Table of figures
iii
                                                                                   Document no. M1031520-02
Table of figures 
Figure 1 Compact Anesthesia Monitor ................................................................................................................ 1
Figure 2 Basic structure of Display Unit .............................................................................................................. 4
Figure 3 Central Processing Board Block Diagram...

Page 212

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
iv
Document no. M1031520-02                                                                    

Page 213

Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
1
                                          Document no. M1031520-02
Introduction
This section provides information about the maintenance and service of the following products:
•S/5™ Compact Anesthesia Monitor F-CM1 Display Unit and S/5™ Compact Critical Care 
Monitor F-CMC1, Display Unit (including UPI function, LCD display, Command board function, 
Display Controller function, NET function, MEM function, DIS function, Display Unit power 
supply...

Page 214

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
2
Document no. M1031520-02                                                                           
1 Specifications
1.1 S/5 Compact Monitor, F-CM1, F-CMREC1, F-CMC1, F-CMCREC1; 
Display Unit
Display Unit Dimensions, 
D x W x H 55 x 335 x 270 mm (2.2 x 13.2 x 10.6 in)
Weight 

Page 215

Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
3
                                          Document no. M1031520-02
Data transmission IEEE 802.11b compliant, 
Direct Sequence Spread Spectrum (DSSS)
Roaming Automatic roaming between the Datex-Ohmeda Centrals
Access Points
Certificates Wi-Fi certified
Encryption Wired Equivalent Privacy (WEP) 40 and 128 bit
WLAN adapter PC card type with antenna
1.6 Recorder option
Power consumption 3 W
Printer type Thermal array
Print resolution Vertical 8...

Page 216

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
4
Document no. M1031520-02                                                                           
2 Display Unit functional description
The Compact Monitor Display Unit includes the following parts:
•Central Processing Board (CPB): includes Display Controller, UPI, CPU and optionally Network 
and MemCard functions
•Color LCD display and its backlight inverter 
•Connector board
•Command board 
•Display Unit DC/DC power supply
•Fan...

Page 217

Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
5
                                          Document no. M1031520-02
•2 serial channel signals in RS232-level
•Programmable alarm sound generator
•PC-card slot for software updates
2.1.1 Standby and Display Unit power down
When the monitor is turned to standby, NMI-interrupt is generated by the control logic. The interrupt 
signal in the Central Processing Board means that all the outputs from the CPB are switched off (i.e. 
display, alarm leds,...

Page 218

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
6
Document no. M1031520-02                                                                           
Figure 3 Central Processing Board Block Diagram 
2.1.3 SRAM MT48T37V 
The lithium battery back-up is a 32 Kbytes static RAM with a real-time clock.
CAUTION The IC contains a lithium battery. Discard the battery according to local 
regulations.
486
Pro
cessor
PC
MCIASe
rialEth
ernetSo
undUP
I
Chipse
t
Displ
ay
Co
ntroller486
 LOCAL...

Page 219

Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
7
                                          Document no. M1031520-02
2.2 UPI section
The UPI functions as a general I/O section. It performs I/O duties assigned to it by the CPU. The main 
processor and the processor on the UPI communicate through a dual-port memory (shared RAM) 
which is located on the UPI section.
2.2.1 Functional Blocks
The UPI section contains the CPU bus interface, a processor, program and dual-port memories, I/O 
block, and...

Page 220

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
8
Document no. M1031520-02                                                                           
2.3 Software License, L-CANE05, L-CANE05A, L-CICU05, L-CICU05A
L-CICU05/L-CICU05A software is developed for critical care monitoring and L-CANE05/L-CANE05A 
is for anesthesia monitoring. Anesthesia record keeping capability is included in the anesthesia 
monitor software L-CANE05/L-CANE05A as standard. The software L-CANE05A and...