Page 171

Frame for FM 29
Document no. M1187335-004
29. Data Card recognitionSelect  Status.  Insert a Memory card labelled ”Data” in  the memory card slot. Check that 
the message ‘Data Card inserted’ appears on the message field and the green menu card 
symbol on the upper right hand corner of the screen within 1 minute.
Wait until the information regarding SLOT2 is fully updated in the service menu, then 
check that the ‘Card type’ states DATA and the ‘File system’ ATA.
Check that the rest of the information...

Page 172

Datex-Ohmeda S/5 FM
30
Document no. M1187335-004
3.2.8 Network
32. Connection to networkCheck that the Mon-Net cable connector is  clean and intact, then connect it to the 
monitor.
Check that the monitor connec ts to the Datex-Ohmeda Network, i.e. the network symbol 
appears under the clock in the upper right hand corner of the screen. Also a message 
regarding the connected Central should appear in the message field of the screen.
If the message regarding the Central does not appear, check the status...

Page 173

Frame for FM 31
Document no. M1187335-004
Reconnect the Identification plug and che ck that the monitor connects to the Datex-
Ohmeda Network.

3.2.9 Wireless LAN option
38. Signal strength
Check that the WLAN signal strength symbol  roams or stays fixed on the monitor screen. 
39. Configure access point Configure one access point appropriately. Bri ng the monitor to good WLAN coverage 
area. Check that the field le vel indicator stops scrolling.
Go to the  WLAN Status page  (... Service -> Frame ->...

Page 174

Datex-Ohmeda S/5 FM
32
Document no. M1187335-004
Clear the content of the Service Log by selecting RESET LOG from the menu.

44. Electrical safety checkPerform electrical safety check and leakage current test.
Check that the monitor functions normally af ter the performed electrical safety check.

45. Final cleaning
Switch the monitor to standby, disconnect the power cord and perform final cleaning.

46. Fill in all necessary documents

3.3 S/5 FM disassembly and reassembly
WARNING The S/5 FM is...

Page 175

Frame for FM 33
Document no. M1187335-004
Tools needed
- screwdrivers, TORX; T8, T10, T15
- crosshead screwdriver; M15
- flat blade screwdriver
- sockets; M4.5, M7, M11
- antistatic wristband
To separate the back cover from 
the monitor frame
1. From the back side of the monitor, remove:
- both batteries 
- Multi I/O adapter
- the fuse holder
- three (T8) screws from the top of the back cover
- two (T15) screws with star washers
- one (T15) screw
Notice the places of the star washers. 
2. Remove the two...

Page 176

Datex-Ohmeda S/5 FM
34
Document no. M1187335-004
3. Remove the two (T8) screws from the datacard slot. 
4. Place the monitor face down on a non-adrasive, static-free surface.
NOTE: Be careful not to damage the ComWheel.
- Lift the back cover up.
Now you can access the following field replaceable parts. 
Follow the instructions below:
-“To remove the networ k connector board unit”
- “To remove the main power connector unit”
- “To remove the Multi I/0 connector (25 pins)”
- “To remove the DC/DC board”
-...

Page 177

Frame for FM 35
Document no. M1187335-004
To remove the ne twork connector 
board unit
- Remove the screw (T10) holding the unit to the frame.
- Disconnect the cable coming from the Network 
connector board
- Remove the Network connector board
To remove the main  power connector unit
- Remove the two screws (T10) holding the unit 
to the frame.
- Disconnect the X1 connector and the  protective earth connector from the 
receptacle.
To remove the Multi I/ 0 connector (25 pins)
- Remove the two terminal...

Page 178

Datex-Ohmeda S/5 FM
36
Document no. M1187335-004
To remove the DC/DC board
- Remove the AC/DC board connector.
- Remove the screw (T10) holding the unit to the frame.
- Disconnect the DC/DC board from the frame  by lifting it up carefully.
- While reassembling, route the AC/DC board -  DC/DC board cable according to the picture. 

Page 179

Frame for FM 37
Document no. M1187335-004
To remove the AC/DC  unit and the battery 
brake
Disconnect:
- the AC/DC unit - DC/DC unit cable connector
- the AC/DC board - Filter board cable 
connector
- the protective earth cable connector
Remove:
- the four screws (T10) holding the AC/DC unit  to the frame
- the AC/DC unit
NOTE: It is not allowed to repair the AC/DC 
power unit in the field. Return the faulty AC/DC 
power unit to GE Healthcare for service.
To remove the battery brake:
- remove the two...

Page 180

Datex-Ohmeda S/5 FM
38
Document no. M1187335-004
WARNINGWear a grounded antistatic wristban d and soft cotton gloves (dust free) 
when handling the LCD display parts. Hold  them by their corners or edges. 
Do not touch the connector pins.
In normal circumstances it is very difficult  to keep the LCD display component and the display 
shield surfaces free of dust when the LCD display  unit is exposed to room air. If dust particles 
remain on the LCD display component and the display shield surfaces, they...