Page 251
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 45 Document no. M1215098-002 •Remove the module from the frame and plug it back again. Then perform “Calibration check” (see the preceding page) to verify the new calibration. 3.4.2 Temperature calibration NOTE: For the temperature calibration, separate, accurate test plugs (25 °C and 45 °C) are needed. A test set of two plugs is available from GE Healthcare, order code 884515-HEL. A Dual temperature adapter cable, order code 2016998-001 is also required...
Page 252
E-modules 46 Document no. M1215098-002 4Troubleshooting 4.1 Troubleshooting charts See also the “User’s Reference Manual” for more troubleshooting procedures. 4.1.1 NIBP ProblemCause What to do No NIBP value displayed NIBP not selected on screen. Check monitor setup. NIBP menu fading No E-PSM(P) module, module not properly connected. Plug in the module. ‘Artifacts’ message Unsuccessful measurement due to patient movement, shivering, external artifact or weak signal. Check the patient status. ‘Weak...
Page 253
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 47 Document no. M1215098-002 Cuff ID not working 1. Defective cuff ID holes in the NIBP cuff hose - Replace NIBP cuff hose. 2. NIBP flex board connector wrongly connected - Check that the NIBP flex board connector is properly connected to the STP input board: all pins have to be connected. 3. Cuff ID switches defective - To check the switches, attach a NIBP cuff hose without the cuff ID and check that the message ‘Select inflation limit’ appears. If...
Page 254
E-modules 48 Document no. M1215098-002 ‘Cuff occlusion’ message 1. Cuff and/or hose occluded. Reason: - cuff tube kinked - Straighten tube. - tubes inside the shield kinked - Straighten tubes. - tubes inside module kinked - Straighten tubes. - occlusion inside/outside module - Remove occlusion. 2. Cuff, hose, and tubes OK. Reason: - fault in pressure transducer - Replace the NIBP board. - fault in A/D converter - Replace the NIBP board. - faulty calibration - Check calibration. ‘Calibration switch on’...
Page 255
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 49 Document no. M1215098-002 4.1.2 NIBP error code explanation Code Problem What to do 0 RAM failure; memory failure Change the NIBP board. 1 ROM checksum error; memory failure Change the NIBP board. 2 Pump current failure Check short circuits. Change the NIBP board. 3 Safety CPU internal test failure or pressure sensor reference voltage failure Change the NIBP board. 4 EEPROM protection error Press module buttons Auto ON/OFF and Start/ Cancel...
Page 256
E-modules 50 Document no. M1215098-002 4.1.3 Pulse oximetry (SpO2) ProblemCause What to do Message ‘NO PROBE’ No sensor connected to the module SpO2 connector. Check sensor connections. Sensor faulty. Change the sensor. Flat cable connecting the SpO2 connector to the STP board loosen or broken. Check the Flat cable, replace if necessary. Message ‘PROBE OFF’ though sensor properly attached to the patient Unsuitable site. Try another site. Sensor faulty. Try another sensor. Sensor connection cable...
Page 257
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 51 Document no. M1215098-002 4.1.4 Invasive blood pressure ProblemCause What to do Abnormally low pressure Transducer wrongly positioned. Check mid-heart level and reposition transducer. No pressure Defective transducer. Check transducer. No pressure module plugged in. Check the module. No waveform selected on screen. Check the selected pressure waveforms by pressing Monitor Setup key and selecting Screen 1 Setup - Waveform Fields . Check that the...
Page 258
E-modules 52 Document no. M1215098-002 4.2 Troubleshooting flowcharts 4.2.1 Troubleshooting for NIBP parameter Figure 13 Troubleshooting flowchart for NIBP parameter Insert new module and turn power on. Does fault still appear? Module ID on screen?Does another module work in same slot? Error message on screen? See error code explanation inservice manual and fix it. Do they work? Check tubes and connectors. Find leak and fix it.Leak test in Service Menu OK? Pump check in Service Menu OK? Key check in...
Page 259
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 53 Document no. M1215098-002 4.2.2 Troubleshooting for ESP parameters Figure 14 Troubleshooting flowchart for ESP Parameters Is it correct?Select the correctconfiguration YES NO Does another module work in same slot? ECG module ID on screen? Remove ECG boardSTP module ID on screen? Replace STP board Replace ECG board Check front panel key functions in Service menu Do they wo rk? Che ck front panel unit, ECG board, and STP board to find culprit. Replace if...
Page 260
E-modules 54 Document no. M1215098-002 5 Earlier revisions Patient Side Modules E-PSM, E-PSMP (Rev. 00)