Page 91
Service Menu3 Document no. M1187329-003 1 Frame 1.1 Memory The Frame menu includes service menus common for the frame. A service menu to check the status of the memory used in the CPU board of the monitor. Test Memory tests the condition of the EEPROM/Flash memory component of the CPU board. If the result of the test is Fail , see section “Error messages” in the slot. Test SRAM tests the Static RAM memory of the CPU board in a similar way as the EEPROM/Flash memory. If the result of the test...
Page 92
Datex-Ohmeda S/5 FM 4 Document no. M1187329-003 1.2 Network 1.2.1 Network Status The Network Status view shows the general status of the network Location ID: Monitor’s location given at the setup. DRI level : Shows the selected level of network communication. The network communication is set according to the network software used (e.g. S-CNET02) Interfaces: The field indicates if there is a connection to the Datex-Ohmeda Network Current Interface : The field indicates the active network interface...
Page 93
Service Menu5 Document no. M1187329-003 Subnet Status The Subnet status view gives more accurate information of the different subnet id:s connected. All four Subnet status menus have a similar structure. The number of different packets transmitted and received by the monitor are shown in the columns below Tx and Rx. The packet types are: Waveforms: Waveform data Phys. data: Physiological numerical data Alarms: Alarms, alarm profiles and alarm limits Link mgmt: Network management messages...
Page 94
Datex-Ohmeda S/5 FM 6 Document no. M1187329-003 1.2.2 Network Config The DRI Level is for setting the monitors network communication. The network communication is set ac cording to the network software used (e.g. S-CNET01). Network software S-CNET 01 ->DRI Level = 2001 Network software S-CNET02 ->DRI Level = 2001 or 2002 (WLAN) Network software L-NET03 ->DRI Level = 2003 Network software L-NET05 ->DRI Level = 2003 Virtual Plug ID indicates the monitor’s location (Network ID). Virtual Plug ID is...
Page 95
Service Menu7 Document no. M1187329-003 1.2.4 WLAN Driver: WLAN driver chip name. Connection indicates the state of the WLAN connection. Ethernet Addr : Ethernet address of the WLAN card used. AP MAC: Ethernet address of the access point where the monitor is currently connected. Signal Quality indicates the quality of the radio signal between the monitor and the access point (0…100%). Channel is the WLAN channel configured to the access point where the monitor is currently connected. Speed...
Page 96
Datex-Ohmeda S/5 FM 8 Document no. M1187329-003 AP List Access Points: shows the access points (max. eight) which are visible for the monitor. CH is the WLAN channel configured to this particular access point. NL indicates the noise level of the channel used (0…100%) SL indicates the signal level of the channel used (0…100%) WLAN Config Network ID is for setting the correct WLAN network ID name. WEP Algorithm is for setting the level of the encryption (width of encryption key) used for...
Page 98
Datex-Ohmeda S/5 FM 10 Document no. M1187329-003 1.3 MemCards 1.3.1 Status Present and Active indicate the status of the PCMCIA controller on the CPU board. Possible values are YES and NO. ROM indicates the status of the ROM memory. Possible values are OK and ERR. RAM indicates the status of the RAM memory. Possible values are OK and ERR. PCMCIA indicates the status of the PCMCIA controller. Possible values are OK and ERR. EEPROM indicates the status of the EEPROM memory. Possible values are...
Page 99
Service Menu11 Document no. M1187329-003 1.3.2 Communication Interface status indicates the status of the data link between the monitor and memory module. The status should always be on ACTIVE. If the status blinks between ACTIVE and CLOSED, a communications error has occurred. Message types indicates the type of data packets that have been sent ( Tx) and received ( Rx) since the last monitor start. Data types are listed on the lines below Message types text. Record K indicates the...
Page 100
Datex-Ohmeda S/5 FM 12 Document no. M1187329-003 1.3.3 PCMCIA Board Chip type is the type of the PCMCIA controller chip used. Number of slots is the number of PCMCIA card slots on the CPU board (0, 1 or 2). Slot A/B State and Drv Active indicate if the PCMCIA card slot is active. Driver Ptr, Detect Int and PcCard Int: PCMCIA card driver software related data. PcCard info : General PCMCIA card related information.