Page 191
Frame for FM 49 Document no. M1187335-004 3.7 Adjustments and calibrations It is not necessary to perform calibrations or adjustments on S/5 Frame for FM.
Page 192
Datex-Ohmeda S/5 FM 50 Document no. M1187335-004 4Troubleshooting 4.1 S/5 Frame for FM Table 6 S/5 Frame for FM troubleshooting chart ProblemCause What to do Monitor is not starting. 1. The batteries are empty. 2. Fuses may be blown. 3. If power cord connected, AC/DC unit may be faulty. 4. If AC/DC unit is working, the DC/ DC board may be faulty. 5. On/Stby key may be faulty. 1. Connect the power cord. 2. Replace fuses. 3. Replace the AC/DC power unit. 4. Replace the DC/DC unit. 5. Replace the command...
Page 193
Frame for FM 51 Document no. M1187335-004 4.2 Memory option Table 3 Memory option troubleshooting chart ‘Battery failure’ message on the screen. 1. Problem in communication between battery and DC/DC board. 2. Battery too old or defected. 3. DC/DC board may be faulty. 1. Replace battery. 2. Replace battery. 3. Replace the DC/DC board. Frame temperature high message. The temperature inside the frame is too high. Check monitor ventilation holes. ‘Battery temperature high’ message on the screen....
Page 194
Datex-Ohmeda S/5 FM 52 Document no. M1187335-004 4.3 NET section Table 4 NET section troubleshooting chart ProblemCause What to do Monitor does not connect to the network. Monitor connects to the network, but disconnects unexpectedly (‘Network connection down’ message on the monitor screen). Patch panel Patch cable not connected to HUB or to panel. Patch cable Patch cable or connector defective. HUB not connected to power supply. HUB port closed due to physical layer problems. HUB port temporarily...
Page 195
Frame for FM 53 Document no. M1187335-004 ‘Check network connectors’ message shows on the monitor screen Monitor-Network cable Cable not pro perly connected to the wallplate or to the monitor. Cable or connector defective. Network cable (inside the monitor) defective Replace the Network cable. Network Connector Board defective Replace the Network Connector Board Unit. Multi I/O adapter faulty Replace multi I/O adapter. Identification plug There is no identification plug properly attached to the...
Page 197
Appendix A Service check form, Datex-Ohmeda S/5 Frame for FM, F-FM 1(4) Document no. M1187335-004 APPENDIX A: Service check form, Datex-Ohmeda S/5 Frame for FM, F-FM OK = Test OK N.A. = Test not applicable Fail = Test failed Customer Service Module type S/N Service engineer Date Monitor Installation L- E- F-N- N-N-N-N- N- General OK N.A. Fail OK N.A. Fail 1. Internal parts...
Page 198
Datex-Ohmeda S/5 FM 2(4) Document no. M1187335-004 Recorder unit OK N.A. Fail 22. Recorder messages 23. Recording 24. Paper speed 25. Quality of recording Notes MemCard Option S/N OK N.A. Fail OK N.A. Fail 26. Module recognition 27. Memories and PCMCIA controller...
Page 199
Appendix A Service check form, Datex-Ohmeda S/5 Frame for FM, F-FM 3(4) Document no. M1187335-004 Wireless LAN option S/N OK N.A. Fail OK N.A. Fail 38. Signal strength 39. Configure access point 40. Connection to Network Notes General OK N.A. Fail OK N.A. Fail 41....