Page 181
Frame for FM 39 Document no. M1187335-004 To change the LCD backlight - Disconnect the connector from the inverter board. - Push the lid beside the backlight to release the backlight tube. Pull the backlight carefully out. NOTE: Do not bend the backlight tube. NOTE: Do not touch the backlight tube with your hands. To separate the LCD display from the frame unit - Remove the four (T10) screws holding the display to the frame unit. - Disconnect the connector from the inverter board. - Lift the display...
Page 182
Datex-Ohmeda S/5 FM 40 Document no. M1187335-004 3.3.2 Handling and storage of LCD display component Handling of LCD display comp onent and protective window If the LCD Display component surface becomes dusty, wipe it gently with absorbent cotton, chamois or other soft material. If necessary, breathe onto the display surface and wipe immediately. The display surface may also be cleaned using a small amount of normal hexane. Do not use acetone, toluene or alcohol because they cause chemical damage to...
Page 183
Frame for FM 41 Document no. M1187335-004 To remove the inverter board NOTE: The inverter board is sensitive to bending. - Remove the two (T10) screws and the insulator plate. - Disconnect the backlight connector(s) and the inverter cable connector. NOTE: When reassembling make sure that the connectors are fastened properly. Inverter Board F-FM(W)-01 Inverter Board F-FM(W)-00 NOTE: When replacing the Inverter board, fasten the inverter screws gently, using only as much force as necessary to fasten...
Page 184
Datex-Ohmeda S/5 FM 42 Document no. M1187335-004 To remove the EMC shield - Remove the 14 screws (T10) holding the shield to the frame unit. One of the screws is under the Network connection cable. Lift the shield carefully up. NOTE: When reassembling the monitor, remember to connect the grounding wire of the Sync connector unit to the EMC shield. Now you can access the following field replaceable spare parts. Follow the instructions below: - “To remove the handle guide unit” - “To remove the CPU...
Page 185
Frame for FM 43 Document no. M1187335-004 To remove the CPU board - Disconnect the following cables from the CPU board: • loudspeaker cable • Sync connector cable • interconnection board connector cable • network connection cable - Remove the eight screws (T10) holding the CPU board to the frame. - Lift the CPU board carefully up and remove it from the frame. To remove the loudspeaker unit - Remove the two screws (T10) holding the loudspeaker unit to the frame. - Disconnect the loudspeaker...
Page 186
Datex-Ohmeda S/5 FM 44 Document no. M1187335-004 To remove the interconnection board - Remove the three screws (T10) holding the plastic flex board cover. - Turn the frame around. Remove the two screws (T10) holding the flex board to the frame. - Disconnect the interconnection board connector from the module bus cable. NOTE: When reassembling the monitor, check that a cellular rubber sealing is attached to the module bus cable connector. - Remove the interconnection board carefully from the frame.
Page 187
Frame for FM 45 Document no. M1187335-004 To remove the power supply frame - Remove the two screws (T15) holding the power supply frame to the inner frame. - Now you have access to replace the inner frame and Modulebus cable unit sub- assembly.
Page 188
Datex-Ohmeda S/5 FM 46 Document no. M1187335-004 3.4 Batteries 3.4.1 Battery indicators The S/5 FM messages, screen symbols and front panel LED indicators tell the user about the status of the batteries. For screen symbols, see “Part 1 Symbols”. For LED indicators, consult the table below and for messages, see section “Troubleshooting.” Table 5 Battery indicators Screen symbol Explanation Front panel battery LED indicators Monitor is battery powered. Batteries are fully charged and the size of the...
Page 189
Frame for FM 47 Document no. M1187335-004 3.4.2 To change the batteries Figure 14 Inserting a battery (1) Open the lid of the battery compartment located behind the handle by sliding it to the left. (2) Lift up the battery you want to change. Check the indicators and messages on the screen to make sure that you change the battery with lower charge. (3) Push in the new battery. Make sure that the charging indicator is facing forward and push the battery down all the way. Check the monitor...
Page 190
Datex-Ohmeda S/5 FM 48 Document no. M1187335-004 3.4.4 Conditioning the batteries Condition the batteries regularly to maintain their useful life. Condition a battery every six months or when the message ‘Condition Battery x’‘ appears on the screen. Always observe the messages and symbols on the screen to see the battery status. You can also check the status through Monitor Setup - Battery Setup . Conditioning a battery is best done on an external charger. If you do not have an external charger,...