarding SLOT1 is fully updated in the service menu, then check that 
the Card type states MENU and the File system states ATA.
Check that the rest of the information for SLOT1 is reliable and no errors have been detected.

39. Data Card recognition
Insert a Memory card labelled Data in the right hand side memory card slot....

Page 242

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
30
Document no. M1031520-02                                                                           
Reset Case
Reset All --> Yes
or
Admit/Discharge
Discharge --> Yes
Check that the trends have been erased.
Reload the trends from the Data card by pressing the key 
Patient Data or Admit/Discharge, 
selecting Patient from Card, pressing the ComWheel on the last saved file (the file information 
is shown at the bottom of the menu) and...

Page 243

Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
31
                                          Document no. M1031520-02
NOTE: The counters may show values higher than 0. However, if any of the values is increasing 
continuously, it indicates a problem.

46. Hardware error counters
Check that the counters for hardware errors (Intern., Missed, FIFO, Overrun) show all 0. If any of 
the counters show a value higher than 0, replace the network board.

47. Disconnection recognition
Disconnect the...

Page 244

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
32
Document no. M1031520-02                                                                           
51. Connection to the Network
Make sure that the monitors wireless LAN antenna card is in line of sight with a wireless LAN 
access point antenna. Check that the monitor connects to the Datex-Ohmeda Network, i.e. the 
wireless LAN network symbol appears in the upper right-hand corner of the screen.

General
52. Trend retaining
Check...

Page 245

Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
33
                                          Document no. M1031520-02
56. Electrical safety check
Perform an electrical safety check and a leakage current test.
Check that the monitor functions normally after the performed electrical safety check.

57. Final cleaning
Switch the monitor to standby, disconnect the power cord and perform final cleaning.

Fill in all necessary documents. 

Page 246

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
34
Document no. M1031520-02                                                                           
3.3 Disassembly and reassembly
WARNINGThe S/5 Compact Anesthesia/ Critical Care Monitor is always energized by the 
internal battery. A short circuit may cause internal damage. Do not touch any 
exposed wiring or conductive surface inside, this may cause an electric shock.
CAUTION Perform leakage current measurement whenever service...

Page 247

Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
35
                                          Document no. M1031520-02
3.3.2 To disassemble the Display Unit of F-CM(C)(REC)1 rev.00...03
1. Set the Compact Monitor lying on its rear side.
2. Turn the quick lock screw at the bottom of the Display Unit counter clockwise to release it.
3. Pull the Display Unit carefully upwards and detach the Network and NetID cables.
4. Set the Display Unit lying on its front side; be careful not to damage the...

Page 248

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
36
Document no. M1031520-02                                                                           
3.3.3 To remove the Central Processing Board (CPB) from the Display Unit Back Unit
1. Disassemble Display Unit. 
2. Detach the fan and loudspeaker cables from the CPB, connectors X17 (loudspeaker) and X16 
(fan).
3. Detach the power cables coming from the DC/DC board, connector X10.
4. Detach the Signal cables (two) coming from the...

Page 249

Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
37
                                          Document no. M1031520-02
3.3.5 To remove the Synchronization Unit
1. Disassemble the Display Unit.
2. Detach carefully the Synchronization board foil cable by opening the lock at the connector X9 
and pulling the foil out.
3. Open the two screws holding the whole unit in the Display Unit back cover.
NOTE: Do not open the screws holding the Synchronization board in the Synchronization Unit.
3.3.6 To...

Page 250

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
38
Document no. M1031520-02                                                                           
In normal circumstances it is very difficult to keep the LCD Display component and the display shield 
surfaces free of dust when the LCD Display unit package is opened and they are exposed to room air. 
If dust particles remain on the LCD Display component and the display shield surfaces, they may 
impair the picture quality on the...