Page 241
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 35 Document no. M1215098-002 3.3 Disassembly and reassembly 3.3.1 Before disassembly WARNINGWear a grounded, antistatic wristband when handling PC boards. Electrostatic discharge may damage components on the board. Handle all PC boards by their edges. 3.3.2 Tools needed CAUTION When reassembling the module, make sure to reconnect all cables properly. 3.3.3 To disassemble the module - torx screwdrivers; T6, T8 - flat blade screwdriver -pincers - antistatic...
Page 242
E-modules 36 Document no. M1215098-002 3. To remove the NIBP board:NOTE: You may remove the NIBP filter cover and the filter before disconnecting the flex cable. - Disconnect the module bus co nnector, pump connector and NIBP flex connector. - Disconnect the hoses (2 pcs) coming from the manifold. NOTE: Note the positions of the hoses; mark them if necessary to ensure they are replaced correctly. - Remove the NIBP board. 4. Disconnect the air intake hose from the NIBP manifold. NOTE: The lips of the...
Page 243
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 37 Document no. M1215098-002 5. Lift the NIBP-STP insulator plate carefully up. 6. To remove the STP board - Carefully open the connector lock and then disconnect the STP input flex cable from the STP board. NOTE: When reassembling, ensure that the flex cable is aligned properly and the connector is locked. - Lift the STP board a little to disconnect the module bus connector. Remove the STP board. 7. Remove the STP-ECG insulator plate. Be careful not to...
Page 244
E-modules 38 Document no. M1215098-002 Reassemble the module in reverse order. Always perform the “Service check” after reassembling the module. 9. To remove the ECG board - While holding the ECG input unit out of the way, lift the ECG board a little and disconnect the module bus connector. - Remove the ECG board. 10. Remove the NIBP filter cover and the filter. (If not removed already.) 11. Remove the four screws (T6) with washers holding the NIBP pump to the frame. 12. Flip the module over and...
Page 245
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 39 Document no. M1215098-002 To remove the pump unit 14. Disconnect the hoses from the manifold. The hoses follow the pump. 3.3.4 To remove the manifold unit 15. Disconnect the two (T6) screws holding the manifold to the front cover unit. 16. Open the connector lock from the NIBP flex board and disconnect the membrane keyboard flex.
Page 246
E-modules 40 Document no. M1215098-002 17. Lift the manifold carefully aside. Be careful not to damage the NIBP flex board. Disconnect the NIBP flex board connector from the STP input board. NOTE: When reassembling, make sure that the NIBP flex board connecto r is connected properly (all pins connected) to the STP input board. Reassemble the module in reverse order. Always perform the “Service check” after reassembling the module.
Page 247
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 41 Document no. M1215098-002 Reassemble the module in reverse order. Always perform the “Service check” after reassembling the module. 3.3.5 To remove the module bus connector 18. Use a flat blade screwdriver to unlock the module bus connecto r insulator cover. Put the screwdriver in the hole and move the blade backwards (away from the flex cable) until the insulato r cover unlocks. 19. Pull the module bus connector carefully through the hole in the...
Page 248
E-modules 42 Document no. M1215098-002 3.3.6 To replace the NIBP filter Reassemble the module in reverse order. Always perform the “Service check” after reassembling the module.1. Follow the disassemble instruction steps 1 and 2. 2. Remove the NIBP filter cover and replace the filter.
Page 249
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 43 Document no. M1215098-002 3.4 Adjustments and calibrations NOTE: Use only properly maintained, calibrated and traceable measurement equipment for the specified calibrations and adjustments to ensure accuracy. 3.4.1 NIBP calibrations The electronics of the NIBP pressure measurement is calibrated at the factory. The processor automatically maintains the zeroing pressure. If the zero point of the pressure transducer drifts more than specified, an error...
Page 250
E-modules 44 Document no. M1215098-002 Table 3 NIBP calibration check pressures If the error of pressure channel B1 is larger than specified above, the module should be recalibrated. The error of B2 is allowed to be even twice as large because it has no effect on blood pressure measurement accuracy. Howeve r, we recommend recalibrating the module when the error of B2 is larger than spec ified above to ensure best possible operation. Calibration 1. Enter Calibration menu. 2. Remove the hoses from the...