Service Menu23
Document no. M1187329-003
4.1.1 General
 
Service Data
Module configuration  shows which measurement 
options are available, i.e. are detected by the module.
Timeouts  is a cumulative number that indicates how 
many times the module has not responded to the 
monitors inquiry. 
Bad checksums  is a cumulative number that 
indicates how many times communication from the 
module to monitor has broken down.
Bad c-s by mod  is a cumulative number that indicates 
how many communication  errors...

Page 112

Datex-Ohmeda S/5 FM
24
Document no. M1187329-003
4.1.2 Gases
      
Noise Measactivates the noise measurement.
Sample gain adj adjusts the sampling pump gain,  i.e. for adjusting the sample flow 
measurement.
Pump ctrl A manual control for the sampling 
pump.
Zero valve ctrl A manual control for the zero 
valve.
Record Data prints out the shown service data 
and board information (id. serial 
number and software id.) onto the 
recorder module, N-FCREC, N-
FREC.
Service Data
CO2% field  shows real-time...

Page 113

Service Menu25
Document no. M1187329-003
4.2 ECG Module
Power freq:  Set power frequency; 50 Hz/60 Hz.
Filter low
: Set filter low frequency; 0.05 Hz/0.5 Hz.
Filter high:  Set filter high frequency; 30 Hz (40 Hz if 
power freq is 60 Hz) /150 Hz.
Service Data
Power freq, and Cable type  show the values chosen 
or detected,  Filter low and high  defines the selected 
filter (Monitor/Diagnostic/ST).
Quick zero  module is ON when the ECG signal is 
beyond the scale, and therefore, is quickly returned to...

Page 114

Datex-Ohmeda S/5 FM
26
Document no. M1187329-003
4.2.1 ECG Setup
Filter filters the ECG signal high frequency noise and 
slow respiratory artefacts.
Monit  (monitor) filter is used  in routine monitoring. It 
effectively filters the artefacts caused by the 
electrosurgery unit and respiration. 
Diagn  (diagnostic) filter is used if more accurate 
information of the waveform is needed (e.g. of P-
wave or AV block). The diagnostic filter is more 
susceptible both to high frequencies and baseline 
wander...

Page 115

Service Menu27
Document no. M1187329-003
4.3 PSM STP
Record Data  prints out the shown service 
data and board information (id, serial number 
and sw id) onto the recorder.
Temp Test  activates the automatic 
temperature test for the temperature 
channels T1 and T2. The result from the test is 
shown in the service data field.
NOTE: The Temp Test needs to be selected 
twice before the test starts.
Service Data field
Gain  is a coefficient to co mpensate gain error. 
Usually the values for P1 and P2 are...

Page 116

Datex-Ohmeda S/5 FM
28
Document no. M1187329-003
Balance between leds is adjusted by changing the intensity of red/infrared. Intensity of infrared (Ired int.) is in 
the range of 40 to 255 and red intensity ( Red int .) is in the range of 40 to 255.
DC gain  shows the gain of DC signal adjusted by the module.
IDC  is the value of infrared signal.
RDC  is the dc value of red signal.
AC gain  is the gain of infrared and red ac signals. AC gain  values can be 1 or 0. Value 1 means high ac gain 
and 0 means...

Page 117

Service Menu29
Document no. M1187329-003
4.3.1 CalibrationsHow to calibrate T1/ T2
 The calibrations are possible only when the protection is set  OFF. The temperature calibration 
requires accurate test plugs of value 25 °C and 45 °C.
1. Select  Calibrate T1/Calibrate T2.
2. Insert the test plug 25  °C into the T1/T2 connector.
3. Press the ComWheel.
4. Insert the test plug 45  °C into the T1/T2 connector.
5. Press the ComWheel.
How to calibrate P1/ P2
 The calibrations require a pressure transducer...

Page 118

Datex-Ohmeda S/5 FM
30
Document no. M1187329-003
4.4 NIBP Module
   
Service menu for Non-invasive Blood Pressure 
Modules.
Service Data
Pressure shows the measured pressure multiplied by 
10. This value is automatically zero-drift 
compensated.
Zero  shows the difference between the zeroing value 
in the permanent memory (stored when the module 
is calibrated) and the current automatic zero-drift 
compensation multiplied by 10. The value can 
change between +20 and -20 mmHg. If the zero drift 
exceeds...

Page 119

Service Menu31
Document no. M1187329-003
4.4.1 NIBP Demo
A service menu for demonstrating the 
oscillometric method of  NIBP measurement. The 
menu shows the real-time  pressure signals that 
are measured from the NIBP cuff. The 
measurement result is shown in the adjoining digit 
field.
Wave Recording  is for selecting the recording 
option. If ON is selected, the pressure signals are 
printed out in real-time on the recorder after the 
recorder 
Start/Stop button is pressed.
Remove menu widens the...

Page 120

Datex-Ohmeda S/5 FM
32
Document no. M1187329-003
4.4.2 NIBP CalibrationHow to Calibrate
Calibration selection is available only when protection is OFF.  
NOTE: Both channels B1 and B2 must be calibrated simultaneously.
NOTE: The module must be in the frame during the whole procedure.
NIBP calibration can be performed in the NIBP Service menu as follows:
1. If Protection is  ON, change it to OFF  by pressing the NIBP module buttons 
Auto ON/OFF 
and 
Start Cancel simultaneously for 3 seconds, which...