Interfacing33
Document no. M1187317-009
4.1.1 Device Interfacing Solution components
The Device Interfacing Solution consists of:
a device specific interfacing module
 a device specific cable
 a bus cable
 a connector for another bus cable
 label specifying the external device
4.1.2 Connections
Connect the device specific cable to the external  device and the bus cable to the monitors DIS 
connector or to another interfacing module. 
Figure 8 Connection cables and LED indicators
WARNING The monitor,...

Page 42

Datex-Ohmeda S/5 FM
34
Document no. M1187317-009
4.1.3 Mounting
The DIS module can be mounted on the side panel of the external device. Also IV pole 
placement is possible.
NOTE: As the Device Interfacing Solution works on ly with the device specified in the label of the 
interfacing module, it is recommended that the  interfacing module always travels along with 
the external device.
For mounting accessories, please refer to the  “Supplies and Accessories” catalog. See the 
following figure for an...

Page 43

Interfacing35
Document no. M1187317-009
4.1.5 Functional check
There are two ways to check the function of the Device Interfacing Solution:
1. Press the 
Monitor Setup key.
2. Select  Interfacing and open the  Status Page menu. The status page shows you the 
current communication status of  the interfacing module (1 - 4). 
NOTE: The status message ‘Connected ’ appears on the Status Page after you have connected 
the external device to the interfacing module an d turned it on. Note also that the monitor...

Page 44

Datex-Ohmeda S/5 FM
36
Document no. M1187317-009
4.2 Interfacing printer
It is possible to interface a laser printer (serial or parallel) with the S/5 FM via the monitors’ serial 
port. The printer should be PCL5 compatible and should contain at least 2 MB of memory. 
Parallel printers require the us e of Serial-to-Parallel Converter, order code 78030, model PI 
130-R2.
WARNING Always make sure that the combinat ion complies with the international 
safety standard IEC 60601-1-1 for medical electrical...

Page 45

Interfacing 37
Document no. M1187317-009
3. Check the DIP-switch settin gs on the converter (order code 78030, model PI130-R2):
4. Connect a DB25 female-to-female gender changer (order code 78032) between the  printer cable and converter.      
Figure 10 Connecting S/5 FM monitor  to printer, converter model PI130-R2
5. Connect a common RS-232 modem cable’s  (order code 78031) DB9 connector to the 
serial port connector X9 on the monitor,  and the DB25 connector to the converter.
The new Serial to...

Page 46

Datex-Ohmeda S/5 FM
38
Document no. M1187317-009
4.4 Output signals
Analog/digital output signals on the connectors X5 and X8 can be used for interface with other 
devices. The pin assignments are illu strated in tables/pictures below.
Table 5 NET ID connector, X8 on Multi I/O adapter 
Table 6 Defib & IABP sync connector, X5 
4.4.1 Digital outputs
The digital output signals are as follows:
Defibrillation Sync (X5 pin 1)
Def ibrillation Sync indication is generated by ECG. When active, the signal is state...

Page 47

Interfacing 39
Document no. M1187317-009
The pacemaker pulses have been replaced with 2 ms /5 V fixed digital pulses at the ECG analog 
output for IABP or defibrillator  equipment. A device that fulfils  the requirements of the IEC 
60601-1 standard can be connected to the defibrillator & IABP synchronization connector. 
There are no other limitations, because the signals of the connector are galvanically isolated 
from patient applied part of the ECG and invasive blood pressure measurements.
Direct ECG...

Page 48

Datex-Ohmeda S/5 FM
40
Document no. M1187317-009
5 Functional check
These instructions include procedures for a functional check for Datex-Ohmeda S/5 FM. The 
functional check is mandatory after monitor installation.
These instructions include a  “Functional check form, Datex-Ohmeda S/5 FM” which may be 
used when performing the procedures. The symbol 
 in the instructions  indicates that the 
check form contains space to record the results of the particular procedure. The procedures 
should be...

Page 49

 Functional check41
Document no. M1187317-009
5.1 Recommended tools
NOTE: Use only properly maintained, calibrated and traceable measurement equipment for the 
specified calibrations and adjustments to ensure accuracy. 
For details on recommended accessories s ee the “Supplies and Accessories” catalog.
Table 7 Patient simulators’ compatibility with each hemodynamic module
For product(s)ToolOrder No.
Airway modules
N-FC(REC) Calibration gas755580
Regulator 755534-HEL
N-FC(REC) CO
2 Sampling line 3m/10 ft...

Page 50

Datex-Ohmeda S/5 FM
42
Document no. M1187317-009
Table 8 Adapter cables for hemodynamic patient simulators
5.2 Visual inspection
Make sure that the monitor is switched to standby.
Disconnect the mains power cord from the monitor. 
If the monitor is connected to the Datex-Ohmeda Network, disconnect the Mon-Net cable 
from the monitor.
1. Check all units visually. Check that all parts are intact and that the cables and screws are connected and 
tightened properly.  Especially check the following parts:...