GE S5 Service Manual
Here you can view all the pages of manual GE S5 Service Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.
Page 221
Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01) 9 Document no. M1031520-02 Figure 5 Display Controller Block Diagram 2.5 MEM section 2.5.1 Memory Card interface The memory interface consists of a PD6722 controller and two PCMCIA compatible card sockets. Both sockets consist of 60 signal and 8 power connections. The controller communicates with the CPU. Card insertions and removal are also detected by the controller. The MemCards are PCMCIA-ATA...
Page 222
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 10 Document no. M1031520-02 Figure 6 Memory Section Block Diagram 2.6 NET section The Ethernet interface controller is basically the heart of the NET section. The interface controller communicates with the CPU section through the CPU bus. The section also contains a 4k SRAM memory. The network interface controller transmits data packets to the Datex-Ohmeda...
Page 223
Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01) 11 Document no. M1031520-02 Figure 7 Central Processing Board, block diagram 2.6.1 Ethernet interface The data transformer is designed by Datex-Ohmeda and it is hospital grade approved. Adapter’s 10BASE-T is an interface with low-pass filters on the unisolated side of the transformer. On the isolated side there is a common mode choke for both transmitting and receiving lines. The three LEDs on the...
Page 224
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 12 Document no. M1031520-02 2.7 Wireless LAN Datex-Ohmeda Wireless Network is an extension to, or an alternative for, a wired Datex-Ohmeda Network. It provides same network services than the wired Datex-Ohmeda Network. The monitor network is a local area network based on standard Ethernet technology. The monitor network is formed by connecting one Central and...
Page 225
Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01) 13 Document no. M1031520-02 Figure 8 Central Processing Board parts 2.8 Front Unit NOTE: The LCD display backlight circuit runs on a high voltage. Do not touch the inverter board or the backlight tube leads when powered. Front Unit includes an LCD display component, Inverter Board, Connector board and Keyboard. The Front Unit is connected to the Central Processing Board with two cables: video signal...
Page 226
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 14 Document no. M1031520-02 2.8.1 Backlights The backlight consists of two replaceable cold-cathode fluorescent tubes driven by a separate inverter board. 2.8.2 Connector board The connector board is used for connecting the backlight power, keyboard and ComWheel signals between the Central Processing Board and the Front Unit. 2.8.3 Keyboard There are two...
Page 227
Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01) 15 Document no. M1031520-02 2.10 Connectors and signals 2.10.1 External connectors Figure 9 External connections of Compact Monitor frame F-CM(C)1 rev.04 (1) Air filter (2) NET connector (3) NET ID connector (not used with N-CMW) (4) Recorder (optional) (5) SYNC connector (6) Connector for external keyboard (anesthesia record keeping solution keyboard), K-CREMCO or Barcode Reader (7) Potential...
Page 228
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 16 Document no. M1031520-02 Figure 10 External connectors of Compact Monitor Frame rev.00...03 (1) Dust filters (2) Recorder unit (optional) (3) Connector for factory use only (4) Synchronization connector (5) Connector for external keyboard (anesthesia record keeping solution keyboard), K-CREMCO or Barcode Reader (6) Potential equalization connector (7)...
Page 229
Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01) 17 Document no. M1031520-02 DIS connector, X5 DIS interface Network connector, X5 Coding element connector, X4 Direct ECG (pin 7) Delay (max.): 15 ms Gain ECG (in)/ECG (out): 1 mV/1V NOTE: The direct ECG out signal is not available with modules M-ESTP rev. 01, M-EST rev. 00 and M-ETP rev. 00. Nurse call (pin 8) The nurse call signal is generated by the red and yellow alarms. When activated, the signal...
Page 230
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 18 Document no. M1031520-02 Keyboard connector, X9 Synchronization connector, X8 Direct ECG (PIN 3): Delay (max.): 15 ms Gain ECG (in)/ECG (out): 1 mV/1V NOTE: The direct ECG out signal is not available with modules M-ESTP rev. 01, M-EST rev. 00 and M-ETP rev. 00. Pressure out (PIN 4): P1 from hemodynamic module Gain Signal (out) / Pressure (in): 1V/100 mmHg...