GE S5 Owners Manual
Here you can view all the pages of manual GE S5 Owners Manual. The GE manuals for Medical Monitor are available online for free. You can easily download all the documents as PDF.
Page 21
Introduction13 Document no. M1187317-009 1.2.2 ESD precautionary procedures • To avoid electrostatic charges building up, it is recommended to store, maintain and use the equipment at a relative humidity of 30% or greater. Floors should be covered by ESD dissipative carpets or similar. Non-synthetic clothing should be used when working with the component. •To prevent applying a possible electrostatic di scharge to the ESD sensitive parts of the equipment, one should touch the metallic fram e of the...
Page 22
Datex-Ohmeda S/5 FM 14 Document no. M1187317-009 2 System description 2.1 Introduction Datex-Ohmeda monitors build up a freely configurable modular system. The architecture is designed to enable different module combinations so that the user is able to get the desirable parameter and feature set. This modular approach makes it possible to add new features when they are needed. 2.2 Bus structure The operation of Datex-Ohmeda FM is based on two communication channels, the CPU bus and module bus. All...
Page 23
System description15 Document no. M1187317-009 The RS-485 type of serial communication su pports so-called multidrop or party line connections. This means that all parameter modules connected to the module bus use exactly the same lines for communication. The advant age of this is that all bus connectors are identical and the modules can be connected in any order and position. 2.3 Distributed processing A system assembled from Datex-Ohmeda products is a multiprocessor system. All parameter modules...
Page 24
Datex-Ohmeda S/5 FM 16 Document no. M1187317-009 The UPI processor collects and stores all data that is received from the parameter modules into a shared SRAM, which is mapped directly to the address space of the main CPU. The main CPU reads data from the memory while the UPI processor guarantees that the data is up to date. This operation also works in the other direction. In this the main CPU fills the shared SRAM with data and the UPI processor distributes it to the parameter modules. 2.5 Software...
Page 25
System description17 Document no. M1187317-009 2.6 Parameter modules Figure 4 General structure of parame ter modules with patient isolation The detailed structure of a parameter module depends on the specific needs for each individual parameter. However, some common parts are used in the parameter modules. The electronics inside the module is usually divided into isolated (floating) and non-isolated sections. Typically, the non-isolated section cons ists of buffers to interface the parameter...
Page 26
Datex-Ohmeda S/5 FM 18 Document no. M1187317-009 3 System installation 3.1 Unpacking instructions 1. Confirm that the packing box is undamaged. If the box is damaged, contact the shipper. 2. Open the top of the box and carefully unpack all components. 3. Confirm that all components are undamaged. If any of the components is damaged, contact the shipper. 4. Confirm that all components are included. If any of the components is missing, contact your GE Healthcare distributor. 3.2 Choosing location Consider...
Page 27
System installation19 Document no. M1187317-009 Figure 5 E-PSM(P) mounting accessories 1. M1051025 Module Bus Adapter for PSM 2. M1049197 Pole Mount for PSM, short 3. M1051023 Pole Mount for PSM, long WARNINGMake sure that the Pole Mo unt for PSM is always used in vertical position to prevent water from enteri ng the E-PSM(P) module. Pole Mount for PSM – Instructions for connecting to an IV pole, vertical position Pole Mount for PSM - Instructions fo r installing in horizontal position. Remove the 2...
Page 28
Datex-Ohmeda S/5 FM 20 Document no. M1187317-009 3.3.2 Monitor connections Figure 6 S/5 FM front panel (1) Battery compartment (2) Transportation handle (3) Alarm light (4) Alarm LED indicators (5) Side panel keys (6) The ComWheel (7) Command Board keys (8) Guide rail for GCX mounting (9) Mains power and battery LEDs (10) ON/standby key (11) Connector for the Device Interfacing Solution (X6) (12) Connector for defibrillator synchronization (X5) (13) Measurement modules (14) Slot for Data Card or...
Page 29
System installation21 Document no. M1187317-009 Figure 7 Rear panel connections (1) Slot for infusion pole mounting (2) Module connector (3) Guide rail for GCX mounting (4) Receptacle for power cord (5) Fuse holder (6) Serial port X9 (7) Network ID X8 (8) Connector for K-CREMCO X7 (9) Optional: Multi I/O adapter incorporating connectors 6 - 8 mentioned above (10) Network connector X1 (11) Equipotential connector (12) Slot for Data Card or Menu Card (13) Defibrillator & IABP sync connector X5...
Page 30
Datex-Ohmeda S/5 FM 22 Document no. M1187317-009 3.3.3 Connection to mains Connect the power cord to the mains power inlet (4) at the back of the monitor and to the wall socket. NOTE: Before taking the monitor into use for the first time, the batteries should be fully charged. Keep the monitor connected to the mains until the Battery charging symbol disappears, or in STBY mode the Orange Battery condition LED is off (may take up to 5 hours if the batteries are fully discharged). WARNING The power...