GE S5 Owners Manual

Here you can view all the pages of manual GE S5 Owners Manual. The GE manuals for Medical Monitor are available online for free. You can easily download all the documents as PDF.

Page 91

Service Menu3
Document no. M1187329-003
1 Frame
1.1 Memory
The Frame menu includes service menus 
common for the frame.
A service menu to check the status of the 
memory used in the CPU board of the monitor.
Test Memory  tests the condition of the 
EEPROM/Flash memory component of the CPU 
board. If the result of the test is Fail , see section 
“Error messages” in the  slot.
Test SRAM  tests the Static RAM memory of the 
CPU board in a similar way as the EEPROM/Flash 
memory. If the result  of the test...

Page 92

Datex-Ohmeda S/5 FM
4
Document no. M1187329-003
1.2 Network
1.2.1 Network Status
The Network Status view shows the general 
status of the network
Location ID: Monitor’s location given at the 
setup. 
DRI level : Shows the selected level of network 
communication. The network communication 
is set according to the network software used 
(e.g. S-CNET02)
Interfaces: The field indicates if there is a 
connection to the Datex-Ohmeda Network
Current Interface : The field indicates the 
active network interface...

Page 93

Service Menu5
Document no. M1187329-003
Subnet Status
The Subnet status view gives more accurate 
information of the different subnet id:s connected. 
All four Subnet status  menus have a similar 
structure. The number of different packets 
transmitted and received by the monitor are 
shown in the columns below Tx and Rx. The packet 
types are:
Waveforms:  Waveform data
Phys. data:  Physiological numerical data
Alarms:  Alarms, alarm profiles and alarm
limits
Link mgmt:  Network management messages...

Page 94

Datex-Ohmeda S/5 FM
6
Document no. M1187329-003
1.2.2 Network Config
The DRI Level is for setting the monitors 
network communication. The network 
communication is set ac cording to the network 
software used (e.g. S-CNET01).
Network software S-CNET 01 ->DRI Level = 2001
Network software S-CNET02 ->DRI Level = 2001 
or 2002 (WLAN)
Network software L-NET03 ->DRI Level = 2003
Network software L-NET05 ->DRI Level = 2003
Virtual Plug ID  indicates the monitor’s location 
(Network ID). Virtual Plug ID is...

Page 95

Service Menu7
Document no. M1187329-003
1.2.4 WLAN
Driver: WLAN driver chip name.
Connection indicates the state of the WLAN 
connection.
Ethernet Addr : Ethernet address of the WLAN card 
used. 
AP MAC:  Ethernet address of the access point 
where the monitor is currently connected.
Signal Quality  indicates the quality of the radio 
signal between the monitor and the access point 
(0…100%).
Channel  is the WLAN channel configured to the 
access point where the monitor is currently 
connected.
Speed...

Page 96

Datex-Ohmeda S/5 FM
8
Document no. M1187329-003
AP List
Access Points: shows the access points (max. 
eight) which are visible for the monitor. 
CH  is the WLAN channel configured to this 
particular access point.
NL  indicates the noise level of the channel used  
(0…100%) 
SL  indicates the signal level of the channel used  
(0…100%)
WLAN Config
Network ID  is for setting the correct WLAN 
network ID name. 
WEP Algorithm is for setting the level of the 
encryption (width of encryption key) used for...

Page 97

Service Menu9
Document no. M1187329-003
1.2.5 Socket
For factory use only. 

Page 98

Datex-Ohmeda S/5 FM
10
Document no. M1187329-003
1.3 MemCards
1.3.1 Status
Present and Active  indicate the status of the 
PCMCIA controller on the CPU board. Possible 
values are YES and NO.
ROM  indicates the status of the ROM memory. 
Possible values are OK and ERR.
RAM  indicates the status of the RAM memory. 
Possible values are OK and ERR.
PCMCIA  indicates the status of the PCMCIA 
controller. Possible values are OK and ERR.
EEPROM  indicates the status of the EEPROM 
memory. Possible values are...

Page 99

Service Menu11
Document no. M1187329-003
1.3.2 Communication
Interface status  indicates the status of the data 
link between the monitor and memory module. 
The status should always be on ACTIVE. If the 
status blinks between ACTIVE and CLOSED, a 
communications error has occurred.
Message types  indicates the type of data packets 
that have been sent ( Tx) and received ( Rx) since 
the last monitor start. Data types are listed on the 
lines below  Message types text.
Record K  indicates the...

Page 100

Datex-Ohmeda S/5 FM
12
Document no. M1187329-003
1.3.3 PCMCIA Board
Chip type is the type of the PCMCIA controller 
chip used.
Number of slots  is the number of PCMCIA card 
slots on the CPU board (0, 1 or 2).
Slot A/B State  and Drv Active  indicate if the 
PCMCIA card slot is active.
Driver Ptr, Detect Int  and PcCard Int: PCMCIA 
card driver software related data. 
PcCard info : General PCMCIA card related 
information. 