Page 161
Frame for FM 19 Document no. M1187335-004 DIS connector, X6 DIS interface (RS485) 2.2.2 Digital and analog outputs Digital outputs There are separate digital output s. Both signals use TTL-level. The outputs are: Defibrillation Sync. and Nurse Call. Def ibrillation Sync (X5, pin1) indication is generated by ECG. When active, the signal is in state 1. After 10 ms the signal is reset to state 0. Ne w Defibrillation Sync is not generated before the indication is deactivated. The delay from the R wave...
Page 163
Frame for FM 21 Document no. M1187335-004 3 Service procedures 3.1 General service information The field service of S/5 FM is limited to replacin g faulty PC boards and mechanical parts. Return the faulty PC boards to GE Healthcare for repair. GE Healthcare is always available for service advice. Provide the unit serial number, full type designation and a detailed fault description. WARNING Only trained personnel with appropri ate tools and equipment should perform the tests and repair s outlined in...
Page 164
Datex-Ohmeda S/5 FM 22 Document no. M1187335-004 Table 2 Patient simulators’ compatibility with each hemodynamic module Table 3 Adapter cables for hemodynamic patient simulators 3.2.3 Recommended parts Table 4 Recommended parts 3.2.4 Before beginning • Make sure that the monitor is switched to standby. •If the monitor is connected to Datex-Ohm eda Network, disconnect the Mon-Net cable from the monitor. Take out th e memory card, if installed. •Disconnect the power cord and detach the batteries. •Detach...
Page 165
Frame for FM 23 Document no. M1187335-004 3.2.5 General 1. Internal partsCheck that: the screws are tightened properly the cables are connected properly the connector pins are intact there are no loose objects inside the monitor Reassemble the monitor. 2. External parts Check that: all stickers are intact the LCD display screen is intact the ComWheel cover is intact and is attached properly all mechanical parts are functioning properly 3. Module installation Check that the modules...
Page 166
Datex-Ohmeda S/5 FM 24 Document no. M1187335-004 6. Module recognitionConfigure the monitor screen so that a ll needed parameter information is shown: Monitor Setup Screen 1 Setup Waveform Fields Digit Fields Check that the selected parameter information is shown on the screen. NOTE: InvBP waveforms are not shown without a patient simulator. Do no connect the simulator to the monitor at this stage. 7. LCD display picture Check that the picture on the LCD display screen is straight, clear and stable. ...
Page 167
Frame for FM 25 Document no. M1187335-004 11. Monitor softwareEnter the service menu: Monitor Setup Install Service (Password 16-4-34) Service (Password 26-23-8) Take down the information regarding Monitor software. 12. Content of the service log Select Service Log from the menu. Check the contents of the Service Log for possible problems. If the monitor contains the recorder unit, record the Service Log onto the recorder by selecting Record Log. 13. Voltages Check the power supply output...
Page 168
Datex-Ohmeda S/5 FM 26 Document no. M1187335-004 Delay --> Off Paper Speed --> 6.25 mm/s Length --> 30 s 15. Alarm LEDs Enter the Keyboard service menu: Monitor Setup Install Service (Password 16-4-34) Service (Password 26-23-8) Keyboard Select the text Upper Led. Check that the red alarm LED turns on and off when pressing the ComWheel. Check also the yellow alarm LED by selecting Lower Led from the menu. 16. Membrane keys Press the keys on the monitor command board one by one. Check that each...
Page 169
Frame for FM 27 Document no. M1187335-004 19. BatteriesEnter the BATTERY service menu: Monitor Setup Install Service (Password 16-4-34) Service (Password 26-23-8) Frame Power Supply WPM Battery Smart Batt1 Smart Batt2 Check that all values are within reasonable limits. NOTE: In order to update values for a desired battery, disconnect the other battery shortly. 20. Battery operation Disconnect the power cord (without switching the monitor to standby). Check that the monitoring continues norm ally....
Page 170
Datex-Ohmeda S/5 FM 28 Document no. M1187335-004 NOTE: InvBP waveforms are not recorded without a patient simulator. NOTE: If no recording appears, check that the paper roll is installed in a correct way. Only one side of the paper is printable. 24. Paper speed Press the key Start/Stop again and this time wait until the recording stops. Check that the length of the recorded waveform scale is 18.7 cm ( 1.5 cm). Change the paper speed setting to 1 mm/s: Print/Record Record Waveforms Paper Speed -->...