Page 161

Frame for FM 19
Document no. M1187335-004
DIS connector, X6 DIS interface (RS485) 
2.2.2 Digital and analog outputs Digital outputs
There are separate digital output s. Both signals use TTL-level. The outputs are: Defibrillation 
Sync. and Nurse Call.
Def ibrillation Sync  (X5, pin1)  indication is generated by ECG. When active, the signal is in state 
1. After 10 ms the signal is reset to state 0. Ne w Defibrillation Sync is not generated before the 
indication is deactivated. The delay from the R wave...

Page 162

Datex-Ohmeda S/5 FM
20
Document no. M1187335-004
For your notes: 

Page 163

Frame for FM 21
Document no. M1187335-004
3 Service procedures
3.1 General service information
The field service of S/5 FM is limited to replacin g faulty PC boards and mechanical parts. Return 
the faulty PC boards to GE Healthcare for repair.
GE Healthcare is always available for service  advice. Provide the unit serial number, full type 
designation and a detailed fault description. 
WARNING Only trained personnel with appropri ate tools and equipment should 
perform the tests and repair s outlined in...

Page 164

Datex-Ohmeda S/5 FM
22
Document no. M1187335-004
Table 2 Patient simulators’ compatibility with each hemodynamic module 
Table 3 Adapter cables for hemodynamic patient simulators
3.2.3 Recommended parts
Table 4 Recommended parts
3.2.4 Before beginning •
Make sure that the monitor is switched to standby.
•If the monitor is connected to Datex-Ohm eda Network, disconnect the Mon-Net cable 
from the monitor. Take out th e memory card, if installed.
•Disconnect the power cord and detach the batteries.
•Detach...

Page 165

Frame for FM 23
Document no. M1187335-004
3.2.5 General
1. Internal partsCheck that:
the screws are tightened properly
 the cables are connected properly
 the connector pins are intact
 there are no loose objects inside the monitor
Reassemble the monitor.

2. External parts Check that:
 all stickers are intact
  the LCD display screen is intact
  the ComWheel cover is intact and is attached properly
  all mechanical parts are functioning properly

3. Module installation
Check that the modules...

Page 166

Datex-Ohmeda S/5 FM
24
Document no. M1187335-004
6. Module recognitionConfigure the monitor screen so that a ll needed parameter information is shown:
Monitor Setup
Screen 1 Setup
Waveform Fields
Digit Fields
Check that the selected parameter information is shown on the screen.
NOTE: InvBP waveforms are not shown without a patient simulator.
Do no connect the simulator to the monitor at this stage.

7. LCD display picture
Check that the picture on the LCD display screen is straight, clear and stable.
...

Page 167

Frame for FM 25
Document no. M1187335-004
11. Monitor softwareEnter the service menu:
Monitor Setup
Install Service  (Password 16-4-34)
Service  (Password 26-23-8)
Take down the information regarding Monitor software.

12. Content of the service log Select  Service Log from the menu. 
Check the contents of the Service Log for possible problems. If the monitor contains the 
recorder unit, record the Service Log onto the recorder by selecting  Record Log.

13. Voltages
Check the power supply output...

Page 168

Datex-Ohmeda S/5 FM
26
Document no. M1187335-004
Delay --> Off
Paper Speed --> 6.25 mm/s
Length --> 30 s
15. Alarm LEDs Enter the  Keyboard  service menu:
Monitor Setup
Install Service  (Password 16-4-34)
Service  (Password 26-23-8)
Keyboard
Select the text  Upper Led. Check that the red alarm LED turns on and off when pressing 
the ComWheel. Check also the yellow alarm LED by selecting  Lower Led from the menu.

16. Membrane keys
Press the keys on the monitor command board one by one. Check that each...

Page 169

Frame for FM 27
Document no. M1187335-004
19. BatteriesEnter the  BATTERY service menu:
Monitor Setup
Install Service  (Password 16-4-34)
Service  (Password 26-23-8)
Frame
Power Supply
WPM Battery
Smart Batt1
Smart Batt2
Check that all values are within reasonable limits.
NOTE: In order to update values for a desired battery, disconnect the other battery 
shortly. 

20. Battery operation Disconnect the power cord (without switching the monitor to standby).
Check that the monitoring continues norm ally....

Page 170

Datex-Ohmeda S/5 FM
28
Document no. M1187335-004
NOTE: InvBP waveforms are not recorded without a patient simulator.
NOTE: If no recording appears, check that the paper roll is installed in a correct way. Only 
one side of the paper is printable.

24. Paper speed
Press the key 
Start/Stop again and this time wait until the recording stops. Check that 
the length of the recorded waveform scale is 18.7 cm (  1.5 cm).
Change the paper speed setting to 1 mm/s:
Print/Record
Record Waveforms
Paper Speed -->...