Page 101

Service Menu13
Document no. M1187329-003
1.4 Power supply
The menu shows the voltages and temperature 
measured by the DC/DC board and CPU board. The 
values in the column under  Mean are the mean 
values of last one second, the  Min column shows 
the minimum mean value, and the  Max column the 
maximum mean value of the voltages and 
temperature measured during the current power 
ON.
The voltages and currents are measured by the 
DC/DC Board, except the four lowest under 
heading CPU Board A/D, which...

Page 102

Datex-Ohmeda S/5 FM
14
Document no. M1187329-003
Power Temp Power supply unit temperature, measured at DC/DC Board.
Therm Not CHG This is subject to change. 
Dummy CHG  This is subject to change.
CPU Board Temp  Temperature measured at CPU Board.
VSYS_OUT VSYS that is connected to CPU Board and Backlight Inverter by a FET. 
Measured at CPU Board. Range 9…16.5V.
+3.3V  Measured at CPU Board. Range 3.15…3.45V.
+5V Measured at CPU Board. Range 4.75…5.25V. 

Page 103

Service Menu15
Document no. M1187329-003
1.4.1WPM Battery
Batts
This page contains information related to the 
batteries and power supplies. The power supply 
part is practically the same as in Power Page. 
Battery information includes also data measured 
by the smart batteries themselves and 
transmitted to the DC/DC Board via SMBus.
SMBus is System Management Bus, a two-wire 
interface closely resembling I2C. SMBus is used 
for battery communication. 
Batts information section of the page has two...

Page 104

Datex-Ohmeda S/5 FM
16
Document no. M1187329-003
Smart Batt1, Smart Batt2
Clear Temp Maxs
This command is useful only when the WPM Battery Batts view is selected. 
The Clear Temp Maxs command clears the maximum values of Batt Temp (SMBus) and 
Batt Temp (PMC).
Note: Power temp and CPU temp maxs are not cleared. 
These menus give additional battery information via SMBus. Batt1 stands for Battery A, Batt2 
for  Battery B. If Smart Battx is selected for  the battery connected to the SMBus, the menu...

Page 105

Service Menu17
Document no. M1187329-003
Smart Batt1
Battx information from SMBus (all this 
information comes from the battery via SMBus)
Temperature Battery temperature
Voltage  Battery voltage
Current  Battery current (discharge or 
charge)
Avg. Current Rolling average of the battery 
current
Rel. State of Charge Relative state of charge of the 
battery. Expressed as a 
percentage of Rem. Cap. (mAh)/ 
Full Cap.(mAh).
Abs. State of Charge        
Absolute state of charge. 
Expressed as a percentage of...

Page 106

Datex-Ohmeda S/5 FM
18
Document no. M1187329-003
Smart Batt2
If Smart Battx is selected for a battery NOT 
connected to the SMBus, the menu contains the 
following: 
Battx information from Memory and PMC (This 
information comes from the DC/DC Board 
memory or is measured by the DC/DC Board. 
SMBus data in this menu is not real time, because 
this battery is not connected to the SMBus). 
NOTE: This page may not contain information if 
SMBus has been connected only to the other 
battery. Page can be...

Page 107

Service Menu19
Document no. M1187329-003
2Display
Geometry  views the geometry of the display.
Colors  views the color of the screen.
Readability  views the readabilit y of the screen. 

Page 108

Datex-Ohmeda S/5 FM
20
Document no. M1187329-003
3 Keyboard
Keyboard
The service menu for testing the command board 
functions.
Upper Led is for testing the upper alarm LED (red) 
on the command board. When the text is 
highlighted, the upper alarm LED can be turned 
on and off by pressing the ComWheel.
Lower Led  is for testing the lower alarm LED 
(yellow) on the command board. When the text is 
highlighted, the lower alarm LED can be turned on 
and off by pressing the ComWheel.
Dummy Press  is for...

Page 109

Service Menu21
Document no. M1187329-003
3.1 Keyboard Log
Keyboard Scroll Log
All the keyboard presses and the commands 
given by the ComWheel are recorded in the 
Keyboard Log. The keyboard log is saved in the 
permanent memory of th e monitor. The length 
of the log is 1150 events. The log is FIFO type.
Scroll Stat  enables to scroll the keyboard 
events. 

Page 110

Datex-Ohmeda S/5 FM
22
Document no. M1187329-003
4 Parameters
NOTE: Parameter values in Service Data fields are only for reference in this section.
4.1 Gas Unit 