Page 241

Patient Side Module, E-PSM, E-PSMP (Rev. 01) 35
Document no. M1215098-002
3.3 Disassembly and reassembly
3.3.1 Before disassembly
WARNINGWear a grounded, antistatic wristband when handling PC boards. 
Electrostatic discharge may damage components on the board. 
Handle all PC boards by their edges.
3.3.2 Tools needed
CAUTION When reassembling the module, make  sure to reconnect all cables properly.
3.3.3 To disassemble the module
- torx screwdrivers; T6, T8
- flat blade screwdriver
-pincers
- antistatic...

Page 242

E-modules
36
Document no. M1215098-002
3. To remove the NIBP board:NOTE: You may remove the NIBP filter cover and the filter before 
disconnecting the flex cable.
- Disconnect the module bus co nnector, pump connector and 
NIBP flex connector. 
- Disconnect the hoses (2 pcs) coming from the manifold.
NOTE: Note the positions of the hoses; mark them if necessary 
to ensure they are replaced correctly.
- Remove the NIBP board.
4. Disconnect the air intake hose from the NIBP manifold. 
NOTE: The lips of the...

Page 243

Patient Side Module, E-PSM, E-PSMP (Rev. 01) 37
Document no. M1215098-002
5. Lift the NIBP-STP insulator plate carefully up. 
6. To remove the STP board
- Carefully open the connector lock and then disconnect the  STP input flex cable from the STP board.
NOTE: When reassembling, ensure that the flex cable is aligned 
properly and the connector is locked.
- Lift the STP board a little to disconnect the module bus  connector. Remove the STP board.
7. Remove the STP-ECG insulator plate. Be careful not to...

Page 244

E-modules
38
Document no. M1215098-002
Reassemble the module in reverse order.
Always perform the  “Service check” after reassembling the module.
9. To remove the ECG board
- While holding the ECG input unit out of the way, lift the ECG 
board a little and disconnect the module bus connector.
- Remove the ECG board.
10. Remove the NIBP filter cover and the filter. (If not removed  already.)
11. Remove the four screws (T6)  with washers holding the NIBP 
pump to the frame.
12.  Flip the module over and...

Page 245

Patient Side Module, E-PSM, E-PSMP (Rev. 01) 39
Document no. M1215098-002
To remove the pump unit
14. Disconnect the hoses from the manifold. The hoses follow the pump.
3.3.4 To remove the manifold unit
15. Disconnect the two (T6) screws holding the manifold to the front cover unit.
16.  Open the connector lock from the NIBP flex  board and disconnect the membrane 
keyboard flex.  

Page 246

E-modules
40
Document no. M1215098-002
17. Lift the manifold carefully aside. Be careful 
not to damage the NIBP flex board. 
Disconnect the NIBP flex board connector 
from the STP input board.
NOTE: When reassembling, make sure that the 
NIBP flex board connecto r is connected properly 
(all pins connected) to the STP input board.
Reassemble the module  in reverse order.
Always perform the  “Service check” after reassembling the module.
  

Page 247

Patient Side Module, E-PSM, E-PSMP (Rev. 01) 41
Document no. M1215098-002
Reassemble the module in reverse order. 
Always perform the  “Service check” after reassembling the module.
3.3.5 To remove the module bus connector
18. Use a flat blade screwdriver to unlock the 
module bus connecto r insulator cover. 
Put the screwdriver in the hole and move the 
blade backwards (away from the flex cable) 
until the insulato r cover unlocks.
19. Pull the module bus connector carefully  through the hole in the...

Page 248

E-modules
42
Document no. M1215098-002
3.3.6 To replace the NIBP filter
Reassemble the module in reverse order.
Always perform the  “Service check” after reassembling the module.1. Follow the disassemble instruction steps 1 
and 2.
2. Remove the NIBP filter cover and replace the  filter. 

Page 249

Patient Side Module, E-PSM, E-PSMP (Rev. 01) 43
Document no. M1215098-002
3.4 Adjustments and calibrations
NOTE: Use only properly maintained, calibrated and traceable measurement equipment for the 
specified calibrations and adjustments to ensure accuracy.
3.4.1 NIBP calibrations
The electronics of the NIBP pressure measurement is calibrated at the factory. The processor 
automatically maintains the zeroing pressure. If the zero point of the pressure transducer drifts 
more than specified, an error...

Page 250

E-modules
44
Document no. M1215098-002
Table 3 NIBP calibration check pressures
If the error of pressure channel B1 is larger than specified above, the module should be 
recalibrated. The error of B2 is allowed to be even twice as large because it has no effect on 
blood pressure measurement accuracy. Howeve r, we recommend recalibrating the module 
when the error of B2 is larger than spec ified above to ensure best possible operation.
Calibration
1. Enter Calibration  menu.
2. Remove the hoses from the...