Page 111
TroubleshootingTroubleshooting Cardiograph Issues PageWriter Trim Cardiograph Service Manual5-13 One or more leads show leads off (red dashed line) periodicallyPIM communications have been lost due to USB patient data cable connection failure, or defective leads are connected to the PIM. 1Check that PIM patient data cable is firmly seated to the PIM connector ( ) on the rear of the cardiograph. 2If the PIM is firmly attached, perform the PIM test as described in “Maintenance Tests (Trim...
Page 112
TroubleshootingTroubleshooting Cardiograph Issues 5-14PageWriter Trim Cardiograph Service Manual Real Time Screen Issues One or more leads show leads off (red dashed line) periodically6If replacing the lead wires does not resolve signal quality issues, check the PIM Kernel Revision field in the Service Utility (see page 4-7). – If the PIM Kernel version is G.001, upgrade the cardiograph software to version A.01.03. If the signal quality issues persist with software version A.01.03 installed,...
Page 113
TroubleshootingTroubleshooting Cardiograph Issues PageWriter Trim Cardiograph Service Manual5-15 Archive Screen Issues Table 5-5 Archive Screen Issues SymptomPossible Cause & Investigation StepConfirmation Solution After attempting an Archive screen operation, an error message appears: Command timed-out The TraceMaster ECG Management system is busy. Action: Retry TraceMaster operation. Operation succeeds on subsequent try. Transfer operations use network and server resource, which may...
Page 114
TroubleshootingTroubleshooting Cardiograph Issues 5-16PageWriter Trim Cardiograph Service Manual When an error occurs during transfer of multiple ECGs to a TraceMaster system, the transfer process does not continue and transferred files are not deleted. Application limitation Action: None. Current system limitation. The current PageWriter Trim operations fail in a known safe manner when multiple ECG transfer operations are interrupted. Retry the transfer for the remaining ECGs. After...
Page 115
TroubleshootingTroubleshooting Cardiograph Issues PageWriter Trim Cardiograph Service Manual5-17 Configuration Screen Issues Some ECGs retrieved from TraceMaster and printed on the cardiograph look different from TraceMaster prints or XLi printouts. Older or non-PageWriter Trim source ECG files were retrieved. Action: Inspect the printed report date and information ECG was not created by a PageWriter Trim cardiograph.Retrieving and printing older ECG files that were not sourced by the...
Page 116
TroubleshootingTroubleshooting Cardiograph Issues 5-18PageWriter Trim Cardiograph Service Manual Printer Issues Table 5-7 PageWriter Trim I Printer Issues Software Version A.00.03 and lower Error Code Possible Cause & Investigation Step Solution 0201 No battery or dead battery.1Replace battery. 2Replace motherboard, contact Philips Medical Systems for further assistance (page 1-32). 3Replace power supply cord. 0202UPR running: a new print job has been sent to the printer and the printer cannot...
Page 117
TroubleshootingTroubleshooting Cardiograph Issues PageWriter Trim Cardiograph Service Manual5-19 0206 Temperature shutdown: the temperature of the print head is too high1Shut down cardiograph and wait 20 minutes. 2If error persists, send cardiograph for repair, contact Philips Medical Systems for further assistance (page 1-32). 0207Unknown error 1Shut down cardiograph 2Remove battery 3Reinsert battery and press On/ Standby button 4If error persists, contact Philips Medical Systems for further...
Page 118
TroubleshootingTroubleshooting Cardiograph Issues 5-20PageWriter Trim Cardiograph Service Manual 0204 USB printer - paper jam or out of paper1Open and firmly close the printer drawer. 2Insert paper into drawer, if necessary. 3Open the printer drawer and check to make sure that the paper is properly inserted and aligned. 4Replace TOF sensor assembly, contact Philips Medical Systems for further assistance (page 1-32). 5Replace paper tray assembly. 0207No PIM or defective PIM is connected to...
Page 119
TroubleshootingTroubleshooting Cardiograph Issues PageWriter Trim Cardiograph Service Manual5-21 After the paper drawer was opened during printing, the print image was corrupted or incorrect and the paper does not stop feeding.Printer controller error Action: Press the rear restart button (see page 5-33). The cardiograph restarts and boots correctly. Paper stops feeding.Occasional printer controller status errors can occur when opening the paper drawer during printing. The message Printer...
Page 120
TroubleshootingTroubleshooting Cardiograph Issues 5-22PageWriter Trim Cardiograph Service Manual Printer Error: Out of Paper/Paper Jam/ Door Open„Paper drawer is not fully closed „Out of paper, or paper loaded incorrectly „Defective TOF sensor assembly „Defective paper tray assembly Error does not occur again.„Open and firmly close the paper drawer. „Open paper drawer and check to make sure paper is aligned correctly. „Replace TOF sensor assembly, contact Philips Medical Systems for further...