Page 151

Frame for FM 9
Document no. M1187335-004
Connectors
Interconnection board connector X3
Flat panel connector X7 (to display connection board)
Speaker connector X17
Analog output board connector  X4 (to sync connector board)
Ethernet connector X20 (network  cable to network connector board)
System watchdog and voltage supervision
There are two voltage supervision chips that control the system reset signals. 
The +3.3V supervision chip outputs reset signal s for +3.3V devices. Reset is activated when...

Page 152

Datex-Ohmeda S/5 FM
10
Document no. M1187335-004
The boost converter can be set to two different output levels. The higher one is used to give the 
battery charger adequate input voltage when the DC/DC board input comes from an external 
DC source. The lower one is set in AC/DC use to keep the boost converter enabled but passed 
by a diode and thus not switching, which minimizes the power loss. When the AC/DC voltage 
drops, the boost converter starts regulating its output and keeps the MOD voltages at a...

Page 153

Frame for FM 11
Document no. M1187335-004
Battery charger
Battery charger is a Level 2 Smart Battery Charger. It communicates with the batteries and 
PMC CPU via SMBus interface. The charger acts as an SMBus slave device that responds to 
charging current and charging voltage values sent to it by a Smart Battery. 
The charger includes an input current limiting feature. In a case where the input current 
exceeds the limit, the charger reduces the output  current to keep the input within the limit. The...

Page 154

Datex-Ohmeda S/5 FM
12
Document no. M1187335-004
2.1.6 Module bus cable unit
Figure 8 Module bus cable unit
Interconnection Board connector 
Module bus connector (pins 21…25)
CPU board connector
DIS connector X4
Sync IABP & DEF connector 
2 pcs grounding wires 
2.1.7 Interconnection board
 
Figure 9 Interconnection board
Interconnection board takes care of most of the monitor internal cabling.
CPU board connector X2
Module bus flex board connector X6
Fuse for External DC input (for future use)
DC/DC...

Page 155

Frame for FM 13
Document no. M1187335-004
Battery 2 connector X4
User interface board connector X1
2.1.8 Display connection board
Figure 10 Display connection board
CPU board connector X1
Display connector X2
2.1.9 Front cover parts
Figure 11 Parts inside Front cover 
(1)  Alarm light board
(2)  Alarm indicator board
(3)  Indicator ribbon cable
(4)  User interface board
(5)  Power indicator board
5
1
2
3
4 

Page 156

Datex-Ohmeda S/5 FM
14
Document no. M1187335-004
2.1.10 Wireless LAN option, F-FMW 
Datex-Ohmeda Wireless Network is an extension to, or an alternative for, a wired Datex-
Ohmeda Monitor network. It provides same network services than the wired Datex-Ohmeda 
Network.
The monitor network is a local area network based on standard Ethernet technology. The 
monitor network is formed by connecting one Central and up to 32 bedside monitors together. 
The Central and the bedside monitors are connected to a 10...

Page 157

Frame for FM 15
Document no. M1187335-004
2.2 Connectors and signals
2.2.1 External connectors
Figure 12 External connectors of Frame for FM
(1) Fuses
(2) Potential equalization connector
X1     NET connector
X2     Multi IO connector
X3     Receptacle for power cord
X4     Module Connector for E-PSM,  E-PSMP or N-FMC, N-FMCREC, N-FMREC
X5     Defib & IABP sync connector
X6     Connector for DIS interfacing system 
X7     Connector for Remote Controller K-CREMCO
X8     NET ID connector
X9     Serial...

Page 158

Datex-Ohmeda S/5 FM
16
Document no. M1187335-004
 Network connector, X1
Multi I/O connector X2
  
RJ45 connectorPinSignal
1
2
3
4
5
6
7
8 Tx +
Tx -
Rx +
N/C
N/C
Rx -
N/C
N/C
25 pin female connector Pin
Signal
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25 GND 
NET_ID_DO, TTL out
NET_ID_DI, TTL input
NET_ID_CLK, TTL out
NET_ID_CS, TTL out 
TXD, RS-232 output
RXD, RS-232 input
NURSE_CALL, CMOS output
reserved
Data -, RS-485 I/O
CONNECTOR_ID, TTL input 
EXTDC+, 10-16V, inpu
t (for future...

Page 159

Frame for FM 17
Document no. M1187335-004
Multi I/O adapter
Figure 13 Multi I/O adapter 
Connector for K-CREMCO X 7  
Net ID connector X8
Serial port X9  
5 pin connector PinSignal
1
2
3
4
5 KB_CLK
KB_DATA 
not in use
GND
+4.75 - +5.25V
9 pin female connector Pin Signal
1
2
3
4
5
6
7
8
9 NET_ID_CS, TTL out.
NET_ID_CLK, TTL out.
NET_ID_DO, TTL out.
NET_ID_DI, TTL input
GND
+4.75 . . . +5.25V
Not in use
NURSE_CALL, CMOS output
GND for Network ID
9 pin male connector Pin Signal
1
2
3
4
5
6
7
8
9 GND
RXD,...

Page 160

Datex-Ohmeda S/5 FM
18
Document no. M1187335-004
Main power X3
Module connector X4
Defib & IABP sync connector X5
DIRECT_ECG: Output signal follows first user lead (ECG1 lead) including augmented leads. If 
pace is on, there is an enhanced pacer pulse  (5 volts at 2 milliseconds) when pacemaker 
triggering is detected.
In the event of a LEAD FAIL, refer to the following: 
Lead Failed ECG Analog Out
Right arm III
Left arm II
Left leg I
Chest II
On LEADS OFF situation ECG analog output is flat line (0V)....