Page 171
Frame for FM 29 Document no. M1187335-004 29. Data Card recognitionSelect Status. Insert a Memory card labelled ”Data” in the memory card slot. Check that the message ‘Data Card inserted’ appears on the message field and the green menu card symbol on the upper right hand corner of the screen within 1 minute. Wait until the information regarding SLOT2 is fully updated in the service menu, then check that the ‘Card type’ states DATA and the ‘File system’ ATA. Check that the rest of the information...
Page 172
Datex-Ohmeda S/5 FM 30 Document no. M1187335-004 3.2.8 Network 32. Connection to networkCheck that the Mon-Net cable connector is clean and intact, then connect it to the monitor. Check that the monitor connec ts to the Datex-Ohmeda Network, i.e. the network symbol appears under the clock in the upper right hand corner of the screen. Also a message regarding the connected Central should appear in the message field of the screen. If the message regarding the Central does not appear, check the status...
Page 173
Frame for FM 31 Document no. M1187335-004 Reconnect the Identification plug and che ck that the monitor connects to the Datex- Ohmeda Network. 3.2.9 Wireless LAN option 38. Signal strength Check that the WLAN signal strength symbol roams or stays fixed on the monitor screen. 39. Configure access point Configure one access point appropriately. Bri ng the monitor to good WLAN coverage area. Check that the field le vel indicator stops scrolling. Go to the WLAN Status page (... Service -> Frame ->...
Page 174
Datex-Ohmeda S/5 FM 32 Document no. M1187335-004 Clear the content of the Service Log by selecting RESET LOG from the menu. 44. Electrical safety checkPerform electrical safety check and leakage current test. Check that the monitor functions normally af ter the performed electrical safety check. 45. Final cleaning Switch the monitor to standby, disconnect the power cord and perform final cleaning. 46. Fill in all necessary documents 3.3 S/5 FM disassembly and reassembly WARNING The S/5 FM is...
Page 175
Frame for FM 33 Document no. M1187335-004 Tools needed - screwdrivers, TORX; T8, T10, T15 - crosshead screwdriver; M15 - flat blade screwdriver - sockets; M4.5, M7, M11 - antistatic wristband To separate the back cover from the monitor frame 1. From the back side of the monitor, remove: - both batteries - Multi I/O adapter - the fuse holder - three (T8) screws from the top of the back cover - two (T15) screws with star washers - one (T15) screw Notice the places of the star washers. 2. Remove the two...
Page 176
Datex-Ohmeda S/5 FM 34 Document no. M1187335-004 3. Remove the two (T8) screws from the datacard slot. 4. Place the monitor face down on a non-adrasive, static-free surface. NOTE: Be careful not to damage the ComWheel. - Lift the back cover up. Now you can access the following field replaceable parts. Follow the instructions below: -“To remove the networ k connector board unit” - “To remove the main power connector unit” - “To remove the Multi I/0 connector (25 pins)” - “To remove the DC/DC board” -...
Page 177
Frame for FM 35 Document no. M1187335-004 To remove the ne twork connector board unit - Remove the screw (T10) holding the unit to the frame. - Disconnect the cable coming from the Network connector board - Remove the Network connector board To remove the main power connector unit - Remove the two screws (T10) holding the unit to the frame. - Disconnect the X1 connector and the protective earth connector from the receptacle. To remove the Multi I/ 0 connector (25 pins) - Remove the two terminal...
Page 178
Datex-Ohmeda S/5 FM 36 Document no. M1187335-004 To remove the DC/DC board - Remove the AC/DC board connector. - Remove the screw (T10) holding the unit to the frame. - Disconnect the DC/DC board from the frame by lifting it up carefully. - While reassembling, route the AC/DC board - DC/DC board cable according to the picture.
Page 179
Frame for FM 37 Document no. M1187335-004 To remove the AC/DC unit and the battery brake Disconnect: - the AC/DC unit - DC/DC unit cable connector - the AC/DC board - Filter board cable connector - the protective earth cable connector Remove: - the four screws (T10) holding the AC/DC unit to the frame - the AC/DC unit NOTE: It is not allowed to repair the AC/DC power unit in the field. Return the faulty AC/DC power unit to GE Healthcare for service. To remove the battery brake: - remove the two...
Page 180
Datex-Ohmeda S/5 FM 38 Document no. M1187335-004 WARNINGWear a grounded antistatic wristban d and soft cotton gloves (dust free) when handling the LCD display parts. Hold them by their corners or edges. Do not touch the connector pins. In normal circumstances it is very difficult to keep the LCD display component and the display shield surfaces free of dust when the LCD display unit is exposed to room air. If dust particles remain on the LCD display component and the display shield surfaces, they...