Page 291
S/5 Extension Module for FM, N-FC, N-FCREC, N-FREC 21 Document no. M1187338-003 11. Remove the CO2 unit module bus flex cable from the REC board. NOTE: Be careful with the in sulation plate covering the cable connector on the REC board. 12. Lift the CO2 unit carefully up and remove the unit from the module cover. 13. Remove the screw (T8) holding the module bus flex cable to recorder unit metal plate and module cover. 14. Lift the REC board carefully up and disconnect the two module bus flex cables...
Page 292
Datex-Ohmeda S/5 FM 22 Document no. M1187338-003 16. Use a flat blade screwdriver to unlock the module bus connector insulator cover. Put the screwdriver in the hole and move the blade backwards (away from the module bus flex cable) until the insulator cover unlocks. 17. By assisting with a screwdriver pull the module bus connector through the hole in the module cover. 18. Flip the module cover over and remove the two (T6) screws holding the lock un it to the frame. While pulling the tab push the...
Page 293
S/5 Extension Module for FM, N-FC, N-FCREC, N-FREC 23 Document no. M1187338-003 3.4 Adjustments and calibrations 3.4.1 Calibrating The airway module should be calibrated once every six months or whenever there are indications of errors in the gas readings. Calibrate the gas measurement with the Datex-Ohmeda calibration gas. Do not use any other calibration gases. −Use the regulator 755534-HEL. − Use gas 755580 only and set the FiO2 le vel in the CO2 setup menu to 21-40% Airway Gas - CO2 setup. .O 2...
Page 294
Datex-Ohmeda S/5 FM 24 Document no. M1187338-003 Figure 9 Connecting sampling line to the gas valve and feeding gas 3. For maximum accuracy, let the monitor warm up for 30 minutes. The menu item Gas calibration remains gray as long as the message ‘Calibrating gas sensor’ is displayed. 4. Press the Airway Gas key and select Gas calibration . 5. Wait until the ‘Zero ok’ and then the ‘Feed gas’ messages appear after each gas on the screen. 6. Open the regulator and feed calibration gas until the...
Page 295
S/5 Extension Module for FM, N-FC, N-FCREC, N-FREC 25 Document no. M1187338-003 •Turn the ComWheel until the displayed value matches the desired value in the gas bottle and push it again. If the message ‘Zero error’ is displayed, press the Normal Screen key and repeat the calibration procedure. The time of the last calibration is shown at the bottom of the menu page. 3.4.2 Gas sampling system adjustment For flow rate measurements, a flow meter with a low flow resistance and the capability to...
Page 296
Datex-Ohmeda S/5 FM 26 Document no. M1187338-003 A change of 0.050 in the Gain value changes the flow approximately 7.5 ml/min. After you have changed the gain, wait until the Sample Flow value on the screen returns near to the original, then check the flow meter reading again. 3.4.5 Gas calibration Gas calibration is performed in the Airway Gas menu. Calibration gas regulator flow check Interval: every 12 months Regulator flow specification: REF 755583-HEL & 755534-HEL: 260 – 410 ml /min at 1-10 bar...
Page 297
S/5 Extension Module for FM, N-FC, N-FCREC, N-FREC 27 Document no. M1187338-003 4Troubleshooting 4.1 Troubleshooting chart for recorder unit 4.2 Troubleshooting chart for CO 2 measurement Problem CauseWhat to do Module not responding to front panel key, but operates through Recorder menu. Membrane switch cable loose or broken. Check the cable. Replace the front panel if necessary. Recorder will not start. No error messages shown. Recorder: Module not properly inserted.Re-insert the module properly....
Page 298
Datex-Ohmeda S/5 FM 28 Document no. M1187338-003 4.2.1 CO2 measurement ‘REPLACE D-FEND’ message Amb – Work pressure difference too big.Indicates residue build-up on the water trap membrane. This decreases air flow. Replace the D-fend. ‘SAMPLE LINE BLOCKED’ message Amb – Work pressure difference too big. Sampling line or water trap is occluded. Water trap container is full. If occlusion persists, check internal tubing for blockages. No response to any gas Check Sample Gas Out. Amb - Work pressure low,...
Page 299
S/5 Extension Module for FM, N-FC, N-FCREC, N-FREC 29 Document no. M1187338-003 4.3 Gas sampling system troubleshooting The faults which can occur in the sampling system are: leaks or blockages in the tubing, failure of the sampling pump or the magnetic valves, or diminishing of the flow rates because of dirt accumulating in the internal tubing. The following checks should help in localizing the fault. Whenever suspecting the sampling system and always after having done any work on the sampling...
Page 300
Datex-Ohmeda S/5 FM 30 Document no. M1187338-003 4.4 MiniC unit troubleshooting WARNINGThe miniC measuring unit can only be repaired and calibrated at the factory. Attempts to repair/calibrate the unit elsewhere will adversely affect operation of the unit. In case of failure, the complete miniC unit should be sent to GE Heal thcare for factory exchange. 4.5 Error messages Message Explanation Occlusion or Sample Line Blocked The sample tube inside or outside the monitor is blocked or water trap is...