Page 91
Maintenance TestsTrim I Diagnostic and Performance Verification Tests PageWriter Trim Cardiograph Service Manual4-17 When the self-tests are completed, a status report is printed automatically. Figure 4-11 on page 4-17 shows a sample report. Figure 4-11 Self-Test Report Trim I Diagnostic and Performance Verification Tests Diagnostic tests are included in the Biomed Service Program and are used to identify and to troubleshoot a technical problem with the PageWriter Trim I cardiograph. CAUTIONEnsure...
Page 92
Maintenance TestsTrim I Diagnostic and Performance Verification Tests 4-18PageWriter Trim Cardiograph Service Manual 3Quickly, while the screen showing five triangles is displayed, press the Patient ID and Auto buttons at the same time. NOTEThe screen showing five triangles disappears after five seconds. If you miss it, you must restart the cardiograph again. An Access Code window appears. If the Access Code window does not appear, restart the cardiograph. Follow steps 1 to 3 again. 4Enter the access...
Page 93
Maintenance TestsTrim I Diagnostic and Performance Verification Tests PageWriter Trim Cardiograph Service Manual4-19 Battery Test This test analyzes if the battery voltage is below or above the voltage limits. To perform the keyboard test: 1 From the Biomed Service Program, turn the Trim Knob to highlight Test and push to select it. 2Turn the Trim Knob to highlight Test Battery and push to select it. 3Turn the Trim Knob to highlight Start and push to select it. The test results display on the...
Page 94
Maintenance TestsTrim I Diagnostic and Performance Verification Tests 4-20PageWriter Trim Cardiograph Service Manual 5If you pressed Patient ID to pass the test, now press Patient ID to print a report or Stop to return to the previous screen. If the screen test fails, try it again. Onboard Flash Test Verifies that the Onboard Flash is working properly. To test the Onboard Flash: 1 From the Biomed Service Program, turn the Trim Knob to highlight Test and push to select it. 2Turn the Trim Knob to...
Page 95
Maintenance TestsTrim I Diagnostic and Performance Verification Tests PageWriter Trim Cardiograph Service Manual4-21 3Turn the Trim Knob to highlight Start and push to select it. 4The test results display on the screen. Press Patient ID to print a report or Stop to return to the previous screen. If the PIM test fails, try it again. If the failure persists, contact the nearest Philips Response Center for further assistance, see “Contacting a Philips Response Center” on page 1-32. Printer Test To...
Page 96
Maintenance TestsTrim I Diagnostic and Performance Verification Tests 4-22PageWriter Trim Cardiograph Service Manual Self-Test The Self-Test option will run all of the tests listed in Table 4-37. For information on any specific test, see the page number referenced in the table. To run the self-test: 1 Select Test from the Main Menu. 2Select Self-Test and then select Start to start the self-test program. 3At the conclusion of the Self-Test, press the Patient ID button to print out a report of all test...
Page 97
Maintenance TestsTrim I Diagnostic and Performance Verification Tests PageWriter Trim Cardiograph Service Manual4-23 4The test results display on the screen. Press Patient ID to print a report or Stop to return to the previous screen. NOTEThe Version Test will fail if the PIM is not connected to the cardiograph. Voltage Test The Voltage Test verifies that the cardiograph is operating within acceptable voltage limits. To perform the Voltage Test: 1 From the Biomed Service Program, turn the Trim Knob to...
Page 98
Maintenance TestsTrim I Diagnostic and Performance Verification Tests 4-24PageWriter Trim Cardiograph Service Manual
Page 99
5 5-1 Chapter 5 Troubleshooting This chapter provides information for localizing cardiograph problems to the subassembly level. This information is designed for use with the cardiograph’s diagnostic self-tests to help you efficiently repair the cardiograph with a minimum of equipment. Contacting a Philips Response Center The Philips Response Center can assist with product troubleshooting and provide technical expertise to help with any issue with the PageWriter Trim cardiograph or any of its...
Page 100
TroubleshootingPower On and Power Off Sequence 5-2PageWriter Trim Cardiograph Service Manual Figure 5-14 Trim I Power On Sequence