S/5 Extension Module for FM, N-FC, N-FCREC, N-FREC 1 Document no. M1187338-003 Introduction This Technical Reference Manual Slot provides information for the maintenance and service of the Datex-Ohmeda S/5 Extension Module for FM, N-FC, N-FCREC and N-FREC. The Extension Modules for FM are designed for use with the S/5 FM Monitor and FM Light. Later in this manual the module can be called w/o the system name Datex-Ohmeda S/5. Please also refer to the Part I sections of the manual for information...
Page 272
Datex-Ohmeda S/5 FM 2 Document no. M1187338-003 1Specifications 1.1 General specifications N-FCREC Module size, W × D × H90 × 169 × 191 mm, 3.54 × 6.65 × 7.51 in Module weight 1.0 kg/2.2 lb. Power consumption 2.7 W standby, 11.5 W printing N-FREC Module size, W × D × H90 × 169 × 191 mm, 3.54 × 6.65 × 7.51 in Module weight 0.8 kg/1.7 lb. Power consumption 1.2 W standby, 10 W printing N-FC Module size, W × D × H46 × 169 × 191 mm, 1.81 × 6.65 × 7.51 in Module weight 0.5 kg/1.1 lb. Power...
Page 273
S/5 Extension Module for FM, N-FC, N-FCREC, N-FREC 3 Document no. M1187338-003 1.2.2 Technical specifications Airway humidity 0...100%, condensing Sampling rate 150 ±25ml/min (sampling line 2 to 3 m, normal conditions) Sampling delay 2.1 seconds typical with a 3-m sampling line Total system response time 2.4 seconds typical with a 3-m sampling line, including sampling delay and rise time (typically 3.7 seconds with a 6-m sampling line). Automatic compensation fo r barometric pressure, CO 2-NO2 and...
Page 274
Datex-Ohmeda S/5 FM 4 Document no. M1187338-003 1.2.4 Conditions exceeding normal Accuracy specifications under the following conditions nop: n Ambient temperature 10...40 °C, within ±5 °C of calibration Ambient pressure 500...800 mmHg, ±50 mmHg of calibration Ambient humidity 10...98% RH, ±20% RH of calibration o During warm-up 1 to 10 minutes, under normal conditions p During warm-up 10 to 30 minutes, under normal conditions 1.3 Respiration Rate (RR) Measurement range 4 to 80 breaths /min...
Page 275
S/5 Extension Module for FM, N-FC, N-FCREC, N-FREC 5 Document no. M1187338-003 2 Functional description 2.1 Measurement principle 2.1.1 CO2 measurement MiniC is a side stream gas analyzer, measuring real time concentrations of CO2. It is a non dispersive infrared analyzer, measuring absorption of the gas sample using an optical narrow band filter. The infrared radiation detector is thermopile. Concentration of CO 2 is calculated from absorption measured at 4.2 to 4.3 µm. Figure 2 Absorbance of CO2...
Page 276
Datex-Ohmeda S/5 FM 6 Document no. M1187338-003 2.2.1 Gas sampling system The sampling system draws a gas sample to the analyzer at a fixed rate. The gas sampling system samples the measured air to the module, and removes water and impurities from it. A sampling line is connected to the water trap. The pump draws gas through the sampling line to the gas measuring unit. After the measurement, the gas is exhausted from the sample gas out connector. The sample flow is nominally 150 ml/min. Figure 3...
Page 277
S/5 Extension Module for FM, N-FC, N-FCREC, N-FREC 7 Document no. M1187338-003 NafionTM tube 1) A Nafion tube is used between the water trap and the zero valve to balance the sample gas humidity with that of ambient air. The tube prev ents errors caused by the effect of water vapor on gas partial pressure when humid gases are measured after calibration with dry gases. Gas analyzers After the zero valve and Nafion tube, the gas passes through the miniC measuring unit. Sample flow differential pressure...
Page 278
Datex-Ohmeda S/5 FM 8 Document no. M1187338-003 Figure 4 Gas tubing layout 2.2.2 MiniC measuring unit The miniC measuring unit is a non-dispersive infrared analyzer measuring absorption of the gas sample at 4.2 to 4.3 µm infrared wavelength, which is selected using an optical narrow band filter. The IR lamp is a filament surrounded by thermal isolation. There is a hole in the isolation, passing the radiation to a conical measuring chamber with 3 mm length. From the sample chamber, the radiation...
Page 279
S/5 Extension Module for FM, N-FC, N-FCREC, N-FREC 9 Document no. M1187338-003 2.2.3 CPU board The CPU board contains a processor, memories and all the analog signal processing needed. A MiniC measuring unit is attached to the board with a flexible PCB. Also supply voltage and an RS485 serial channel are connected to the CPU board using another flexible PCB. Analog signals (CO 2, temperature, absolute and differential pressures and lamp current signals) are fed to the 16-bit A/D converter. The proces...
Page 280
Datex-Ohmeda S/5 FM 10 Document no. M1187338-003 Module Bus Flex boardModule Bus Flex board connects the CPU board of the miniC unit and the recorder board to the module bus. Figure 7 Module bus flex board 2.3 Connectors and signals 2.3.1 Module bus connector 5 pin female connector Pin No. Signal 1GND 2Vmod 13.8 - 16 V 3Data + 4Data - 5Shield