Page 71
Planned maintenance instructions3 Document no. M1187318-003 1.2.1 Hemodynamic patient simulators The following tables present the patient simulators’ compat ibility with each hemodynamic module, and the accessories needed: Table 1 Patient simulators’ compatibility with each hemodynamic module Table 2 Adapter cables for hemodynamic patient simulators 1.3 Recommended parts Module Parameter Patient simulator M1010831 MedSimLionheart & MPS450 E-PSM(P) ECG Cable included Mult ilink ECG acc. Multilink ECG...
Page 72
Datex-Ohmeda S/5 FM 4 Document no. M1187318-003 1.4 Planned maintenance parts 1.4.1 PM parts for Airway Module, N-FC(REC) 1.4.2 PM parts for S/5 FM frame, F-FM For details on recommended accessories s ee the “Supplies and Accessories” catalog. Part Order No. For product(s) Special tube (Nafion 2 pcs) 733382-HEL All Airway modules Mini D-fend O-ring (2 pcs) 656565 N-FC(REC) Mini D-fend 8002174 (pkg of 10 pcs) N-FC(REC) Zero valve air filter M1011471 N-FC(REC), every 3 years CO 2 Sampling line 3.0 m...
Page 73
Planned maintenance check list5 Document no. M1187318-003 2 Planned maintenance check list 2.1 Visual inspection/preparation 2.1.1 General WARNINGWear a grounded antistatic wristband when handling PC boards. Electrostatic discharge may damage components on the board. Make sure that the monitor is turned off. Disconnect the mains power cord. If the monitor is connected to the Datex-Ohmeda Network, disconnect the Mon-Net cable from the monitor. Remove any memory cards. 1. Check all the units visually....
Page 74
Datex-Ohmeda S/5 FM 6 Document no. M1187318-003 2.1.3 Recorder unit 1. Clean the recorder.−Open the paper compartment hatch and remove the paper roll, if installed. − Remove any paper chaff from the paper compartment. − Clean the thermal printhead and the small glass window in front of the static brush with a cotton swab dipped in isopropyl alcohol, if necessary. NOTE: Avoid contact with the rubber paper roller. Be careful to limit the application of alcohol to the thermal printhead and the window....
Page 75
Planned maintenance check list7 Document no. M1187318-003 4. Check that the monitor goes to battery use if the main cord is disconnected. 2.2.2 Display 1. Check that the picture on the screen is correct. 2.2.3 Keyboard(s) 1. Tests with the Command Board: −Press the Monitor Setup key. Turn the ComWheel in bo th directions and check that the cursor in the menu moves correspondingly. Select Normal Screen and check that the menu disappears from the screen. Check the rest of the menu keys by pressing...
Page 76
Datex-Ohmeda S/5 FM 8 Document no. M1187318-003 NOTE: Noisy sampling pump might indicate possible problems with motor bearing. Replace the noisy sampling pump by a new one if needed. 2.2.5 Multiparameter Hemodynamic Modules ECG and RESP measurements 1. Enter the ECG service menu. Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing faster than by 5 per second. Check that the ECG/RESP board memories have passed the internal memory test, i.e. RAM, ROM and EEPROM all...
Page 77
Planned maintenance check list9 Document no. M1187318-003 Invasive blood pressure measurement 8. Check the function of the module front panel membrane keys. 9. Check the InvBP channels with a patient simulator.Zero the InvBP channels, then check that the values and waveforms correspond to the simulator settings. SpO2 measurement 10. Check that the message ‘No probe’ is shown, when no SpO2 sensor is connected. Connect an SpO 2 finger probe to the module. Check that the message ‘Probe off’ is shown,...
Page 78
Datex-Ohmeda S/5 FM 10 Document no. M1187318-003 3. Select Communication . Check that the Interface status states Active continuously and the error counter values on the bottom part of the menu are stable. 4. Select Status. Insert Data card into the slot. Wait until the info rmation is fully updated in the service menu, then check that the Card types are correct and the ‘File system’ states ATA. Check that the rest of the information is reliable and no errors have been detected. 2.2.7 Recorder 1....
Page 79
Planned maintenance check list11 Document no. M1187318-003 2.2.10 Device Interfacing Solution, N-DISxxx 1. Enter the DIS Interfacing service menu: Check that the DIS module ‘tout’ and ‘cse’ values do not increase faster than by 5 per second. Check also that the DIS module memories have passed the internal memory test, i.e. Ram, Rom and EEPROM state all OK. Perform the same check for all connected DIS modules. 2.2.11 General 1. Storing trend data Check that the monitor is capable of storing the...