Page 211

Assembling the Cardiograph Cart and Patient Cable ArmAssembling the Patient Cable Arm
PageWriter Trim Cardiograph Service ManualD-17
18Loosen the elbow hinge after fully assembling the patient cable arm. 
CAUTIONLoosen hinge carefully. Patient cable arm will open suddenly. 

Page 212

Assembling the Cardiograph Cart and Patient Cable ArmAssembling the Patient Cable Arm
D-18PageWriter Trim Cardiograph Service Manual 

Page 213

E
E-1
Appendix AUpgrade Kits
The following upgrade kits are available for the PageWriter Trim II, III, and Rx model 
cardiographs:
„Interpretation
„Wireless LAN
„Ethernet LAN
For more information on ordering the upgrade kits, see “Upgrades and Accessories” on 
page 1-18.
NOTEThere are no upgrades available for the PageWriter Trim I cardiograph.
Upgrade Kit Contents 
Each upgrade kit includes the following items as described in Table E-1.
Table E-1 Upgrade Kit Contents
Interpretation 
Upgrade KitWireless...

Page 214

Upgrade KitsUpgrade Kit Installation
E-2PageWriter Trim Cardiograph Repair Manual
Figure E-1 The Upgrade Kit Contents
Upgrade Kit Installation
Each upgrade kit is provided with a set of instructions that describe how to enter the token 
number information. Entering the token number activates the software option on the 
cardiograph. The following instructions must be completed before configuring the specific 
features of the upgrade option. For more information on token numbers and where to affix the...

Page 215

Upgrade KitsUpgrade Kit Installation
PageWriter Trim Cardiograph Repair ManualE-3
Figure E-2 The Please Input Token window
7
Enter the token number in the Token field.
8Press the Ta b  key or turn the Trim Knob to highlight the Enable Token button. 
9Press the space bar or the Trim Knob to select the button. The option is enabled.
10Remove the paper drawer. Locate the existing token label inside the paper drawer 
compartment. Remove the existing label. Affix the new token label to the same location. 
For...

Page 216

Upgrade KitsUpgrade Kit Installation
E-4PageWriter Trim Cardiograph Repair Manual 

Page 217

F
F-1
Appendix ASpecifications
Technical Specifications
ECG Acquisition
PageWriter Trim I
„Auto (12 leads)
„Rhythm (3 or 6 leads)
PageWriter Trim II/III/Rx
„Auto (12 leads)
„Rhythm (up to 12 leads)
„Disclose (1 lead)
Keyboard
PageWriter Trim II/III/Rx
„Full alphanumeric capability
Screen Display 
PageWriter Trim I
„40 character LCD display
PageWriter Trim II/III/Rx
„640 x 480 pixel resolution
„Color TFT display
Patient Interface Module 
„Remote, microprocessor-controlled module 

Page 218

SpecificationsTechnical Specifications
F-2PageWriter Trim Cardiograph Service Manual
Cardiograph Cart
„The cardiograph cart (optional accessory) has been tested for stability using the IEC 
60601-1 test protocol. 
„The IEC 60601-1 protocol calls for the cardiograph cart to not tip over using a 10 degree 
angle of incline in any direction. It is possible to exceed this amount of incline when the 
cardiograph and cart are moved over a roadside curb or some other small but steep incline. 
Use care when...

Page 219

SpecificationsTechnical Specifications
PageWriter Trim Cardiograph Service ManualF-3
Report Formats
PageWriter Trim I
„3x4, 1R
„6x2 (Cabrera lead sequence only)
„3 and 6 lead Rhythm report
PageWriter Trim II/III/Rx
„3x4 1R, 3R (Standard, Cabrera)
„6x2 (Standard, Cabrera)
„Panoramic 12 (Cabrera)
„12x1(Standard, Cabrera)
„Rhythm (up to 12 selected leads)
„12-Lead Extended Measurements
„Disclose (1 minute of continuous waveform data for 1 selected lead)
Battery Operation
Capacity 
„Typically 30 ECGs on a...

Page 220

SpecificationsTechnical Specifications
F-4PageWriter Trim Cardiograph Service Manual
Modem (optional for USA and Canada)
PageWriter Trim II/III/Rx
„V.90, K56flex, enhanced V.34, V.32bis, V.32, V.22 bis and below
Barcode Reader (optional)
PageWriter Trim II/III/Rx
„Reads Code 39 (standard and full ASCII) Symbology
Magnetic Card Reader (optional)
PageWriter Trim II/III/Rx
„Reads cards adhering to ISO 7810, 7811-1, -2, -3, -4, -5, and JIS X 6301 and X6302
ECG Storage 
„150 ECGs typical for the optional...
Start reading Philips Trim 1 Service Manual