Page 141
List of figuresiii Document no. M1187335-004 Table of f igures Figure 1 S/5 Frame for FM ...................................................................................................... ................................................................... 1 Figure 2 FM Monitor connection block diag ram .............................................................................................................................. 6 Figure 3 CPU board F-FM(W)-01...
Page 143
Frame for FM 1 Document no. M1187335-004 Introduction This section provides information about th e maintenance and service of the following products: •S/5 Frame for FM, F-FM. •Software Licenses L-FICU04, L- FICU04A. L-FICU06, L-FICU06A. •WLAN option F-FMW. Figure 1 S/5 Frame for FM
Page 144
Datex-Ohmeda S/5 FM 2 Document no. M1187335-004 1Specifications 1.1 Frame for FM Dimensions Height 302 mm (11.8 in)Handle Up 274 mm (10.7 in)Handle Down Width 302 mm (11.8 in) Without modules 337 mm (13.3 in) With modules Depth 143 mm (5.6 in) Without extension module 156 mm (6.1 in) With extension module 200 mm (7.8 in) With extension and PSM modules Weight FM 5.0 kg FM without modules, batteries included FM + E-PSM 5.6 kg FM with PSM FM + N-FCREC + E-PSM 6.6 kg FM with PSM...
Page 145
Frame for FM 3 Document no. M1187335-004 Vertical viewing angles Lower side 45° Upper side 35° Backlight cold cathode fluorescent lamp, replaceable Backlight lifetime 40 000 h Backlight intensity 250 Cd/m 2 typ. Integrated Command Board with direct function ke ys, menu keys and ComWheel for selections and adjustments in menus. 1.3 Batteries Number of batteries 2 Battery type Exchangeable Lithium-Ion Voltage 11.1V (nominal) Capacity 3.5 Ah Capacity indicator Fuel gauge on monitor screen and on battery...
Page 146
Datex-Ohmeda S/5 FM 4 Document no. M1187335-004 1.7 Analog Outputs, direct ECG and synchronization ECGFrom first user lead (ECG1) Gain: 1 V/1mV ±20% Delay: < 15 ms DC offset: ±100 mV max. Output range: ±4 V Noise: 50 mVpp max. Frequency response: 0.05 Hz to 40 Hz InvBP From pressure channel P1 Gain: 10 mV/1 mmHg ±20% Delay: < 35 ms DC offset: ±100 mV Output range: -0.4 V to +3.2 V Noise: 50 mVpp max. The pacemaker pulses have been replaced with 2 ms/5 V fixed digital pulses at the ECG analog...
Page 147
Frame for FM 5 Document no. M1187335-004 2 Functional description 2.1 Main components 2.1.1 Keyboards User interface parts Vertical and Horizontal Membrane keypad cont aining 20 keys. The keypads are foil membrane keypads. The keypads are connected to the UPI section of the CPU board. ComWheel is used for menu selection. 2.1.2 Display The 10.4” LCD display with SVGA 800 x 600 reso lution has bright long live lamps and a wide viewing angle. NOTE: The LCD display backlight circuit runs on a high...
Page 148
Datex-Ohmeda S/5 FM 6 Document no. M1187335-004 Figure 2 FM Monitor connection block diagram FLEXIBLE MONITOR (FM) CONNECTION BLOCK DIAGRAM EXTERNAL DC SUPPLY 10-16.5VSMART BATTERY A Li-ION 9-12.6V SMART BATTERY B Li-ION 9-12.6VAC INLET 90-264Vac 47-63Hz FILTER BOARD AC/DC POWER SUPPLY DC/DC BOARD SW LOAD PCMCIA CARD CF WLAN CARD NETWORK CONNECTOR BOARD NETWORK CONNECTOR RJ-45 10BASE-T DISPLAY CONNECTION BOARD BACKLIGHT INVERTER LCD DISPLAY 10.4" SVGA CCFL BACKLIGHT LAMPS CPU BOARD MULTI I/O ADAPTER...
Page 149
Frame for FM 7 Document no. M1187335-004 2.1.3 CPU boardF-FM(W)-01 Figure 3 CPU board F-FM(W)-01 The board is based on AMD Élan SC520 microcontroller. Other features include a flat panel display controller, 10/100 Mbit/s Ethernet interface, PCMCIA interface, sound system and Renesas H8SX based UPI. The CPU section takes care of the central processing. The main features are: •Élan SC520 processor •Internal clock frequency 133MHz •32 Mbytes SDRAM •16+16 Mbytes flash memory •32 Kbytes static RAM with...
Page 150
Datex-Ohmeda S/5 FM 8 Document no. M1187335-004 Ethernet connector X14 (network cable to network connector board) System watchdog and voltage supervision There is a voltage supervisor chip that controls +1.2V, +2.5V, +3.3V and +5V internal operating voltages and it generates system reset if undervoltage situation occurs in monitored voltages. There is a separate watchdog supervisor chip that generates system reset if watchdog refresh period exceeds 1.6 seconds. F-FM(W)-00 Figure 4 CPU board...