Page 191

Wireless LAN Installation InstructionsInstalling the Wireless LAN Card
PageWriter Trim Cardiograph Service ManualC-9
13Click the OK button (upper right corner of screen) to close the Aironet Client Utility. The 
PageWriter Trim Wireless LAN Installation screen appears. Click the 
Reset Cardiograph 
button. The cardiograph automatically restarts. 
14
When the R/T (real-time) ECG screen appears, press the Ta b key ( ) to highlight the 
Config button on the Command Toolbar. Press the space bar to select the...

Page 192

Wireless LAN Installation InstructionsEnabling LEAP Credentials
C-10PageWriter Trim Cardiograph Service Manual
Enabling LEAP Credentials
Follow this procedure after entering the user name, user password, and user domain settings 
used with LEAP in the Aironet Client Utility (ACU). For information on entering settings in 
the ACU, see “Installing the Wireless LAN Card” on page C-3. The following procedure must 
be completed in order to enable LEAP credentials used with LEAP authentication. This 
procedure...

Page 193

Wireless LAN Installation InstructionsConfiguring a TraceMasterVue Remote Site
PageWriter Trim Cardiograph Service ManualC-11
information on registering to use the InCenter site, see “Using the Philips InCenter Site” on 
page 1-26. The guide is updated on a regular basis, check for updates at the Philips InCenter 
web site. 
To configure a Remote Site with the wireless network connection:
1
Press the Ta b key ( ) on the keyboard to highlight the Config button. Press the space 
bar to select the button....

Page 194

Wireless LAN Installation InstructionsConfiguring a TraceMasterVue Remote Site
C-12PageWriter Trim Cardiograph Service Manual
14Press the space bar. The Add New Remote Site window appears.Type in a name for this 
Remote Site. Press the Ta b key to highlight the 
OK button. Press the space bar to select the 
button. 
The Remote Site appears in the 
Remote Sites Archive drop-down list (top of screen) and 
may be selected in the Archive. Press the Enter key (on keyboard) to save the changes and 
to exit the...

Page 195

D
D-1
Appendix AAssembling the Cardiograph Cart and
Patient Cable Arm
This Appendix includes information on how to assemble the optional cardiograph cart, and 
optional patient cable arm for the cardiograph cart. For information on ordering the 
cardiograph cart or patient cable arm, see “Upgrades and Accessories” on page 1-18. 
Figure D-1 The PageWriter Trim Cardiograph Cart
1 

Page 196

Assembling the Cardiograph Cart and Patient Cable ArmAssembling the Cart
D-2PageWriter Trim Cardiograph Service Manual
Assembling the Cart 
Follow this procedure to assemble the cardiograph cart.
To assemble the cart:
1
Insert the beam into the cart base. 
2Hold the beam steadily. Turn the cart onto the side to expose the bottom of the cart.
3Place the ground strap onto the screw end. 

Page 197

Assembling the Cardiograph Cart and Patient Cable ArmAssembling the Cart
PageWriter Trim Cardiograph Service ManualD-3
4Attach the bolt screws and tighten using the provided wrench.
5Turn the cart upright.
6Attach the bin to the cart beam.
7Insert the bin dividers. 

Page 198

Assembling the Cardiograph Cart and Patient Cable ArmAttaching the Cardiograph to the Cart
D-4PageWriter Trim Cardiograph Service Manual
8Pull the tray out from the top shelf. Attach the top shelf to the beam using the provided 
bolts. Tighten the bolts.
Attaching the Cardiograph to the Cart 
CAUTIONFollow the procedure below to ensure that the cardiograph is securely fastened to the cart before use.
To attach the cardiograph to the cart:
1
Align the front feet of the cardiograph with the front locking...

Page 199

Assembling the Cardiograph Cart and Patient Cable ArmUsing the Cart Wheel Positioners and Brake
PageWriter Trim Cardiograph Service ManualD-5
2Insert the screws through the bottom of the cart and through the screw holes. Tighten the 
screws with a Phillips head screwdriver.
Using the Cart Wheel Positioners and Brake
The cart includes one wheel brake and two wheel positioners. Lock the wheel positioners at all 
times when using the cart. The wheel positioners keep the cart straight when moving forward 
or...

Page 200

Assembling the Cardiograph Cart and Patient Cable ArmConnecting the Patient Data Cable Bracket
D-6PageWriter Trim Cardiograph Service Manual
2Step on the gray rear wheel brake to lock the cart wheels. The cart will not move. Step on 
the wheel brake again to unlock the wheels.
Connecting the Patient Data Cable Bracket 
To connect the patient data cable bracket:
1
Insert patient data cable into the bracket.
2Attach the patient data cable bracket to the mounting hole located on the rear of the 
cardiograph...