Patient Side Module, E-PSM, E-PSMP (Rev. 01)
Technical Reference Manual Slot
Conformity according to the Council Directive 93/42/EEC concerning Medical Devices amended by 2007/47/EC
CAUTION: U.S. Federal law restricts this device to sale by or on the order of a li censed medical practitioner.
Outside the USA, check local laws  for any restriction that may apply.
All specifications subject to change without notice.
Document number M1215098-002
12 August, 2011
GE Healthcare Finland Oy
Kuortaneenkatu 2...

Page 202


				            

Page 203

Table of contentsi
Document no. M1215098-002
Table of contents
Table of contents i
Table of figures iii
List of tables iv
Introduction1
1 Specif ications 3
1.1 General specif ications  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  3
1.2 Typical performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  . . . . . . . . . . . . . .  3
1.2.1 NIBP. . . . . . . . . . . ....

Page 204

E-modules
ii
Document no. M1215098-002
3 Service procedures25
3.1 General service information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  . . . . . . . . . . 25
3.2 Service check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  . . . . . . . . . . . . . . . . 25
3.2.1 Recommended tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ....

Page 205

 Table of figuresiii
Document no. M1215098-002
Table of f igures
Figure 1 Patient Side Module, E-PSMP ............................................................................................ ...................................................... 1
Figure 2 Absorption of infrared light in  the finger ............................................................................. .............................................. 9
Figure 3 Pulse oximetry probe parts layout and sc hematic diagram...

Page 206

E-modules
iv
Document no. M1215098-002
List of tables
Table 1 Patient Side Module options................................................................................................................................. 1
Table 2 Module bus connector description ........................................................................................ ........................ 22
Table 3 ECG connector ..............................................................................................................

Page 207

Patient Side Module, E-PSM, E-PSMP (Rev. 01) 1
Document no. M1215098-002
Introduction
This Technical Reference Manual Slot provides  information for the maintenance and service of 
the Patient Side Modules E-PSMP and E-PSM. 
Please also refer to “Technical Reference Ma nual” of the monitor for system specific 
information e.g. related documentation, conv entions used, symbols on equipment, safety 
precautions, system description, system installation, interfacing, functional check and planned...

Page 208

E-modules
2
Document no. M1215098-002
Impedance respiration measurement is indicated for patients aged 3 and up. The NIBP 
measurement is indicated for patients who weigh 5 kg (11 lb.) and up. This device is indicated 
for use by qualified medical personnel only. 
Monitor software compatibility
Patient Side Module, E-PSM(P) Rev. 00 and 01 are designed for use with Datex-Ohmeda 
monitors as follows: 
−S/5 FM monitors using software L-FICU04(A) or later.
− S/5 Anesthesia Monitors using software L- ANE04(A)...

Page 209

Patient Side Module, E-PSM, E-PSMP (Rev. 01) 3
Document no. M1215098-002
1Specifications
1.1 General specifications
Module size51 x 132 (171 w/ tab) x 140 mm
W x D x H 2 x 5.2 (67 w/ tab) x 5.5 in
Module weight 0.6 kg /1.4 lb.
Power consumption 2.3 W typical (NIBP pump off)
7.5 W typical (NIBP pump on)
Operation temperature 10 to 40°C / 50 to 104 °F
1.2 Typical performance
1.2.1 NIBP
WARNING Non-invasive blood pres sure measurement is intended for patients 
weighing over 5 kg (11 lb.) 
Oscillometric...

Page 210

E-modules
4
Document no. M1215098-002
1.2.2 ECG
Lead selectionI, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Sweep speeds  12.5, 25, 50 mm/sec.
Display filter
Diagnostic0.05 to 150 Hz
Monitoring 0.5 to 30 Hz (-3 dB, with 50 Hz reject filter)
0.5 to 40 Hz (-3 dB, with 60 Hz reject filter)
ST filter 0.05 to 30 Hz (-3 dB, with 50 Hz reject filter)
0.05 to 40 Hz (-3 dB, with 60 Hz reject filter)
Heart rate from ECG
Range 30 to 250 bpm
Accuracy ±5 bpm or ±5%, whichever is greater
Resolution 1 bpm
Update...