Home > GE > Medical Monitor > GE S5 Owners Manual

GE S5 Owners Manual

Here you can view all the pages of manual GE S5 Owners Manual. The GE manuals for Medical Monitor are available online for free. You can easily download all the documents as PDF.

Page 191

Frame for FM 49
Document no. M1187335-004
3.7 Adjustments and calibrations
It is not necessary to perform calibrations or adjustments on S/5 Frame for FM. 

Page 192

Datex-Ohmeda S/5 FM
50
Document no. M1187335-004
4Troubleshooting
4.1 S/5 Frame for FM
Table 6 S/5 Frame for FM troubleshooting chart
ProblemCause What to do
Monitor is not starting. 1. The batteries are empty. 2. Fuses may be blown.
3. If power cord connected, AC/DC unit may be faulty.
4. If AC/DC unit is working, the DC/ DC board may be faulty.
5. On/Stby key may be faulty. 1. Connect the power cord.
2. Replace fuses.
3. Replace the AC/DC power unit.
4. Replace the DC/DC unit.
5. Replace the command...

Page 193

Frame for FM 51
Document no. M1187335-004
4.2 Memory option
Table 3 Memory option troubleshooting chart
‘Battery failure’ message on the 
screen. 1. Problem in communication 
between battery and DC/DC 
board.
2. Battery too old or defected.
3. DC/DC board may be faulty. 1. Replace battery.
2. Replace battery.
3. Replace the DC/DC board. 
Frame temperature high 
message. The temperature inside the frame is 
too high. Check monitor ventilation holes.
‘Battery temperature high’ 
message on the screen....

Page 194

Datex-Ohmeda S/5 FM
52
Document no. M1187335-004
4.3 NET section
Table 4 NET section troubleshooting chart
ProblemCause What to do
Monitor does not connect to 
the network.
Monitor connects to the 
network, but disconnects 
unexpectedly (‘Network 
connection down’ message 
on the monitor screen). Patch panel 
Patch cable not connected to HUB or to panel.
Patch cable  Patch cable or connector defective.
HUB not connected to power supply.
HUB port closed due to physical layer problems.
HUB port temporarily...

Page 195

Frame for FM 53
Document no. M1187335-004
‘Check network connectors’ 
message shows on the 
monitor screen Monitor-Network cable Cable not pro
perly connected to the wallplate 
or to the monitor.
Cable or connector defective.
Network cable (inside the 
monitor) defective Replace the Network cable.
Network Connector Board 
defective Replace the Network Connector Board Unit.
Multi I/O adapter faulty Replace multi I/O adapter.
Identification plug There is no identification plug properly attached  to the...

Page 196

Datex-Ohmeda S/5 FM
54
Document no. M1187335-004
For your notes: 

Page 197

Appendix A Service check form, Datex-Ohmeda S/5 Frame for FM, F-FM
1(4)
                           Document no. M1187335-004
APPENDIX A: Service check form, 
Datex-Ohmeda S/5 Frame for FM, F-FM 
OK = Test OK            N.A. = Test not applicable            Fail = Test failed
Customer
 
Service Module type
S/N
Service engineer
 Date
 Monitor Installation
 
L- E- F-N-
N-N-N-N-
N-
 General
     OK           N.A.          Fail      OK           N.A.          Fail
1. Internal parts...

Page 198

Datex-Ohmeda S/5 FM
2(4)
Document no. M1187335-004
Recorder unit
     OK           N.A.          Fail
22. Recorder messages                            23. Recording
                           
24. Paper speed                            25. Quality of recording
                           
Notes
 
 
MemCard Option S/N
       OK           N.A.          Fail      OK           N.A.          Fail
26. Module recognition                            27. Memories and PCMCIA 
controller...

Page 199

Appendix A Service check form, Datex-Ohmeda S/5 Frame for FM, F-FM
3(4)
                           Document no. M1187335-004
Wireless LAN option  S/N
      OK           N.A.          Fail      OK           N.A.          Fail
38. Signal strength                            39. Configure access point
                           
40. Connection to 
Network                                                        
Notes
 
 
General 
 
      OK           N.A.          Fail      OK           N.A.          Fail
41....

Page 200

Datex-Ohmeda S/5 FM
4(4)
Document no. M1187335-004
For your notes: 
Start reading GE S5 Owners Manual