Page 281
S/5 Extension Module for FM, N-FC, N-FCREC, N-FREC 11 Document no. M1187338-003 3 Service procedures 3.1 General service information The field service of the Extension Modules for FM is limited to replacing faulty circuit boards or mechanical parts. The circuit boards should be returned to GE Healthcare for repair. GE Healthcare is always available for service ad vice. Please provide the unit serial number, full type designation and a detailed fault description. WARNING Only trained personnel with...
Page 282
Datex-Ohmeda S/5 FM 12 Document no. M1187338-003 The instructions include a check form (“APPENDIX A:”) which should be filled in when performing the procedures. The symbol in the instructions means that the check form should be signed after performing the procedure. 3.2.1 Recommended tools 3.2.2 Recommended parts Detach the module box cover by removing the five screws from the locking side of the module. Be careful with the flex cable, when opening the cover. Remove the grey plastic shield over...
Page 283
S/5 Extension Module for FM, N-FC, N-FCREC, N-FREC 13 Document no. M1187338-003 NOTE: Be careful to limit the application of al cohol to the thermal printhead and the window. Leave the paper compartment empty and close the hatch. 2. Check external parts •all connectors are intact and attached properly •the module box and the locking system are intact Reattach the module and check that the locking system moves properly. CO2 measurement 3. Check Mini D-Fend Detach the Mini D-fend. Che ck the condition...
Page 284
Datex-Ohmeda S/5 FM 14 Document no. M1187338-003 If one of the values is increasing faster, it indicates a failure in module bus communication. 6. Flow measurement offset Enter the service menu Gases: Gas Unit - Gases Check that the flow measurement offset, i.e. the shown sample Zero value is within ±10 ml/min. 7. Ambient pressure Check that the shown Ambient value corresponds with the current ambient pressure (±20 mmHg). 8. Zero valve checkFeed calibration gas and check that the gas readings in...
Page 285
S/5 Extension Module for FM, N-FC, N-FCREC, N-FREC 15 Document no. M1187338-003 11. Check the flow ratesWait until the Sample Flow valu e returns close to 150 ml/min. Connect a flow meter to the 3 meter sampling line and check that the flow (the flow meter reading) is within the following range: Sampling flow (ml/min) 135...165 If necessary, readjust the sampling flow: Select Sample gain adj from the menu. To increase the sampling flow, turn the Co mWheel counterclockwise. To decrease the flow,...
Page 286
Datex-Ohmeda S/5 FM 16 Document no. M1187338-003 Check that the monitor shows the message ‘Apnea’ within 30 seconds after you have stopped feeding the gas. Recorder 17. Paper compartment cleaning Open the paper compartment hatch and take out the paper roll, if installed. Remove any paper chaff from the compartment. Clean the thermal printhead and the small glass window in front of the static brush with a cotton swab dipped in isopropyl alcohol. Avoid contact with the rubber paper roller. NOTE: Be...
Page 287
S/5 Extension Module for FM, N-FC, N-FCREC, N-FREC 17 Document no. M1187338-003 Install a paper roll and close the cover. Check that the message ‘Recorder: Cover open’ disappears from the screen. 20. Quality of recordingPress the Start/ Stop key again and check that the quality of the recordings is acceptable. 21. Recording speed Press the RECORD WAVE key again and this time wait until the recording stops. Check that the length of the recorded waveform scale is 18.7 cm ( ±1.5 cm). Change the...
Page 288
Datex-Ohmeda S/5 FM 18 Document no. M1187338-003 3.3 S/5 N-FCREC disassembly and reassembly CAUTION When reassembling the module, make sure that the tubes and cables are not pinched between the bo ards and the cover. To reassemble the module, reverse the order of the disassembly steps. 3.3.1 Instructions after replacing MiniC unit After replacing the MiniC unit: •perform the sampling system leak test •perform the occlusion test •perform the gas calibration 3.3.2 Before disassembly WARNINGWear a...
Page 289
S/5 Extension Module for FM, N-FC, N-FCREC, N-FREC 19 Document no. M1187338-003 3. Unscrew the two pozidrive screws inside the recorder unit completely (the screws cannot be removed) and remove the recorder unit. 4. Remove the five screws (T8) on the side where the device plate is located. 5. Place the N-F(C)(REC) on a table the device plate side down. 6. Remove the two screws (T8) near the module bus connector. 7. Remove the module cover carefully. 8. Remove the EMC-cover and CO 2 unit insulator...
Page 290
Datex-Ohmeda S/5 FM 20 Document no. M1187338-003 NOTE: The finger of the insulator plate secures the interconnection flex connector. While reassembling the insulator plate ensure that the finger supports the connector correctly. 9. Disconnect the Nafion an d silicon tubes from the mini D-fend connectors. NOTE: While reassembling the nafion tube ensure the correct routing for both modules. If the nafion tube is not correctly routed it may get bent or jammed when reassembling the cover. Bent or...