Page 31

 System installation23
Document no. M1187317-009
3.3.6 Inserting and removing the parameter modules
1. Align the module with  the insertion guides 
2. Push the module into the monitor frame until it stops.
3. Pull the module outwards . Make sure not to drop it when it comes out.
WARNINGWhen detaching modules, be careful no t to drop them. Always support with 
one hand while pulling  out with the other.
NOTE: Only one E-PSM(P) module and one N-F(C)(REC) module can be attached to S/5 FM. 
NOTE: If you want...

Page 32

Datex-Ohmeda S/5 FM
24
Document no. M1187317-009
3. Switch the monitor on.
4. Wait for approximately 80 seconds. After the normal screen appears, enter the service 
menu and make sure that the information regarding monitor software has been updated. 
Memorize the serial number of new software.
5. Remove the software card.
6. Replace the original device plate for monitor software by the new one supplied with the  software card.
7. Perform Factory Reset. Make  sure that the monitor functions normally after...

Page 33

 System installation25
Document no. M1187317-009
3.3.9 Performing Factory Reset
NOTE: The Factory Reset is necessary after  downloading of monitor software and after 
replacing the CPU board or SRAM/Timekeeper battery.
NOTE: The Factory Reset will restore all your customized defaults, includ ing language selection, 
to factory defaults.
1. Press the 
Monitor Setup key.
2. Select  Install/Service  and password (16-4-34).
3. Select  Service and password (26-23-8).
4. Select  Set/Test and perform  Factory...

Page 34

Datex-Ohmeda S/5 FM
26
Document no. M1187317-009
10. Connect the WLAN antennas to the WLAN interface card. Secure the antennas to the PC card drive slot with the holder included in the upgrade kit.
11. Reassemble the monitor.
NOTE: When reassembling the monitor make sure th at the tips of the WLAN halfwave antennas 
do not get pinched between the monitor covers.
12. Switch the monitor ON.
13. Enter the service menu and make sure that the option software was downloaded 
successfully (
Monitor Setup...

Page 35

 System installation27
Document no. M1187317-009
16. Enter the Network Status service menu (Monitor Setup –Install/Service (16-4-34) – 
Service (26-23-8) – Frame - Network – Network Status ).
Check that the Virtual ID sticker that is supplied with the Wireless Network upgrade, 
U-FMW corresponds with the  Location ID that is displayed in the menu.
NOTE: The Location ID is the monitor’s ID on the Datex-Ohmeda Network and cannot be 
changed.
17. Attach the Virtual ID sticker above the RLAN card sticker at...

Page 36

Datex-Ohmeda S/5 FM
28
Document no. M1187317-009
See Appendix C. Channel Mask Selections for S/5 FM .
Save the monitor’s wireless LA N configuration by selecting Save Configuration - Save.
NOTE: The settings must match up with the ones that are defined in the Access Point 
configuration. Refer to the rela ted site configuration document ation for correct settings, if 
necessary.
8. Enter the Network Status service menu (
Monitor Setup  –Install/Service (16-4-34) – 
Service (26-23-8) – Frame - Network –...

Page 37

 System installation29
Document no. M1187317-009
After the patient monitor has restarted fo r the first time after monitor software 
reinstallation, perform a factory reset to the patient monitor. 
2. Enter the service menu and make sure that  the WLAN option was activated successfully 
(
Monitor Setup - Install/Service (16-4-34) - Service (26-23-8) - Frame) :
Options = WLAN
3. Switch the monitor to STANDBY.
4. Disassemble the monitor in order to access the Compact Flash card drive slot on the  monitor...

Page 38

Datex-Ohmeda S/5 FM
30
Document no. M1187317-009
10. Enter the WLAN Configuration service menu (Monitor Setup -Install/Service (16-4-34) - 
Service (26-23-8) - Frame - Network - WLAN - WLAN Config).
Enter the appropriate  Network ID, WEP Algorithm,  Key ID and Encryption Key  using the 
corresponding menu selections.
For selecting allowed communication channels, enter the appropriate  Channel Mask 
using the corresponding menu selection. See  Appendix C. Channel Mask Selections for S/5 
FM .
Save the...

Page 39

 System installation31
Document no. M1187317-009
3.4 Remote Controller, K-CREMCO
To connect a Remote Controller, K-CREMCO to  S/5 FM plug the connector to the 5-pin DIN 
connector X7 in Multi I/O adapter at the back of the monitor.
3.5 Visual indicators
Function Specification             Explanation
External power supply Green LED Indicates when monitor is powered from 
mains or ext. DC (External DC power for 
future use)
Battery operation  Green LED Indicates when monitor is powered from 
internal...

Page 40

Datex-Ohmeda S/5 FM
32
Document no. M1187317-009
4 Interfacing
External devices can be interfaced with the S/5 FM via the monitors serial port and via the 
Device Interfacing So lution, N-DISxxx.  
Printers and computers can be interfaced via the monitor’s serial port.
 
Device specific N-DISxxx modules can be used with:
•Monitors
•Blood-gas analyzers
NOTE: The Device Interfacing Solution, N-DISxxx th at is used on the S/5 FM must be of revision 
01 or later.
4.1 Interfacing external bedside devices via...