Patient Side Module, E-PSM, E-PSMP (Rev. 01) 45
Document no. M1215098-002
•Remove the module from the frame and plug it back again. Then perform  “Calibration 
check”  (see the preceding page) to verify the new calibration.
3.4.2 Temperature calibration
NOTE: For the temperature calibration, separate, accurate test plugs (25 °C and 45 °C) are 
needed. A test set of two plugs is available from GE Healthcare, order code 884515-HEL. 
A Dual temperature adapter cable, order code 2016998-001 is also required...

Page 252

E-modules
46
Document no. M1215098-002
4Troubleshooting
4.1 Troubleshooting charts
See also the “User’s Reference Manual” for more troubleshooting procedures.
4.1.1 NIBP
ProblemCause What to do
No NIBP value displayed NIBP not selected on screen. Check monitor setup.
NIBP menu fading No E-PSM(P) module, module not 
properly connected. Plug in the module.
‘Artifacts’ message Unsuccessful measurement due to  patient movement, shivering, 
external artifact or weak signal. Check the patient status.
‘Weak...

Page 253

Patient Side Module, E-PSM, E-PSMP (Rev. 01) 47
Document no. M1215098-002
Cuff ID not working 1. Defective cuff ID holes in the 
NIBP cuff hose - Replace NIBP cuff hose.
2. NIBP flex board connector  wrongly connected - Check that the NIBP flex board 
connector is properly connected to 
the STP input board: all pins have to 
be connected.
3. Cuff ID switches defective - To  check the switches, attach a NIBP 
cuff hose without the cuff ID and 
check that the message ‘Select 
inflation limit’ appears. If...

Page 254

E-modules
48
Document no. M1215098-002
‘Cuff occlusion’ message 1. Cuff and/or hose occluded. Reason:
- cuff tube kinked - Straighten tube.
- tubes inside the shield kinked - Straighten tubes.
- tubes inside module kinked - Straighten tubes.
- occlusion inside/outside module - Remove occlusion.
2.  Cuff, hose, and tubes OK. Reason:
- fault in pressure transducer - Replace the NIBP board.
- fault in A/D converter - Replace the NIBP board.
- faulty calibration - Check calibration.
‘Calibration switch on’...

Page 255

Patient Side Module, E-PSM, E-PSMP (Rev. 01) 49
Document no. M1215098-002
4.1.2 NIBP error code explanation
Code Problem What to do
0 RAM failure; memory failure Change the NIBP board.
1 ROM checksum error; memory failure Change the NIBP board.
2 Pump current failure Check short circuits. Change the NIBP board.
3 Safety CPU internal test failure or pressure  sensor reference voltage failure Change the NIBP board.
4 EEPROM protection error Press module buttons 
Auto ON/OFF and Start/ 
Cancel...

Page 256

E-modules
50
Document no. M1215098-002
4.1.3 Pulse oximetry (SpO2) 
ProblemCause What to do
Message ‘NO PROBE’ No sensor connected to the module  SpO2 connector. Check sensor connections.
Sensor faulty. Change the sensor.
Flat cable connecting the SpO2 
connector to the STP board loosen or 
broken. Check the Flat cable, replace if 
necessary.
Message ‘PROBE OFF’ 
though sensor properly 
attached to the patient Unsuitable site.
Try another site.
Sensor faulty. Try another sensor.
Sensor connection cable...

Page 257

Patient Side Module, E-PSM, E-PSMP (Rev. 01) 51
Document no. M1215098-002
4.1.4 Invasive blood pressure
ProblemCause What to do
Abnormally low pressure Transducer wrongly positioned. Check mid-heart level and reposition  transducer.
No pressure Defective transducer. Check transducer.
No pressure module plugged in. Check the module.
No waveform selected on screen. Check the selected pressure  waveforms by pressing 
Monitor 
Setup
 key and selecting  Screen 1 
Setup -  Waveform Fields .
Check that the...

Page 258

E-modules
52
Document no. M1215098-002
4.2 Troubleshooting flowcharts
4.2.1 Troubleshooting for NIBP parameter
 
Figure 13 Troubleshooting flowchart for NIBP parameter
Insert new module and turn
power on.
Does fault still appear?
Module ID on screen?Does another
module work in
same slot?
Error message on screen?
See error code explanation inservice manual and fix it.
Do they work?
Check tubes and connectors.
Find leak and fix it.Leak test in Service Menu OK?
Pump check in
Service Menu OK?
Key check in...

Page 259

Patient Side Module, E-PSM, E-PSMP (Rev. 01) 53
Document no. M1215098-002
4.2.2 Troubleshooting for ESP parameters
 
Figure 14 Troubleshooting flowchart for ESP Parameters
Is it correct?Select the correctconfiguration
YES NO
Does another
module work in same slot?
ECG module ID
on screen?
Remove
ECG boardSTP module ID on screen?
Replace STP
board
Replace ECG board
Check front panel
key functions in
Service menu
Do
they wo rk?
Che ck front panel unit,
ECG board, and STP board
to find culprit. Replace if...

Page 260

E-modules
54
Document no. M1215098-002
5 Earlier revisions
Patient Side Modules E-PSM, E-PSMP (Rev. 00) 