Page 71
Cardiograph Care and MaintenanceCalibrating the Barcode Reader PageWriter Trim Cardiograph Service Manual3-11 Calibrating the Barcode Reader The optional barcode reader is shipped with configured settings that provide optimal use with the cardiograph. If the barcode reader operates with errors or missed data, follow the procedure below. If the barcode reader incorrectly scans data into fields on the Patient ID screen, follow the procedure “Removing the Carriage Return” on page 3-13. The barcodes must...
Page 72
Cardiograph Care and MaintenanceCalibrating the Barcode Reader 3-12PageWriter Trim Cardiograph Service Manual Figure 3-2 Barcode Calibration Sequence 1 2 3 4
Page 73
Cardiograph Care and MaintenanceCalibrating the Barcode Reader PageWriter Trim Cardiograph Service Manual3-13 Removing the Carriage Return Scan the following barcode when the barcode reader incorrectly scans data in any field on the Patient ID screen. The following procedure removes a carriage return from the end of a data scan that can cause errors when entering Patient ID information. 6 7 8 5
Page 74
Cardiograph Care and MaintenanceMaintenance Tests for Trim II, III and Rx 3-14PageWriter Trim Cardiograph Service Manual To calibrate the barcode reader: 1 Hold the barcode reader at a 45o angle and push the button (top of scanner). Scan the barcode in Figure 3-3. Three beeps are heard. Figure 3-3 Barcode scan to remove the automatic carriage return 2 Perform the Barcode Test (see page 4-2) to verify performance. Maintenance Tests for Trim II, III and Rx Use the diagnostic utilities and maintenance...
Page 75
4 4-1 Appendix DMaintenance Tests This appendix provides instructions for performing maintenance tests on the PageWriter Trim II/III/Rx cardiograph from the normal cardiograph operating mode, and includes instructions for using the Biomed Service Utility. Using the Biomed Service Utility requires a full restart of the cardiograph. This appendix also includes instructions on using the PageWriter Trim I cardiograph Biomed Service Utility. The tests contained in this appendix help you to identify...
Page 76
Maintenance TestsMaintenance Tests (Trim II/III/Rx only) 4-2PageWriter Trim Cardiograph Service Manual Patient Interface Module (PIM) Test This test is used to confirm that the Patient Interface Module (PIM) is communicating with the cardiograph. This test can be performed when the cardiograph displays PIM error messages, or when the cardiograph is unable to acquire data from the PIM. Be sure that the PIM patient data cable is securely attached to the PIM connector on the rear of the cardiograph...
Page 77
Maintenance TestsMaintenance Tests (Trim II/III/Rx only) PageWriter Trim Cardiograph Service Manual4-3 Magnetic Card Reader Test NOTEYou must attach the magnetic card reader to the port on the rear of the cardiograph before powering on the unit. The Magnetic Card Reader Test is used with the optional magnetic card reader. This test can be used to confirm that the magnetic card reader is correctly reading data from a magnetic card. Ensure that the magnetic card reader is securely attached to the...
Page 78
Maintenance TestsMaintenance Tests (Trim II/III/Rx only) 4-4PageWriter Trim Cardiograph Service Manual Test PointDescription C The diagonal lines should be straight and printed cleanly without distortion or breaks in the lines D The character set is printed cleanly without distortion or missing characters, and all characters are clearly legible Table 4-35 Printer Test Page Description (continued)
Page 79
Maintenance TestsMaintenance Tests (Trim II/III/Rx only) PageWriter Trim Cardiograph Service Manual4-5 Figure 4-4 Printer Test Page 4 If the printer test page matches the printer test points as described in Table 4-35, “Printer Test Page Description,” on page 4-3, the cardiograph has passed the printer test. If the printer test page does not match the printer test points, the cardiograph has failed the print test. AA B C D
Page 80
Maintenance TestsTrim II/III/Rx Diagnostic and Performance Verification Tests 4-6PageWriter Trim Cardiograph Service Manual Network Test The Network Test displays the configured network information for the cardiograph. All cardiograph network configuration is completed on the Network screen. From any Configuration screen, press and hold down the Alt key on the keyboard, and then press the N key to select the Network tab and to open the Network screen. Before performing the Network test, ensure that...