Page 171
Installing PageWriter Trim II, III, and Rx SoftwareInstalling the Software Upgrade PageWriter Trim Cardiograph Service ManualB-7 5Connect the cardiograph to AC power. Ensure that the green AC power indicator light is lit. WARNINGAlways connect the cardiograph to AC power when performing the software upgrade. If the cardiograph is not connected to AC power when performing the software upgrade it may be damaged. 6Ensure that the cardiograph is shut down (screen is black) and that the Standby light is...
Page 172
Installing PageWriter Trim II, III, and Rx SoftwareInstalling the Software Upgrade B-8PageWriter Trim Cardiograph Service Manual 9The Software Installation Utility screen appears. Figure B-19 Software Installation Utility screen 10 Ensure that the correct localization option is selected next to the Option field on the Software Installation Utility screen. See Table B-7, “Software Localization Options,” on page B-8 for a complete listing of all available localization options. Table B-7 Software...
Page 173
Installing PageWriter Trim II, III, and Rx SoftwareInstalling the Software Upgrade PageWriter Trim Cardiograph Service ManualB-9 11Press the Ta b key to highlight the Install Operation drop-down menu or click on the menu with the mouse. ABG Australia English AAMI A4 ABH Netherlands Dutch IEC A4 ABM Latin American Spanish AAMI A ABN Norway Norwegian IEC A4 ABS Sweden Swedish IEC A4 ABU UK English IEC A4 ABX Finland Finnish IEC A4 ABZ Italy Italian IEC A4 AC4 Brazil Portuguese AAMI A AC6 South Korea...
Page 174
Installing PageWriter Trim II, III, and Rx SoftwareVerifying the Software Upgrade B-10PageWriter Trim Cardiograph Service Manual 12Press the Ta b key to highlight the Install Operations menu, or click on the menu with the mouse. 13For software version A.01.01 only: press the down arrow key to select Kernel. CAUTIONDo not select Kernel if installing software version A.01.03. Go to step 16. 14Click on the Start Installation button or press Ta b to highlight the button and then press the Enter key ( )....
Page 175
Installing PageWriter Trim II, III, and Rx SoftwareVerifying the Software Upgrade PageWriter Trim Cardiograph Service ManualB-11 Figure B-20 The Software Identification Screen 3 As soon as the software identification screen appears, press the left Ctrl + Shift+ Trim Knob at the same time. 4The Access Code window appears. If it does not appear, press the left Ctrl + Shift+ Trim Knob sequence again. 5Type in 1573 and then press the Enter key. 6The Service Utility screen appears. Under Diagnostic Tests...
Page 176
Installing PageWriter Trim II, III, and Rx SoftwareVerifying the Software Upgrade B-12PageWriter Trim Cardiograph Service Manual 9For the Screen Test, a blank screen appears with the buttons Color and Pattern at the bottom of the screen. Click both buttons to display the test image. NOTEThe color gradations in the image on the cardiograph display will be more gradual and smoother than those shown in Figure B-22. Figure B-22 Screen Test Image 10 Look for the following details in the test image appearing...
Page 177
Installing PageWriter Trim II, III, and Rx SoftwareVerifying the Software Upgrade PageWriter Trim Cardiograph Service ManualB-13 the printer test. If the printer test page does not match the printer test points, the cardiograph has failed the print test. Table B-4 Printer Test Page Description Test Point Description A The stepped bars are sharp edged and printed cleanly without distortion or missing segments B The spacing between the vertical lines is 25 mm with a discrepancy of no more or less than...
Page 178
Installing PageWriter Trim II, III, and Rx SoftwareVerifying the Software Upgrade B-14PageWriter Trim Cardiograph Service Manual Figure B-23 Printer Test Page 14 Click the OK button if the cardiograph passed the printer test. 15When the Self Test is complete, the cardiograph prints out a Self Test Report. On the printed test report, check the software revision level next to Release Version (lower right side of report). The revision level should read A.01.01 (Figure B-24 on page B-15) or A.01.03...
Page 179
Installing PageWriter Trim II, III, and Rx SoftwareVerifying the Software Upgrade PageWriter Trim Cardiograph Service ManualB-15 Figure B-24 Software Version A.01.01 Self Test Report A ASoftware Revision Level and Localization Option Information
Page 180
Installing PageWriter Trim II, III, and Rx SoftwareVerifying the Software Upgrade B-16PageWriter Trim Cardiograph Service Manual Figure B-25 Software Version A.01.03 Self Test Report If an incorrect software revision level or incorrect localization option appears on the Self Test Report, reinstall the software following the same procedure. If an incorrect revision level or localization option appears again on the report, or if the cardiograph fails any portion of the Self Test, contact the nearest...