Page 21

 
4.2  AE SPECIFICATIONS  
4.2.1  Storage Application Entity Specification  
4.2.1.1  SOP  Classes  
EPIQ/Affiniti  provides  Standard Extended
1 Conformance to the following SOP Classes:  
 
Table  4.1  
SOP CLASSES FOR AE STORAGE 
SOP Class Name SOP Class UID SCU SCP 
US Image Storage   1.2.840.10008.5.1.4.1.1.6.1 Yes  Yes 
US Image Storage (retired)  1.2.840.10008.5.1.4.1.1.6  Yes  Yes 
US Multiframe Image Storage  1.2.840.10008.5.1.4.1.1.3.1  Yes  Yes 
US Multiframe Image Storage (retired)...

Page 22

 
Note: Use of Retired SOP Classes only is user configurable for the system on the PSC – Network/DICOM  – DICOM 
Settings  – DICOM Storage Device –  Advanced Settings  configuration page.  All image storage will use Retired SOP 
Classes only when selected.  
4.2.1.2  Association Establishment Policy  
4.2.1.2.1  General  
The DICOM standard application context name for DICOM 3.0 is always proposed:  
Table  4.2  
DICOM APPLICATION CONTEXT  FOR AE STORAGE  
Application Context Name  1.2.840.10008.3.1.1.1...

Page 23

  
Table  4.6  
DICOM IMPLEMENTATION CLASS AND VERSION FOR AE STORAGE  
Implementation Class UID   1.3.46.670589.14.10000.105 
Implementation Version Name  EPIQ-AFFIN_1.05  
 
4.2.1.3 Association  Initiation Policy  
4.2.1.3.1  Activity  – Store Images and Structured Reports  
4.2.1.3.1.1  Description  and Sequencing of Activities  
A user may select exams or individual images from Review and request them  to be sent to multiple destinations (up to 
five ).  Reports may not be selected individually, but...

Page 24

 
StorageAEImage
Manager
1. Open Association
2. C-STORE (Images)
5. C-STORE (Structured Report)
8. N-ACTION (Storage Commitment Request for Images)
14. Close Association
end
exam
3. Close Association
12. N-EVENT-REPORT (Storage Commitment Response - Images)
4. Open Association
6. Close Association
7. Open Association
10. Close Association
11. Open Association
acquire image
store
completion
commit status updated
9. N-ACTION (Storage Commitment Request for SR)
13. N-EVENT-REPORT (Storage Commitment...

Page 25

 
4.2.1.3.1.2 Proposed Presentation Contexts   
EPIQ/Affiniti  is  capable of proposing the Presentation Contexts shown in the following table:  
Table  4.7  
PROPOSED PRESENTATION CONTEXTS FOR ACTIVITY SEND  IM AGES  
Presentation Context Table 
Abstract Syntax Transfer Syntax  
 
Role 
 
Ext. 
Neg. Name   UID  Name List  UID List 
US Image Storage  
1.2.840.10008.5.1.4.1.1.
6.1  Implicit VR Little Endian
 
Explicit VR Little Endian  
JPEG Lossy Baseline  
JPEG Lossless  Baseline 
RLE Lossless...

Page 26

 
Digital Mammography X-
Ray Image Storage  – For 
Processing   1.2.840.10008.5.1.4.1.1.
1.2.1
  Implicit VR Little Endian
 
Explicit VR Little Endian  
JPEG Lossy Baseline  
JPEG Extended 12- bit Lossy 
JPEG Lossless  Baseline 1.2.840.10008.1.2 
1.2.840.10008.1.2.1
 
1.2.840.10008.1.2.4.50  
1.2.840.10008.1.2.4.51  
1.2.840.10008.1.2.4.70  SCU 
None 
MR Image Storage 
1.2.840.10008.5.1.4.1.1.
4  Implicit VR Little Endian
 
Explicit VR Little Endian  
JPEG Lossy Baseline  
JPEG Extended  12-bit Lossy...

Page 27

 
study contains no JPEG Lossy compressed loops. The Implicit Little Endian Only selection will override the Single 
Frame Compression setting, resulting in only uncompressed export. 
Storage Commitment N -ACTION r equests are  only sent to the device that is configured as the Storage Commitment 
server .  Storage Commitment always works in conjunction with a  designated Storage SCP configured to receive 
storage of the objects to be committed.  
4.2.1.3.1.3  SOP Specific Conformance  
4.2.1.3.1.3.1...

Page 28

 
The Storage AE will request storage commitment for the configured device.  Storage Commitment is reques ted 
automatically by the system immediately after the object are stored (but not necessarily  one-to-one with each storage 
request ).   
Table  4.10  summarizes the behavior of Storage AE when receiving response status codes.  
 
Table  4.10  
STORAGE COMMITMENT N -ACTION RESPONSE STATUS HANDLING BEHAVIOR  
Service Status  
Further Meaning Error Code  Behavior 
Success  Success  0000  The system...

Page 29

 
4.2.1.3.2 Activity  – Query and Request Retrieval of Studies  
4.2.1.3.2.1  Description  and Sequencing of Activities  
The user may set a filter in Review to specify the cri teria used in querying the Query/Retrieve SCP. W hen the user 
initiates the  query, the system sends a  number of Study Root C -FIND request s, using the filter parameters set by the 
user.  The user may then select  from the query results  one or more of the studi es to initiate a Study Root C -MOVE 
re quest for the selected...

Page 30

 Figure 4a 
QUERY/RETRIEVE SEQUENCE DIAGRAM  
 
4.2.1.3.2.2  Proposed Presentation Contexts  
Table 4.13  
PROPOSED PRESENTATION CONTEXTS FOR ACTIVITY QUERY AND RETRIEVE  
Presentation Context Table  
Abstract Syntax  Transfer Syntax   
 
Role    
Ext.  
Neg.  
Name   UID  Name List  UID List 
Study Root 
Query/Retrieve (C -FIND)  1.2.840.10008.5.1.4.1.2.
2.1  Implicit VR Little Endian
 
Explicit VR  Little Endian  1.2.840.10008.1.2 
1.2.840.10008.1.2.1
 SCU 
None 
Study Root 
Query/Retrieve (C -
MOVE)...