Page 221

SpecificationsTechnical Specifications
PageWriter Trim Cardiograph Service ManualF-5
Environmental Operating Conditions 
„10° to 40°C (50° to 104°F) 
„15% to 80% relative humidity (non-condensing)
„Up to 4,550 m (15,000 ft.) altitude
Environmental Storage Conditions
„0° to 50 °C (32° to 122°F)
„15% to 90% relative humidity (non-condensing)
„Up to 4,550 m (15,000 ft.) altitude
Cardiograph Dimensions
PageWriter Trim I
„310 mm width x 388 mm depth x 106 mm height (12.2 x 15.3 x 4.2 inches)
PageWriter Trim...

Page 222

SpecificationsSafety and Performance
F-6PageWriter Trim Cardiograph Service Manual
Cardiograph Cart Weight 
„20.32 kilograms (44.8 pounds) with single bin and counterweight installed 
Fully Assembled Cardiograph Cart 
Shipping Container Dimensions
„547 mm (height) x 827 mm (width) x 1107 mm (length) (21.5 x 32.5 x 43.5 inches)
Partially Assembled Cardiograph Cart 
Shipping Container Dimensions
„229 mm (height) x 534 mm (width) x 1234 mm (length) (9 x 21 x 48.5 inches) Fully 
Fully Assembled Cardiograph...

Page 223

SpecificationsElectromagnetic Compatibility (EMC)
PageWriter Trim Cardiograph Service ManualF-7
cardiograph according to the international standard for EMC for medical devices (IEC 60601-
1-2). This IEC standard has been adopted in Europe as the European Norm (EN 60601-1-2).
The PageWriter Trim cardiograph should not be used adjacent to, or stacked on top of other 
equipment. If the PageWriter Trim cardiograph must be used adjacent to or stacked on top of 
other equipment, verify that the cardiograph...

Page 224

SpecificationsElectromagnetic Compatibility (EMC)
F-8PageWriter Trim Cardiograph Service Manual
RF Emissions
CISPR 11Class B The PageWriter Trim cardiograph is suitable for use in all 
establishments, including domestic establishments and 
those directly connected to the public low-voltage power 
supply network that supplies buildings used for domestic 
purposes. Harmonic Emissions
IEC 61000-3-2Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3Complies
Table F-1 Guidance and Manufacturer’s...

Page 225

SpecificationsElectromagnetic Compatibility (EMC)
PageWriter Trim Cardiograph Service ManualF-9
Voltage dips, short 
interruptions and 
voltage variations 
on power supply 
input lines
IEC 61000-4-1195% dip in UT) 
for 0.5 cycle
40% U
T
(60% dip in UT) 
for 5 cycles
70% U
T
(>30% dip in UT) 
for 25 cycles
95% dip in UT) 
for 5 secondsComplies Mains power quality should be that of a 
typical commercial or hospital environment. If 
the PageWriter Trim cardiograph requires 
continuous operation during power...

Page 226

SpecificationsElectromagnetic Compatibility (EMC)
F-10PageWriter Trim Cardiograph Service Manual
Table F-3 Guidance and Manufacturer’s Declaration: Electromagentic Immunity 
The PageWriter Trim cardiograph is intended for use in the electromagnetic environment 
specified below. The customer or the user of the PageWriter Trim cardiograph should assure 
that it is used in such an environment.
Emissions TestIEC 60601 
Test LevelCompliance 
LevelElectromagnetic Environment: guidance
Portable and mobile RF...

Page 227

SpecificationsElectromagnetic Compatibility (EMC)
PageWriter Trim Cardiograph Service ManualF-11
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, am-
ateur radios, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the elec-
tromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the...

Page 228

SpecificationsElectromagnetic Compatibility (EMC)
F-12PageWriter Trim Cardiograph Service Manual
Table F-4 Recommended Separation Distances Between Portable and Mobile RF Communications 
Equipment and the PageWriter Trim Cardiograph: for equipment and systems that are 
not life-supporting 
The PageWriter Trim cardiograph is intended for use in the electromagnetic environment in which radiated 
RF disturbances are controlled. The customer or the user of the PageWriter Trim cardiograph can help to 
prevent...

Page 229

Index-1
Index
A
AC fuses
remove and replace
 7-3
accessories
enabling
 E-1
Action button
location on Patient Interface Module 
(PIM)
 1-17
Aironet Client Utility (ACU) 
error messages
 5-25
archive
troubleshooting
 5-15
B
barcode reader
calibrating
 3-11
diagnostic test
 4-2, 4-11
part number
 1-18
removing the carriage return
 3-11
battery
long term storage
 3-7
maintenance and care
 3-6
power gauge
 2-14
remove and replace
 7-2
Biomed Service Utility
overview
 4-6
Trim I
 4-17
Trim II, III, Rx
 4-7
C...

Page 230

Index-2
N
network
diagnostic ping test
 4-14
diagnostic test
 4-6
enabling option
 E-1
O
onboard flash
diagnostic test
 4-13
ordering
replacement parts
 8-1
supplies and accessories
 1-20
P
PageWriter Trim I
Biomed Service Program
 4-17
front view
 1-12
installing software
 A-1
rear view
 1-13
PageWriter Trim II/III/Rx
Biomed Service Utility
 4-7
front view
 1-14
installing software
 B-1
maintenance tests
 4-1
rear view
 1-15
token number information
 1-19
paper tray
remove and replace
 7-4
token label...