Philips Epiq 5 Manual
Here you can view all the pages of manual Philips Epiq 5 Manual. The Philips manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 91
Structured Reporting uses codes supplied by DCMR (DICOM Code Mapping Resource, PS 3-16), LOINC, SRT and 99PMSBLUS (Philips Private Codes for Ultrasound). 9.4 GRAYSCALE IM AGE CONS ISTENCY The high- resolution display monitor is calibrated according to the Grayscale Standard Display Function (GSDF). 9.5 STANDARD EXTENDED/SP ECIALIZED/PRIVATE SOP CLASSES 9.5.1 Ultrasound Image and Multi -frame Image SOP Classes The US or US Multiframe Image Storage SOP Classes are extended to create a...
Page 92
Sequence >> Code Value (0008,0100) 1 SH Code Value 121324 >>Code Meaning (0008,0104) 1 LO Code Meaning Source Image 9.5.2 3D Presentation State Private SOP Class EPIQ 7 only may create instances of the following Private SOP Class: Table 9.32 PRIVATE SOP CLASS FOR 3D PRESENTATION STATE SOP Class Name SOP Class UID SCU SCP Private 3D Presentation State 1.3.46.670589.2.5.1.1 Yes No 9.6 PRIVATE TRANSFER SYN T AXES There are no Private Transfer Syntaxes. E7 and...
Page 93
APPENDIX A – Structured Reports A.1 STRUCTURED REPORTS A.1.1 Introduction EPIQ/Affiniti implement DICOM Structured Report Templates TID 5000 ( OB-GYN Ultrasound Procedure Report ), TID 5100 ( Vascular Ultrasound Report ), TID 5200 (Echocardiography Procedure Report ) , TID 5220 ( Pediatric, Fetal and Congenital Cardiac Ultrasound Reports) , p rivate TID 995300 (Ped Echo) based on s upplement 78 version 2, May 8 2004 , and private TID 995400 (Small Parts ). See section A.2 for the DICOM...
Page 94
spatial coordinate shall be the ventricular length measurement, and the 22nd spatial coordinat e shall be the v entricular area measurement. c. For Doppler velocity measurements made with a two- point or trace tool, there shall be two spatial coordinates per measurement. The first shall be the single point associated with the velocity measurement, and t he second shall consist of all the points of the measurement tool . A.1.2 Clinical Scope The supported measurements are located in Calcs...
Page 95
To access the Philips Healthcare Ultrasound Data Portal, • Using a browser, to n avigate to the link below . • Select the ult rasound product of interest. • Provide an email address and press the submit button. • Download “ Philips_EPIQAffiniti_v1.5_DICOM_SR_Specification.xlsx ” Philip Ultrasound Data Portal Website Link: http://129.33.58.186/portal/index.asp?pid=us A.4 PRIVATE TEMPLATE EXT ENSIONS The PDE and Study Info data that is not already part of the DICOM...
Page 96
“Gestational Age”) shall be present 4 >> INFERRED FROM CODE DCID (228) Equation or Table 1 U IF row 4.1 is absent DCID (12013) Gestational Age Equations and Tables 4.1 >> INFERRED FROM TEXT DCID (228) Equation or Table 1 U IF row 4 is absent 5 >> R-INFERRED FROM NUM 1-n U A.4.3 TID5101: Vascular Patient Characteristics NL Rel with Parent VT Concept Name VM Req Type Condition Value Set Constraint … … … … … … … … … 7 > CONTAINS TEXT or DATE or NUM DCID (99002) General Study Info...
Page 97
A.4.4 TID5202: Echocardiography Patient Characteristics NL Rel with Parent VT Concept Name VM Req Type Condition Value Set Constraint … … … … … … … … … 9 > CONTAINS TEXT or DATE or NUM DCID (99002) General Study Info 1 U 10 > CONTAINS TEXT or DATE or NUM DCID (99007) Adult Echo Study Info 1 U E7 and E5/A70 and A50 1.5.x.x DICOM Conformance Statement 000490000000111 Rev A Page 97
Page 98
A.5 USER -DEFINED MEASUREMENTS AND CALCULATIONS A.5.1 Description EPIQ/Affiniti provides 2 different DICOM export formats for creating user defined measurements and calculations. The DICOM Encoding format provides a way to create and export user -defined measurements and calculations or user - defined OB authors to PACS using the existing DICOM SR templates. The P hilips Generic export format allows users to create measurements and calculations without specifying complete DICOM encoding...
Page 99
A.5.1.3 Private Template and Template Extensions A.5.1.3.1 TID5000: OB -GYN Ultrasound Procedure Report NL Rel with Parent VT Concept Name VM Req Type Condition Value Set Constraint … … … … … … … … … 25 > CONTAINS INCLUDE TID (9902) Fetal Heart Section 1 U 26 > CONTAINS INCLUDE TID (9900) User-defined concepts 1 U A.5.1.3.2 TID5100: Vascular Ultrasound Report NL Rel with Parent VT Concept Name VM Req Type Condition Value Set Constraint … … … … … … … … … 31 > CONTAINS INCLUDE...
Page 100
NL Rel with Parent VT Concept Name VM Req Type Condition Value Set Constraint 1 CONTAINER $Type 1 M 2 > HAS OBS CONTEXT INCLUDE DTID (1008) Subject Context, Fetus 1 MC IF this template is invoked more than once to describe more than one fetus 3 > HAS CONCEPT MOD TEXT DT (T9900-04, 99PMSBLUS, “Label”) 1 M 4 > HAS CONCEPT MOD CODE DT (G-C171, SRT, “Laterality”) 1 U DCID (244) Laterality 5 > CONTAINS INCLUDE TID (300) 1 1-n IFF $Type = “Measurement” $Measurement = DCID (99008)...