Home > Philips > Ultrasound System > Philips Epiq 5 Manual

Philips Epiq 5 Manual

Here you can view all the pages of manual Philips Epiq 5 Manual. The Philips manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.

Page 41

 
Module Name 
 Attribute Name 
Tag VR BQ  PQ R D IOD  M SD  
 
Patient Medical 
    
        
 Medical Alerts (0010,2000) LO   x  x   
 Additional Patient’s History (0010,21B0) LT   x  x   
 Pregnancy Status (0010,21C0) US   x     
 Last Menstrual Date (0010,21D0) DA   x  x   
 Special Needs (0038,0050) LO   x  x   
 
Note 1: Patient Comments originating from  modality worklist patients are not exported on exams consisting entirely of 
loop images . 
Note 2: If present, Requested Procedure Description...

Page 42

 
The “Cancel Exam” button causes a “DISCONTINUED” message.  An exam for which an MPPS Instance is sent with a 
state of “DISCONTINUED” can also no longer be updated. 
The system supports creation of “unscheduled cases” by allowing MPPS Instances to be communicated for locally 
registered Patients.  
The system performs a single Performed Procedure Step at a time per Scheduled Procedure Step.  
EPIQ/Affi niti will initiate an Association to issue an:   
—  N-CREATE request according to the Create...

Page 43

 
Note: The Cancel and End Exam commands are mutually exclusive.  They are both represented here for illustration 
purposes only.  Actual workflow uses one or the other for a given exam. 
 
4.2.2.3.2.2  Proposed Pre sentation Contexts  
EPIQ/Affiniti  will propose Presentation Contexts as shown in the following table:  
Table  4.31  
PROPOSED PRESENTATION CONTEXTS FOR REAL -WORLD ACTIVITY ACQUIRE IMAGES  
Presentation Context Table  
Abstract Syntax  Transfer Syntax   
 
Role    
Ext.  
Neg.  
Name   UID...

Page 44

 Table 4.34  
MPPS N -CREATE / N- SET REQUEST IDENTIFIER  
Attribute Name  Tag  VR  N-CREATE   N-SET  
Specific Character Set  (0008,0005) CS  Not Sent    
Modality  (0008,0060) CS  US    
Referenced Patient Sequence  (0008,1120) SQ     
> Referenced SOP Class UID (0008,1150)  UI      
>Referenced SOP Instance UID  (0008,1155)  UI      
Patient’s Name  (0010,0010) PN  From 
Modality Worklist or 
user input.    
Patient ID
  (0010,0020) LO  From Modality Worklist or 
user input.
   
Patient’s Birth Date...

Page 45

 Attribute Name  Tag  VR  N-CREATE   N-SET  
Performed Protocol Code 
Sequence   (0040,0260)
 SQ  Zero length 
OR 
corresponding to built -in 
Staged Protocols OR  
mapped from MWL 
Scheduled Protocol Code 
Sequence  (0040,0008)  Same
 
Scheduled Step Attributes 
Sequence   (0040,0270)
 SQ     
> Accession Number (0008,0050) SH  From MWL or user entered
. 
MWL value may be edited.   
> Referenced Study Sequence 
(0008,1110) SQ  One item per item in the 
MWL Reference Study 
Sequence. Absent if...

Page 46

 Attribute Name  Tag  VR  N-CREATE   N-SET  
> Protocol Name  (0018,1030) LO    See Table  9.9 
> Series Instance UID   (0020,000E )  UI     System generated 
> Referenced Non- Image 
Composite SOP Instance 
Sequence   (0040,0220)
 SQ    One item per referenced 
non-
image instance  
>> Referenced SOP Class UID  (0008,1150)  UI     SOP Class UID of acquired 
instance
 
>> Referenced SOP Instance UID  (0008,1155)  UI     SOP Instance UID of 
acquired instance
 
 
4.2.2.4 Association Acceptance Policy...

Page 47

 
 
4.2.3.2.3 Asynchronous Nature  
EPIQ/Affiniti  does not support asynchronous communication (multiple outstanding transactions over a single 
Association).  
Table 4.38  
ASYNCHRONOUS NATURE AS A SCU FOR AE HARDCOPY  
Maximum number of outstanding asynchronous transactions   1 
 
4.2.3.2.4  Implementation Identifying Information  
The implementation information for this Application Entity is:  
Table  4.39  
DICOM IMPLEMENTATION CLASS AND VERSION FOR AE HARDCOPY  
Implementation Class UID...

Page 48

 
HardcopyAEPrinter
1. Open Association
2. N-GET (Printer)
3. N-CREATE (Film Session)
4. N-CREATE (Film Box)
5. N-SET (Image Box)
6. N-ACTION (Film Box)
7. Print film sheets
8. N-DELETE (Film Box)
9. N-DELETE (Film Session)
10. Close Association
 
Figure 7  
SEQUENCING OF ACTIVI TY – PRINT IMAGES  
Figure 7 illustrates a typical sequence of DIMSE messages sent over an association between Hardcopy AE and a 
Printer. Two DICOM Printers may be simultaneously configured, one for BW and one for Color prints....

Page 49

 Table 4.40  
PROPOSED PRESENTATION CONTEXTS FOR ACTIVITY FILM IMAGES  
Presentation Context Table  
Abstract Syntax  Transfer Syntax   
 
Role    
Ext.  
Neg.  
Name   UID  Name List UID List 
Basic Grayscale 
Print Management 
Meta  1.2.840.10008.5.1.
1.9
  Implicit VR Little Endian
 
Explicit VR Little Endian  1.2.840.10008.1.2
 
1.2.840.10008.1.2.1  SCU
 None  
Basic Color Print 
Management Meta  1.2.840.10008.5.1.
1.18
  Implicit VR Little Endian
 
Explicit VR Little Endian  1.2.840.10008.1.2...

Page 50

 Table 4.43  
PRINTER SOP CLASS N -GET RESPONSE STATUS HANDLING BEHAVIOR  
Service Status  
Further Meaning Error Code  Behavior 
Success  Success  0000 The request to get printer status information was success.   
*  *  Any other status code.  Same as Timeout above.  
 
4.2.3.3.1.3.2  Printer SOP Class Notifications (N -EVENT -REPORT)   
Hardcopy AE is capable of receiving an N -EVENT -REPORT request at any time during an association.  
Table  4.44  summarizes the behavior of Hardcopy AE when receiving...
Start reading Philips Epiq 5 Manual

Related Manuals for Philips Epiq 5 Manual

All Philips manuals