GE S5 Owners Manual
Here you can view all the pages of manual GE S5 Owners Manual. The GE manuals for Medical Monitor are available online for free. You can easily download all the documents as PDF.
Page 11
Introduction3 Document no. M1187317-009 1 Introduction The Datex-Ohmeda S/5 FM is a modular multiparameter patient monitor. The monitor is especially designed for monitoring in intensive care units. It can also be used during transportation within the hospital. The modular design makes the system flexible and easy to upgrade. In addition to patient parameter modularity and easy upgrades, the monitor can be upgraded to wireless networking. External devices can be interfac ed to the monitor with...
Page 12
Datex-Ohmeda S/5 FM 4 Document no. M1187317-009 1.1 Symbols 1.1.1 Symbols on transport packaging 1.1.2 Symbols on equipment The contents of the transport package are fragile and must be handled with care. Indicates the correct upright position of the transport package. The transport package must be kept in a dry environment. Indicates the temperature limitations within which the transport package should be stored. This package can be recycled. This battery contains lead and can be recycled. Dangerous...
Page 13
Introduction5 Document no. M1187317-009 1.1.3 Equipment safety symbols - Attention, consult accompanying documents. - When displayed next to the HR value, indicates that the pacer is set on R. - On the N-FC(REC) module indicates that the airway gases should be calibrated every 6 months in normal use and every two months in continuous use to ensure that the measurement accuracy remains within specifications. - On the E-PSM(P) module: WARNING Protection against cardiac defibrillator discharge is due...
Page 14
Datex-Ohmeda S/5 FM 6 Document no. M1187317-009 1.1.4 Other symbols Type CF (IEC 60601-1) defibrillator-proof protection against electric shock. When displayed in the upper left corner of the screen, indicates that the alarms are silenced. When displayed in the menu or digit fields, indicates that the alarm source has been turned off or alarm does not meet the alarm-specific activation criteria. ESD warning symbol for electrostatic sensitive devices. Pins of connectors identified with the ESD warning...
Page 15
Introduction7 Document no. M1187317-009 Battery (A) is missing In the front panel: mains or external DC power. (External DC power for future use.) The monitor is connected to the Datex-Ohmeda Network (Local Area Network, LAN). The monitor is connected to the Datex-Ohmeda S/5 Network (Wireless Local Area Network, WLAN). Data Card (green) or Menu Card (white) is inserted. WLAN signal strength. The number of segments corresponds to the signal strength: four segments indicate strong signal, one...
Page 16
Datex-Ohmeda S/5 FM 8 Document no. M1187317-009 Date of manufacturer Manufacturer name and address Does not contain Latex. Do not immerse the sensor in liquids. IPX class: IPX0 IPX1 IPX2 IPX3 IPX4 IPX7 IPX8 Degree of protection against harmful ingress of water as detailed in the IEC 60529: - Ordinary equipment - Protection against vertically falling water drops. - Protection against vertically fallin g water drops when enclosure tilted up to 15 °. - Protected against spraying water. - Protected against...
Page 17
Introduction9 Document no. M1187317-009 1.2 Safety The following list contains general warnings an d cautions you should know before installing, maintaining or servicing the system. Warnings and cautions specific to the use of the system can be found in the User’s Guide and User’s Reference Manual. 1.2.1 Safety precautions Warnings WARNING A WARNING indicates a situation in whic h the user or the patient may be in danger of injury or death. •The device is not able to withstand unpacked drops from...
Page 18
Datex-Ohmeda S/5 FM 10 Document no. M1187317-009 •Do not use the monitor in high electromagnetic fields (for example, during MRI.) •A printer or computer must be supplied from an additional transformer providing at least basic isolation (isolating or separating transformer). •If you accidentally drop the monitor, modules or frames, have them checked by authorized service personnel prior to clinical use. •To avoid explosion hazard, do not use the mo nitor in presence of flammable anesthetics. •Do not...
Page 19
Introduction11 Document no. M1187317-009 •Electrostatic discharge through the PC bo ards may damage the components. Before handling PC boards, wear a grounded antistatic wristband. Handle all PC boards by their non-conductive edges and use antistatic containers when transporting them. •Do not break or bypass the patient isolation barrier when testing PC boards. •Always perform an electrical safety check an d a leakage current test on the monitor after service. •Handle the water trap and its...
Page 20
Datex-Ohmeda S/5 FM 12 Document no. M1187317-009 Storage and transport Do not store or transport the monitor outside the specified temperature, pressure and humidity ranges: Temperature -20...+60 °C/-4...140 °F Atmospheric pressure 670...1060 hPa/500...800 mmHg/670...1060 mbar Relative humidity 10...90% noncondensing Disposal •Dispose of the whole device, parts of it and its packing material and manuals in accordance with local environmental and waste disposal regulations. •Dispose the calibration...