Page 51

 Functional check43
Document no. M1187317-009
Preset the measurement settings for those parameters that are connected, for example:
Print/Recorder - Record Waveforms - Waveform 1 - ECG1
                 - Waveform 2 - P1
               - Waveform 3 - P2
Invasive Pressures  -  P1 ‘ART’ Setup -- Label – ART
- P2 ‘CVP’ Setup -- Label – CVP
Others -SPO2 Setup - Pleth Scale –AUTO
Airway Gas - CO2 Setup - Resp Rate Source – AUTO
- Measurement – ON
- Detection Limit – AUTO
5.3.2 Display
1. Check that the...

Page 52

Datex-Ohmeda S/5 FM
44
Document no. M1187317-009
Breathe to the sampling line briefly. Check that the CO2 information is updated on the 
screen.

5.3.6 Multiparameter Hemodynamic Modules
ECG and RESP measurements
1. Connect an ECG cable to the module. Connect the cable leads to a patient simulator. 
Check that all ECG and impedance respiratio n information is shown on the monitor 
screen as configured on the simulator.
Turn the simulator off. Check that the ‘Asystole’ and ‘Apnea’ messages appear on the...

Page 53

 Functional check45
Document no. M1187317-009
5.3.7 Data Card and Menu Card function
1. Insert a Data card or a Menu card to the slot.Check that the corresponding symbol appears on the monitor screen.

5.3.8 Recorder 
1. Press the Start/Stop  sidepanel key and check that the module starts recording the 
selected waveforms. Press the 
Start/Stop  sidepanel  key again to stop recording.
2. Check that the quality of the recordings is acceptable.

5.3.9 Network connection
1. Check that the Mon-Net cable...

Page 54

Datex-Ohmeda S/5 FM
46
Document no. M1187317-009
5.3.12 General•
Switch the monitor to standby
•Perform final cleaning
•Fill in all necessary documents
 

Page 55

 General troubleshooting47
Document no. M1187317-009
6 General troubleshooting    
Figure 11 S/5 FM general  troubleshooting flowchart
No picture on screen
Main CPU board may be
faulty.
Yes
No
NIBP pump starts
pumping and SpO2
probes red LED lit?No
No
FM_general_trbl_flowchart.vsd
Disconnect the mains power cord and
remove the batteries to make a cold start.
Connect the mains power cord
Check the batteries charging levelsby pressing the test buttons on the
batteries.
Insert only one battery with 2 or...

Page 56

Datex-Ohmeda S/5 FM
48
Document no. M1187317-009
For your notes: 

Page 57

Appendix A Functional check form, Datex-Ohmeda S/5 FMA-1(2)
                           Document no. M1187317-009
APPENDIX A: Functional check form, 
Datex-Ohmeda S/5
 FM 
OK = Test OK            N.A. = Test not applicable            Fail = Test failed
Customer
Service
Service engineer
Date
Measuring equipment / test gases used:
Equipment / tool / gas:  Manufacturer: Model/Type/Part 
Number: Serial Number / 
ID:Calibration 
Date:
Monitor Installation
L- E- N-
F-
Visual Inspection          OK...

Page 58

Datex-Ohmeda S/5 FM
A-2(2)
Document no. M1187317-009
5.3.6. Multiparameter Hemodynamic ModulesS/N
                           
. ECG and RESP measurements                            
. Temperature measurement                            
. Invasive blood pressure measurement                            
. SpO2 measurement                            
. Non invasive blood pressure measurement                            
5.3.7. Data Card and Menu Card function                            
5.3.8. Recorder...

Page 59

Appendix B ElectroMagnetic CompatibilityB-1(4)
                           Document no. M1187317-009
APPENDIX B: ElectroMagnetic Compatibility
Table 1 Guidance and manufacturer’s  declaration – electromagnetic 
emissions
(1  S/5 FM equipped with S/5 Frame for FM, F-FM 
(2  S/5 FM equipped with S/5 Frame for FM with WLAN, F-FMW
Guidance and manufacturer’s declar ation – electromagnetic emissions
The S/5 FM is intended for use in the electromagne tic environment specified below. The customer or the 
user of...

Page 60

Datex-Ohmeda S/5 FM
B-2(4)
Document no. M1187317-009
Table 2 Guidance and manufacturer’s declaration – electromagnetic 
immunity
Guidance and manufacturer’s decl aration – electromagnetic immunity
The S/5 FM is intended for use in the electromagnetic en vironment specified below. The customer or the user of 
the S/5 FM should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment -  guidance
Electrostatic 
discharge (ESD)
IEC...