Page 101
Service Menu13 Document no. M1187329-003 1.4 Power supply The menu shows the voltages and temperature measured by the DC/DC board and CPU board. The values in the column under Mean are the mean values of last one second, the Min column shows the minimum mean value, and the Max column the maximum mean value of the voltages and temperature measured during the current power ON. The voltages and currents are measured by the DC/DC Board, except the four lowest under heading CPU Board A/D, which...
Page 102
Datex-Ohmeda S/5 FM 14 Document no. M1187329-003 Power Temp Power supply unit temperature, measured at DC/DC Board. Therm Not CHG This is subject to change. Dummy CHG This is subject to change. CPU Board Temp Temperature measured at CPU Board. VSYS_OUT VSYS that is connected to CPU Board and Backlight Inverter by a FET. Measured at CPU Board. Range 9…16.5V. +3.3V Measured at CPU Board. Range 3.15…3.45V. +5V Measured at CPU Board. Range 4.75…5.25V.
Page 103
Service Menu15 Document no. M1187329-003 1.4.1WPM Battery Batts This page contains information related to the batteries and power supplies. The power supply part is practically the same as in Power Page. Battery information includes also data measured by the smart batteries themselves and transmitted to the DC/DC Board via SMBus. SMBus is System Management Bus, a two-wire interface closely resembling I2C. SMBus is used for battery communication. Batts information section of the page has two...
Page 104
Datex-Ohmeda S/5 FM 16 Document no. M1187329-003 Smart Batt1, Smart Batt2 Clear Temp Maxs This command is useful only when the WPM Battery Batts view is selected. The Clear Temp Maxs command clears the maximum values of Batt Temp (SMBus) and Batt Temp (PMC). Note: Power temp and CPU temp maxs are not cleared. These menus give additional battery information via SMBus. Batt1 stands for Battery A, Batt2 for Battery B. If Smart Battx is selected for the battery connected to the SMBus, the menu...
Page 105
Service Menu17 Document no. M1187329-003 Smart Batt1 Battx information from SMBus (all this information comes from the battery via SMBus) Temperature Battery temperature Voltage Battery voltage Current Battery current (discharge or charge) Avg. Current Rolling average of the battery current Rel. State of Charge Relative state of charge of the battery. Expressed as a percentage of Rem. Cap. (mAh)/ Full Cap.(mAh). Abs. State of Charge Absolute state of charge. Expressed as a percentage of...
Page 106
Datex-Ohmeda S/5 FM 18 Document no. M1187329-003 Smart Batt2 If Smart Battx is selected for a battery NOT connected to the SMBus, the menu contains the following: Battx information from Memory and PMC (This information comes from the DC/DC Board memory or is measured by the DC/DC Board. SMBus data in this menu is not real time, because this battery is not connected to the SMBus). NOTE: This page may not contain information if SMBus has been connected only to the other battery. Page can be...
Page 107
Service Menu19 Document no. M1187329-003 2Display Geometry views the geometry of the display. Colors views the color of the screen. Readability views the readabilit y of the screen.
Page 108
Datex-Ohmeda S/5 FM 20 Document no. M1187329-003 3 Keyboard Keyboard The service menu for testing the command board functions. Upper Led is for testing the upper alarm LED (red) on the command board. When the text is highlighted, the upper alarm LED can be turned on and off by pressing the ComWheel. Lower Led is for testing the lower alarm LED (yellow) on the command board. When the text is highlighted, the lower alarm LED can be turned on and off by pressing the ComWheel. Dummy Press is for...
Page 109
Service Menu21 Document no. M1187329-003 3.1 Keyboard Log Keyboard Scroll Log All the keyboard presses and the commands given by the ComWheel are recorded in the Keyboard Log. The keyboard log is saved in the permanent memory of th e monitor. The length of the log is 1150 events. The log is FIFO type. Scroll Stat enables to scroll the keyboard events.
Page 110
Datex-Ohmeda S/5 FM 22 Document no. M1187329-003 4 Parameters NOTE: Parameter values in Service Data fields are only for reference in this section. 4.1 Gas Unit