Philips Epiq 5 Manual
Here you can view all the pages of manual Philips Epiq 5 Manual. The Philips manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 31
Patient ID (0010,0020) Study Instance UID (0020,000D ) Series Instance UID (0020,000E ) Study ID (0020,0010) 4.2.1.3.2.3.2 Study Root Query Retrieve – MOVE SOP Class After examining the results of the user -initiated Query, the user may select one or more Studies from the query results list and request that the contents of the selected Studies be retrieved. A C -MOVE message is sent to the Query/Retrieve SCP requesting that the selected items be stored back to this system....
Page 32
4.2.1.4.1.3 SOP Specific Conformance for Storage Commitment Push Model SOP Class 4.2.1.4.1.3.1 Storage Commitment Notifications (N -EVENT -REPORT) The Storage AE can receive an N -EVENT -REPORT notification received from the SCP via an association requested by the SCP with reverse- role negotiation. Upon receipt of an N -EVENT -REPORT the timer associated with the Transaction UID will be canceled. Table 4.18 summarizes the behavior of Storage AE when receiving Event Types within the N...
Page 33
Commitment Request Complete – Failures Exist Retrieve AE Title (0008,0054) None Storage Media File- Set ID (0088,0130) None Storage Media File- Set UID (0088,0140) None Referenced SOP Sequence (0008,1199) None >Referenced SOP Class UID (0008,1150) None >Referenced SOP Instance UID (0008,1155) None >Retrieve AE Title (0008,0054) None >Storage Media File- Set ID (0088,0130) None >Storage Media File- Set UID (0088,0140) None Failed SOP Sequence (0008,1198) None...
Page 34
4.2.1.4.2.2 Accepted Presentation Contexts Table 4.21 summarizes Presentation Contexts that t he Storage AE accepts for this activity . Table 4.21 ACCEPTABLE PRESENTAT ION CONTEXTS FOR ACTIVITY STORAGE FRO M A REMOTE SCU Presentation Context Table Abstract Syntax Transfer Syntax Role Ext. Neg. Name UID Name List UID List US Image Storage 1.2.840.10008.5.1.4.1.1. 6.1 Implicit VR Little Endian Explicit VR Little Endian JPEG Lossy Baseline JPEG Lossless Baseline...
Page 35
Digital Mammography X- Ray Image Storage – For Presentation 1.2.840.10008.5.1.4.1.1. 1.2 Implicit VR Little Endian Explicit VR Little Endian JPEG Lossy Baseline JPEG Extended 12- bit Lossy JPEG Lossless Baseline 1.2.840.10008.1.2 1.2.840.10008.1.2.1 1.2.840.10008.1.2.4.50 1.2.840.10008.1.2.4.51 1.2.840.10008.1.2.4.70 SCP None Digital Mammography X - Ray Image Storage – For Processing 1.2.840.10008.5.1.4.1.1. 1.2.1 Implicit VR Little Endian Explicit VR Little Endian...
Page 36
4.2.2 Workflow Application Entity Specification 4.2.2.1 SOP Classes EPIQ/Affiniti provides Standard Conformance to the following SOP Classes: Table 4.22 SOP CLASSES FOR AE WORKFLOW SOP Class Name SOP Class UID SCU SCP MWL Information Model – FIND 1.2.840.10008.5.1.4.31 Yes No Modality Performed Procedure Step 1.2.840.10008.3.1.2.3.3 Yes No 4.2.2.2 Association Establishment Policy 4.2.2.2.1 General The DICOM standard application context name for DICOM 3.0 is always...
Page 37
Table 4.26 DICOM IMPLEMENTATION CLASS AND VERSION FOR AE WORKFLOW Implementation Class UID 1.3.46.670589.14.10000.105 Implementation Version Name EPIQ-AFFIN_1.05 4.2.2.3 Association Initiation Policy 4.2.2.3.1 Activity – Worklist Update 4.2.2.3.1.1 Description and Sequencing of Activities The Worklist Update activity is capable of updating the modality worklist using either a Broad Query or a Patient Query, as described later in this section. A possible sequence of interactions...
Page 38
• Modality = “US” or a user - specified value The user may configure one of the following schemes for updating the worklist using Broad Query : • The worklist may be updated when the user presses “Update W orklist” on the patient data entry screen • The worklist may be updated automatically each time an exam is ended • The worklist may be updated periodically at a configurable time interval between 15 and 120 minutes in 15 minute increments The user at may cancel a worklist...
Page 39
Service Status Further Meaning Error Code Behavior Refused Out of Resources A700 The Association is aborted using A -ABORT. The worklist is not replaced. Failed Identifier does not match SOP Class A900 Same as “Refused” above. Failed Unable to Process C000 – CFFF Same as “Refused” above. Cancel Matching terminated due to Cancel request FE00 The retrieved items are ignored. Pending Matches are continuing FF00 Continue. Pending Matches are continuing – Warning that...
Page 40
Table 4.30 WORKLIST MATCHING KEYS Module Name Attribute Name Tag VR BQ PQ R D IOD M SD SOP Common Specific Character Set (0008,0005) CS x Scheduled Procedure Step Scheduled Procedure Step Sequence (0040,0100) SQ x x > Scheduled Station AE Title (0040,0001) AE (S) x > Scheduled Procedure Step Start Date (0040,0002) DA S x x x > Scheduled Procedure Step Start Time (0040,0003) TM x x > Modality (0008,0060) CS S, (S) x x x x > Scheduled...