Page 41
Interfacing33 Document no. M1187317-009 4.1.1 Device Interfacing Solution components The Device Interfacing Solution consists of: a device specific interfacing module a device specific cable a bus cable a connector for another bus cable label specifying the external device 4.1.2 Connections Connect the device specific cable to the external device and the bus cable to the monitors DIS connector or to another interfacing module. Figure 8 Connection cables and LED indicators WARNING The monitor,...
Page 42
Datex-Ohmeda S/5 FM 34 Document no. M1187317-009 4.1.3 Mounting The DIS module can be mounted on the side panel of the external device. Also IV pole placement is possible. NOTE: As the Device Interfacing Solution works on ly with the device specified in the label of the interfacing module, it is recommended that the interfacing module always travels along with the external device. For mounting accessories, please refer to the “Supplies and Accessories” catalog. See the following figure for an...
Page 43
Interfacing35 Document no. M1187317-009 4.1.5 Functional check There are two ways to check the function of the Device Interfacing Solution: 1. Press the Monitor Setup key. 2. Select Interfacing and open the Status Page menu. The status page shows you the current communication status of the interfacing module (1 - 4). NOTE: The status message ‘Connected ’ appears on the Status Page after you have connected the external device to the interfacing module an d turned it on. Note also that the monitor...
Page 44
Datex-Ohmeda S/5 FM 36 Document no. M1187317-009 4.2 Interfacing printer It is possible to interface a laser printer (serial or parallel) with the S/5 FM via the monitors’ serial port. The printer should be PCL5 compatible and should contain at least 2 MB of memory. Parallel printers require the us e of Serial-to-Parallel Converter, order code 78030, model PI 130-R2. WARNING Always make sure that the combinat ion complies with the international safety standard IEC 60601-1-1 for medical electrical...
Page 45
Interfacing 37 Document no. M1187317-009 3. Check the DIP-switch settin gs on the converter (order code 78030, model PI130-R2): 4. Connect a DB25 female-to-female gender changer (order code 78032) between the printer cable and converter. Figure 10 Connecting S/5 FM monitor to printer, converter model PI130-R2 5. Connect a common RS-232 modem cable’s (order code 78031) DB9 connector to the serial port connector X9 on the monitor, and the DB25 connector to the converter. The new Serial to...
Page 46
Datex-Ohmeda S/5 FM 38 Document no. M1187317-009 4.4 Output signals Analog/digital output signals on the connectors X5 and X8 can be used for interface with other devices. The pin assignments are illu strated in tables/pictures below. Table 5 NET ID connector, X8 on Multi I/O adapter Table 6 Defib & IABP sync connector, X5 4.4.1 Digital outputs The digital output signals are as follows: Defibrillation Sync (X5 pin 1) Def ibrillation Sync indication is generated by ECG. When active, the signal is state...
Page 47
Interfacing 39 Document no. M1187317-009 The pacemaker pulses have been replaced with 2 ms /5 V fixed digital pulses at the ECG analog output for IABP or defibrillator equipment. A device that fulfils the requirements of the IEC 60601-1 standard can be connected to the defibrillator & IABP synchronization connector. There are no other limitations, because the signals of the connector are galvanically isolated from patient applied part of the ECG and invasive blood pressure measurements. Direct ECG...
Page 48
Datex-Ohmeda S/5 FM 40 Document no. M1187317-009 5 Functional check These instructions include procedures for a functional check for Datex-Ohmeda S/5 FM. The functional check is mandatory after monitor installation. These instructions include a “Functional check form, Datex-Ohmeda S/5 FM” which may be used when performing the procedures. The symbol in the instructions indicates that the check form contains space to record the results of the particular procedure. The procedures should be...
Page 49
Functional check41 Document no. M1187317-009 5.1 Recommended tools NOTE: Use only properly maintained, calibrated and traceable measurement equipment for the specified calibrations and adjustments to ensure accuracy. For details on recommended accessories s ee the “Supplies and Accessories” catalog. Table 7 Patient simulators’ compatibility with each hemodynamic module For product(s)ToolOrder No. Airway modules N-FC(REC) Calibration gas755580 Regulator 755534-HEL N-FC(REC) CO 2 Sampling line 3m/10 ft...
Page 50
Datex-Ohmeda S/5 FM 42 Document no. M1187317-009 Table 8 Adapter cables for hemodynamic patient simulators 5.2 Visual inspection Make sure that the monitor is switched to standby. Disconnect the mains power cord from the monitor. If the monitor is connected to the Datex-Ohmeda Network, disconnect the Mon-Net cable from the monitor. 1. Check all units visually. Check that all parts are intact and that the cables and screws are connected and tightened properly. Especially check the following parts:...