Philips Trim 1 Service Manual
Here you can view all the pages of manual Philips Trim 1 Service Manual. The Philips manuals for Cardiograph are available online for free. You can easily download all the documents as PDF.
Page 221
SpecificationsTechnical Specifications PageWriter Trim Cardiograph Service ManualF-5 Environmental Operating Conditions „10° to 40°C (50° to 104°F) „15% to 80% relative humidity (non-condensing) „Up to 4,550 m (15,000 ft.) altitude Environmental Storage Conditions „0° to 50 °C (32° to 122°F) „15% to 90% relative humidity (non-condensing) „Up to 4,550 m (15,000 ft.) altitude Cardiograph Dimensions PageWriter Trim I „310 mm width x 388 mm depth x 106 mm height (12.2 x 15.3 x 4.2 inches) PageWriter Trim...
Page 222
SpecificationsSafety and Performance F-6PageWriter Trim Cardiograph Service Manual Cardiograph Cart Weight „20.32 kilograms (44.8 pounds) with single bin and counterweight installed Fully Assembled Cardiograph Cart Shipping Container Dimensions „547 mm (height) x 827 mm (width) x 1107 mm (length) (21.5 x 32.5 x 43.5 inches) Partially Assembled Cardiograph Cart Shipping Container Dimensions „229 mm (height) x 534 mm (width) x 1234 mm (length) (9 x 21 x 48.5 inches) Fully Fully Assembled Cardiograph...
Page 223
SpecificationsElectromagnetic Compatibility (EMC) PageWriter Trim Cardiograph Service ManualF-7 cardiograph according to the international standard for EMC for medical devices (IEC 60601- 1-2). This IEC standard has been adopted in Europe as the European Norm (EN 60601-1-2). The PageWriter Trim cardiograph should not be used adjacent to, or stacked on top of other equipment. If the PageWriter Trim cardiograph must be used adjacent to or stacked on top of other equipment, verify that the cardiograph...
Page 224
SpecificationsElectromagnetic Compatibility (EMC) F-8PageWriter Trim Cardiograph Service Manual RF Emissions CISPR 11Class B The PageWriter Trim cardiograph is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. Harmonic Emissions IEC 61000-3-2Class A Voltage fluctuations/ flicker emissions IEC 61000-3-3Complies Table F-1 Guidance and Manufacturer’s...
Page 225
SpecificationsElectromagnetic Compatibility (EMC) PageWriter Trim Cardiograph Service ManualF-9 Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-1195% dip in UT) for 0.5 cycle 40% U T (60% dip in UT) for 5 cycles 70% U T (>30% dip in UT) for 25 cycles 95% dip in UT) for 5 secondsComplies Mains power quality should be that of a typical commercial or hospital environment. If the PageWriter Trim cardiograph requires continuous operation during power...
Page 226
SpecificationsElectromagnetic Compatibility (EMC) F-10PageWriter Trim Cardiograph Service Manual Table F-3 Guidance and Manufacturer’s Declaration: Electromagentic Immunity The PageWriter Trim cardiograph is intended for use in the electromagnetic environment specified below. The customer or the user of the PageWriter Trim cardiograph should assure that it is used in such an environment. Emissions TestIEC 60601 Test LevelCompliance LevelElectromagnetic Environment: guidance Portable and mobile RF...
Page 227
SpecificationsElectromagnetic Compatibility (EMC) PageWriter Trim Cardiograph Service ManualF-11 a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, am- ateur radios, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the elec- tromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the...
Page 228
SpecificationsElectromagnetic Compatibility (EMC) F-12PageWriter Trim Cardiograph Service Manual Table F-4 Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the PageWriter Trim Cardiograph: for equipment and systems that are not life-supporting The PageWriter Trim cardiograph is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the PageWriter Trim cardiograph can help to prevent...
Page 229
Index-1 Index A AC fuses remove and replace 7-3 accessories enabling E-1 Action button location on Patient Interface Module (PIM) 1-17 Aironet Client Utility (ACU) error messages 5-25 archive troubleshooting 5-15 B barcode reader calibrating 3-11 diagnostic test 4-2, 4-11 part number 1-18 removing the carriage return 3-11 battery long term storage 3-7 maintenance and care 3-6 power gauge 2-14 remove and replace 7-2 Biomed Service Utility overview 4-6 Trim I 4-17 Trim II, III, Rx 4-7 C...
Page 230
Index-2 N network diagnostic ping test 4-14 diagnostic test 4-6 enabling option E-1 O onboard flash diagnostic test 4-13 ordering replacement parts 8-1 supplies and accessories 1-20 P PageWriter Trim I Biomed Service Program 4-17 front view 1-12 installing software A-1 rear view 1-13 PageWriter Trim II/III/Rx Biomed Service Utility 4-7 front view 1-14 installing software B-1 maintenance tests 4-1 rear view 1-15 token number information 1-19 paper tray remove and replace 7-4 token label...