Page 161

Installing PageWriter Trim I SoftwareVerifying the Software Installation
PageWriter Trim Cardiograph Service ManualA-13
Figure A-11 Software Version A.01.03 Self Test Report
If an incorrect software revision level or incorrect localization option appears on the Self Test 
Report, reinstall the software following the same procedure. If an incorrect revision level or 
localization option appears again on the report, or if the cardiograph fails any portion of the 
Self Test, contact the nearest Philips...

Page 162

Installing PageWriter Trim I SoftwareSpecial Note for Software Version A.01.01
A-14PageWriter Trim Cardiograph Service Manual
Special Note for Software Version A.01.01
The following section provides additional information about software version A.01.01.
Special Note about Software Version A.01.01 Self Test 
Report
Figure A-12 Software Version A.01.01 Self Test Report with Failed Version Test
If the cardiograph fails the Self Test, check the Version Test field under Self-Test Results 
(upper right side of...

Page 163

Installing PageWriter Trim I SoftwareSpecial Note for Software Version A.01.03
PageWriter Trim Cardiograph Service ManualA-15
Special Note for Software Version A.01.03 
NOTEAlways ensure that the Patient Interface Module (PIM) is connected to the cardiograph before pressing 
the On/Off button and returning the cardiograph to active use. 
Figure A-13 Connecting the PIM to the cardiograph
Please note that the following sequence of events may occur if the PIM is not connected to the 
cardiograph, or a...

Page 164

Installing PageWriter Trim I SoftwareSpecial Note about PIM Repairs
A-16PageWriter Trim Cardiograph Service Manual 

Page 165

B
B-1
Appendix CInstalling PageWriter Trim II, III, and
Rx Software
A software installation procedure takes approximately twenty minutes to complete.
Software Upgrades
The upgrade path for a specific cardiograph is dependent upon the software version that is 
currently installed on the unit. Philips recommends that all cardiographs be upgraded to 
software version A.01.03 for optimal performance. See Table B-5 for more information for 
the appropriate upgrade path. 
Upgrading Legacy Software
Legacy...

Page 166

Installing PageWriter Trim II, III, and Rx SoftwareSaving Custom Settings (Software version A.00.03 and lower only)
B-2PageWriter Trim Cardiograph Service Manual
Special Note about Japanese Localization Option
The Japanese AR0 option for the PageWriter Trim II and III model cardiographs is no longer 
available. The supported software revision level for all Japanese option cardiographs is legacy 
software version A.00.03. There is no software upgrade path for the Japanese localization 
option.
Table  B-6...

Page 167

Installing PageWriter Trim II, III, and Rx SoftwareSaving Custom Settings (Software version A.00.03 and lower only)
PageWriter Trim Cardiograph Service ManualB-3
4Press and hold the Alt key on the keyboard, and then press the N key to select the 
Network tab.
All configured cardiograph Network settings appear on the screen. Record all network 
settings as indicated in the following steps. 
5Under Network ID (right side of screen) write down the Computer Name exactly as it 
appears in the field. 
NOTEThe...

Page 168

Installing PageWriter Trim II, III, and Rx SoftwareObtaining Software
B-4PageWriter Trim Cardiograph Service Manual
Obtaining Software 
The software installation procedure requires the use of a computer with a compatible PCMCIA 
card slot, and internet access. The procedure also requires the use of a blank CompactFlash 
(CF) memory card and a PC card adapter. The memory card and adapter, referred to as a PC 
card, may be purchased from Philips using part number 453563479701. The following 
procedure...

Page 169

Installing PageWriter Trim II, III, and Rx SoftwareObtaining Software
PageWriter Trim Cardiograph Service ManualB-5
NOTESThe US English and European software link includes all supported European languages.
The Documentation link contains all user documentation (instructions for use, quick reference 
cards) for the selected software version. 
8
The InCenter software download license agreement appears. If you agree to the terms of 
the agreement, click 
I agree. 
NOTEA message may appear that the Download...

Page 170

Installing PageWriter Trim II, III, and Rx SoftwareInstalling the Software Upgrade
B-6PageWriter Trim Cardiograph Service Manual
Installing the Software Upgrade 
For ease of navigation on the installation screens, connect a USB mouse to the USB connector 
on the rear of the cardiograph. If a USB mouse is not available, use the Tab key and the Enter 
key on the keyboard to navigate on the software installation screens.
To install the software upgrade:
1
Connect a USB mouse to the USB connector ( ) on the...