Service Menu23 Document no. M1187329-003 4.1.1 General Service Data Module configuration shows which measurement options are available, i.e. are detected by the module. Timeouts is a cumulative number that indicates how many times the module has not responded to the monitors inquiry. Bad checksums is a cumulative number that indicates how many times communication from the module to monitor has broken down. Bad c-s by mod is a cumulative number that indicates how many communication errors...
Page 112
Datex-Ohmeda S/5 FM 24 Document no. M1187329-003 4.1.2 Gases Noise Measactivates the noise measurement. Sample gain adj adjusts the sampling pump gain, i.e. for adjusting the sample flow measurement. Pump ctrl A manual control for the sampling pump. Zero valve ctrl A manual control for the zero valve. Record Data prints out the shown service data and board information (id. serial number and software id.) onto the recorder module, N-FCREC, N- FREC. Service Data CO2% field shows real-time...
Page 113
Service Menu25 Document no. M1187329-003 4.2 ECG Module Power freq: Set power frequency; 50 Hz/60 Hz. Filter low : Set filter low frequency; 0.05 Hz/0.5 Hz. Filter high: Set filter high frequency; 30 Hz (40 Hz if power freq is 60 Hz) /150 Hz. Service Data Power freq, and Cable type show the values chosen or detected, Filter low and high defines the selected filter (Monitor/Diagnostic/ST). Quick zero module is ON when the ECG signal is beyond the scale, and therefore, is quickly returned to...
Page 114
Datex-Ohmeda S/5 FM 26 Document no. M1187329-003 4.2.1 ECG Setup Filter filters the ECG signal high frequency noise and slow respiratory artefacts. Monit (monitor) filter is used in routine monitoring. It effectively filters the artefacts caused by the electrosurgery unit and respiration. Diagn (diagnostic) filter is used if more accurate information of the waveform is needed (e.g. of P- wave or AV block). The diagnostic filter is more susceptible both to high frequencies and baseline wander...
Page 115
Service Menu27 Document no. M1187329-003 4.3 PSM STP Record Data prints out the shown service data and board information (id, serial number and sw id) onto the recorder. Temp Test activates the automatic temperature test for the temperature channels T1 and T2. The result from the test is shown in the service data field. NOTE: The Temp Test needs to be selected twice before the test starts. Service Data field Gain is a coefficient to co mpensate gain error. Usually the values for P1 and P2 are...
Page 116
Datex-Ohmeda S/5 FM 28 Document no. M1187329-003 Balance between leds is adjusted by changing the intensity of red/infrared. Intensity of infrared (Ired int.) is in the range of 40 to 255 and red intensity ( Red int .) is in the range of 40 to 255. DC gain shows the gain of DC signal adjusted by the module. IDC is the value of infrared signal. RDC is the dc value of red signal. AC gain is the gain of infrared and red ac signals. AC gain values can be 1 or 0. Value 1 means high ac gain and 0 means...
Page 117
Service Menu29 Document no. M1187329-003 4.3.1 CalibrationsHow to calibrate T1/ T2 The calibrations are possible only when the protection is set OFF. The temperature calibration requires accurate test plugs of value 25 °C and 45 °C. 1. Select Calibrate T1/Calibrate T2. 2. Insert the test plug 25 °C into the T1/T2 connector. 3. Press the ComWheel. 4. Insert the test plug 45 °C into the T1/T2 connector. 5. Press the ComWheel. How to calibrate P1/ P2 The calibrations require a pressure transducer...
Page 118
Datex-Ohmeda S/5 FM 30 Document no. M1187329-003 4.4 NIBP Module Service menu for Non-invasive Blood Pressure Modules. Service Data Pressure shows the measured pressure multiplied by 10. This value is automatically zero-drift compensated. Zero shows the difference between the zeroing value in the permanent memory (stored when the module is calibrated) and the current automatic zero-drift compensation multiplied by 10. The value can change between +20 and -20 mmHg. If the zero drift exceeds...
Page 119
Service Menu31 Document no. M1187329-003 4.4.1 NIBP Demo A service menu for demonstrating the oscillometric method of NIBP measurement. The menu shows the real-time pressure signals that are measured from the NIBP cuff. The measurement result is shown in the adjoining digit field. Wave Recording is for selecting the recording option. If ON is selected, the pressure signals are printed out in real-time on the recorder after the recorder Start/Stop button is pressed. Remove menu widens the...
Page 120
Datex-Ohmeda S/5 FM 32 Document no. M1187329-003 4.4.2 NIBP CalibrationHow to Calibrate Calibration selection is available only when protection is OFF. NOTE: Both channels B1 and B2 must be calibrated simultaneously. NOTE: The module must be in the frame during the whole procedure. NIBP calibration can be performed in the NIBP Service menu as follows: 1. If Protection is ON, change it to OFF by pressing the NIBP module buttons Auto ON/OFF and Start Cancel simultaneously for 3 seconds, which...