Page 61
Event reports, and images received from remote Storage SCP’s . In addition, TLS security for incoming DICOM communication may be e nabled, optional encryption enabled/disabled, and certificates imported, selected, or removed from this interface. 4.4.1.2 Remote AE Title/Presentation Address Mapping Remote AE’s may be configured on the PSC – Network/DICOM – DICOM Settings screen. The screen is organized into DICOM Devices, each specified by • The Device Type which indicates the services...
Page 62
Each Storage Server includes a set of Advanced Settings allowing for device-specific control of the data to be stored to that device. These settings include: • Compression to be used for single- frame images • Compression to be used for multi -frame images • Whether to send images in Monochrome or Color photom etric interpretations • Whether to limit export to Implicit VR Little Endian transfer syntax. Generally used only for the few devices that don’t understand any other transfer...
Page 63
o If there is no response from the SCP within this time period, the Network/Reply timer expires. The system will send a DICOM A -RELEASE Request . Idle Association Timeout o Association inactivity timeout. Additional Configuration Settings for Storage Commit Devices . There is an option provided for user if automatic retry of Commit job is required or not. Number of retry attempts before abandoning storage commitment o Maximum number of retries the s ystem would perform before...
Page 64
Parameter Configurable (Yes/No) Default Value General Parameters AE Title for all AE’s Yes AETitle Listening Port Yes 104 Station Name Yes Station Location Yes Hostname Yes MAX PDU Size Yes 32768 Max Simultaneous Associations Yes 5 ARTIM Timeout Yes 60 5 MEDIA STORAGE 5.1 IMPLEMENTATION MODEL 5.1.1 Application Data Flow Media Application Entity Export to Media Media Storage Medium Import from Media FSCFSU FSR* Figure 9 APPLICATION DATA FLO W...
Page 65
remaining unwritten space on the media is available for use. This restriction does not apply to + / - RW media. Erasing + / - RW media allows the entire disk’s space to be used. If a – R or +R is erased, it is not usable again for DICOM exports due to reliability concerns associated with DVD media. Only DVD+RW/ -RW can be erased and used again for DICOM export. 5.1.2 Functional Definition of AEs 5.1.2.1 Functio nal Definition of Media Application Entity Using “ Export to… Media” will...
Page 66
5.2.1.2 Real -World Activities 5.2.1.2.1 Activity – Export to Media The Media Application Entity acts as an FSC using the interchange option when requested to export SOP Instances from the local database to media. The contents of the export job will be written together with a corresponding DICOMDIR to media. The user can cancel an export job in the job queue. Writing in multi -se ssion format to CDs and DVDs is not supported; a ny CD or DVD with DICOM data on it will be viewed as read-...
Page 67
JPEG Baseline Lossy Compression 1.2.840.10008.1.2.4.50 US Multiframe Image Storage (Retired)* 1.2.840.10008.5.1.4.1.1.3 Explicit VR Little Endian 1.2.840.10008.1.2.1 JPEG Baseline Lossy Compression 1.2.840.10008.1.2.4.50 Secondary Capture Image Storage 1.2.840.10008.5.1.4.1.1.7 Explicit VR Little Endian JPEG Lossy Baseline JPEG Lossless Baseline 1.2.840.10008.1.2.1 1.2.840.10008.1.2.4.50 1.2.840.10008.1.2.4.70 Multi-frame True Color Secondary Capture Image Storage...
Page 68
Positron Emission Tomography Image Storage 1.2.840.10008.5.1.4.1.1.128 Explicit VR Little Endian JPEG Lossy Baseline JPEG Lossless Baseline 1.2.840.10008.1.2.1 1.2.840.10008.1.2.4.50 1.2.840.10008.1.2.4.70 X-Ray Angiographic Image Storage 1.2.840.10008.5.1.4.1.1.12.1 Explicit VR Little Endian JPEG Lossy Baseline JPEG Lossless Baseline 1.2.840.10008.1.2.1 1.2.840.10008.1.2.4.50 1.2.840.10008.1.2.4.70 Comprehensive Structured Report Storage 1.2.840.10008.5.1.4.1.1.88.33...
Page 69
TLS authentication may be used with or without TLS encryption – default is with TLS encryption. For outgoing TLS requests where EPIQ/Affiniti acts as a TLS Client, the received server RSA certificate is validated by verifying its digital signature against a certificate in the local Trusted Certificate Authorities store whose Subject matches the Issuer of the received certificate. If requested by the TLS Server, EPIQ/Affiniti will then send the certificate configured for “This System” from the...
Page 70
9 ANNEXES 9.1 IOD CONTENTS 9.1.1 Created SOP Instances Table 9.1 specifies the attributes of an Ultrasound Image transmitted b y the EPIQ/Affiniti storage application. Table 9.2 specifies the attributes of a Comprehensive Structured Reports transmitted by the EPIQ/Affiniti storage application . Please note that there are differences between which Structured Report Templates are used in each product. The following tables use a number of abbreviations. The abbreviations used in the...