Page 101
TroubleshootingPower On and Power Off Sequence PageWriter Trim Cardiograph Service Manual5-3 Figure 5-15 Trim II/III/Rx Power On Sequence
Page 102
TroubleshootingPower On and Power Off Sequence 5-4PageWriter Trim Cardiograph Service Manual Figure 5-16 Trim I Power Off Sequence Figure 5-17 Trim II/III/Rx Power Off Sequence
Page 103
TroubleshootingPower On and Power Off Sequence PageWriter Trim Cardiograph Service Manual5-5 Special Note About Software Version A.01.03 Please note that the power up sequence for software version A.01.03 may differ from the process described in Figure 5-14 and in Figure 5-15. There are circumstances where the power up sequence can be delayed by up to a maximum of 50 seconds. This can occur with installed software version A.01.03 when the PIM is not connected to the cardiograph (or a defective PIM...
Page 104
TroubleshootingTroubleshooting Cardiograph Issues 5-6PageWriter Trim Cardiograph Service Manual Troubleshooting Cardiograph Issues The troubleshooting information in this section will help you to localize a fault and to correct it. The following charts list symptoms and the steps to follow to investigate and solve the problem. Several failure mode symptoms may actually be hardware induced or triggered, and may require board replacement or opening of the cardiograph for inspection, please contact the...
Page 105
TroubleshootingTroubleshooting Cardiograph Issues PageWriter Trim Cardiograph Service Manual5-7 Screen is dark and shows no imageLCD panel failure Replace with a new LCD panel to see if it functions normally. Replace with a new LCD panel. Contact Philips Medical Systems (see page 1-32). LCD backlight board failure. Replace with a new backlight board to see if the LCD functions normally.Replace the LCD backlight board. Contact Philips Medical Systems (see page 1-32). Main controller board...
Page 106
TroubleshootingTroubleshooting Cardiograph Issues 5-8PageWriter Trim Cardiograph Service Manual Keyboard/Trim Knob/Dedicated Key Issues Table 5-2 Keyboard/Trim Knob/Dedicated Key Issues SymptomPossible Cause & Investigation StepConfirmation Solution Keyboard is unresponsiveKeyboard controller subsystem has failed. Action: restart cardiograph using rear button (see page 5-33) and launch the Service Utility. Perform the Keyboard Test.Unable to enter Service mode after restart. Keyboard is still...
Page 107
TroubleshootingTroubleshooting Cardiograph Issues PageWriter Trim Cardiograph Service Manual5-9 Keyboard is typing incorrect characters Keyboard controller subsystem has failed. Action: restart cardiograph using rear button and launch Service Utility (see page 4-7). Run the Keyboard Test. Keyboard is now functional. Occasional recoverable keyboard failures have been observed. If failures persist, cardiograph must be serviced to inspect and possibly replace keyboard, cabling, or controller...
Page 108
TroubleshootingTroubleshooting Cardiograph Issues 5-10PageWriter Trim Cardiograph Service Manual Signal Acquisition Issues Before attempting to troubleshoot a signal acquisition error as a mechanical issue, ensure that the following patient preparation and basic troubleshooting techniques have first been followed. Always follow good skin and patient preparation techniques prior to taking an ECG. Proper skin preparation helps to ensure a good quality ECG. Skin is a poor conductor of electricity and...
Page 109
TroubleshootingTroubleshooting Cardiograph Issues PageWriter Trim Cardiograph Service Manual5-11 After ensuring that proper patient preparation techniques are being followed, and that fresh electrodes are being applied to the patient, review the following table for further assistance. Table 5-3 Signal Acquisition Issues SymptomPossible Cause & Investigation StepSolution All leads show leads off (red dashed line) continuouslyPIM communications have been lost due to USB patient data cable...
Page 110
TroubleshootingTroubleshooting Cardiograph Issues 5-12PageWriter Trim Cardiograph Service Manual One or more leads show leads off (red dashed line) periodicallyImproper or incomplete patient and skin preparation techniquesEnsure that good patient preparation and skin preparation techniques have been followed as described on page 5-10. Improperly attached electrodes or lead wires; expired disposable gel electrodesEnsure that the lead adapters are clean and have a tight fit. Ensure that the...