Patient Side Module, E-PSM, E-PSMP (Rev. 01) 25 Document no. M1215098-002 3 Service procedures 3.1 General service information The field service of the hemodynamic modules is limited to replacing faulty printed circuit boards or mechanical parts. The circuit boards sh ould be returned to GE Healthcare for repair. GE Healthcare is always available for service ad vice. Please provide the unit serial number, full type designation, and a detailed fault description. WARNING Only trained personnel with...
Page 232
E-modules 26 Document no. M1215098-002 3.2.2 Recommended parts Table 14 Recommended parts 3.2.3 Visual inspection Detach the module cover by removing the four screws from the bottom of the module. Check: 1. Internal parts−screws are tightened properly − connectors are connected properly − NIBP tubing is attached properly − there are no loose objects inside the module 2. External parts −the front cover and the front panel sticker are intact Multi-Link ECG accessories, IEC: - Multi-link 3-leadwire set...
Page 233
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 27 Document no. M1215098-002 −all connectors are intact and attached properly − the module box and latch are intact 3.2.4 Functional inspection 3. NIBP pump filter Replace the NIBP pump filter, if necessary. Reattach the module cover and check that the latch is moving properly. Switch the monitor on and wait until the monitoring screen appears. Configure the monitor screen so that all the needed para meters are shown, for example as follows: Monitor Setup...
Page 234
E-modules 28 Document no. M1215098-002 ECG measurement 6. Module software (serial numbers) Enter the service menu: Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8) Take down the information regarding the module software by selecting Scroll Vers and turning the ComWheel. 7. Communication and memories Enter the Parameters - ECG service menu. Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing faster than by 5 per second. Check also...
Page 235
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 29 Document no. M1215098-002 11. Test with the patient simulatorConnect the leads to a patient simulator. Perform the settings and checks with Dynatech Nevada MedSim 300 Patient Simulator: ECG - BASE - BPM - 160 PACE - WAVE - NSR Check that a normal ECG waveform is shown, the HR value is 160 (±5) and the ‘Pacer count’ value is not increas ing in the service menu. ECG - PACE - WAVE - ASNC Check that pacemaker spikes are shown on the ECG waveform,...
Page 236
E-modules 30 Document no. M1215098-002 Temperature measurement 14. Communication and memoriesEnter the ESTP: STP service menu: Parameters - ESTP : STP Check that the Time-outs, Bad checksums and Bad c-s by mod values do not increase faster than by 5 per second. Check also that the STP board memories have passed the internal memory test, i.e. the RAM, ROM and EEPROM show all OK. 15. Temperature probe detection Check that the ‘Cable’ and ‘Probe’ show OFF for both channels, T1 and T2, when no probes...
Page 237
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 31 Document no. M1215098-002 Invasive blood pressure measurement 19. Membrane keysCheck the front panel membrane keys that are related to the InvBP measurement. Press each of the keys for at least one second. Check that the pressed key is identified, i.e. one of the texts for Buttons changes fr om OFF to ON in the service menu. 20. Cable and transducer detection Check that the Cable and Probe for P1 show OFF. Connect the InvBP adapter cable to the module,...
Page 238
E-modules 32 Document no. M1215098-002 24. Test measurementConnect the SpO 2 probe onto your finger. Check that the reading of 95-99 and SpO2 waveform appears. Check that the HR value is calculated from SpO 2 when ECG and InvBP (P1) are not measured. Non Invasive Blood Pressure measurement 25. Communication and memories Enter the NIBP module service menu: Parameters - NIBP Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing faster than by 5 per second. Check also that...
Page 239
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 33 Document no. M1215098-002 Wait for 15 seconds for the pressure to stabi lize then check that the pressure does not drop more than 6 mmHg per one minute. Release the pressure by pressing the ComWheel once more. 29. Calibration check Recalibrate the NIBP measurement accordin g to the instructions in section “NIBP calibrations” on page 44. Remember to set the calibration protection back on after the calibration. Disconnect the pressure manometer. Select...
Page 240
E-modules 34 Document no. M1215098-002 32. Test measurementCheck that the automatic inflation limits are in use: NIBP - NIBP Setup - Inflation Limits - Auto - Previous Menu Connect the cuff onto your arm, select Start Ven.Stasis in the NIBP menu and press the ComWheel. Check that the module identifies the cuff, i.e. the text Adult appears in the NIBP digit field for a short moment. Keep the pressure inside the cuff for about half a minute in order to find out that the cuff is not leaking, then...