Structured Reporting uses codes supplied by DCMR (DICOM Code Mapping Resource, PS 3-16), LOINC, SRT and 99PMSBLUS (Philips Private Codes for Ultrasound). 9.4 GRAYSCALE IM AGE CONS ISTENCY The high- resolution display monitor is calibrated according to the Grayscale Standard Display Function (GSDF). 9.5 STANDARD EXTENDED/SP ECIALIZED/PRIVATE SOP CLASSES 9.5.1 Ultrasound Image and Multi -frame Image SOP Classes The US or US Multiframe Image Storage SOP Classes are extended to create a Standard Extended SOP Class by addition of standard and private attributes to the created SOP Instanc es as documented in section 8.1 as well as the standard attributes listed in Table 9.30 . Table 9.30 ULTRASOUND IMAGE EXTENDED ATTRIBUTES Tag Attribute Name Added to (0028,0030) Pixel Spacing Images with a single 2D region or dual 2D with same depth (0008,0068) Presentation Intent Type Indicates that Pixel Data (7FE0,0010) contains 3D volume data (0008,2112) Source Image Sequence Extended to use codes from CID 7205. See Table 9.31 . Pixel Spacing (0028,0030) is added to the exported DICOM file when the user has configured this attribute to be included and the image contains only one 2D calibration region and no Doppler or M- Mode calibration regions. This attribute is system generated, if used: • Contain the Pixel Spacing attribute: 2D still, 2D loop, 2D color still, 2D color loop, Mm ode Preview Still, PW Preview Still, CW Preview still, and Dual with same calibration on both images. • Do NOT contain the Pixel Spacing attribute: Mmode live trace, Mmode frozen trace, PW live trace, PW Frozen trace, CW live trace, CW frozen trace, Reports and dual images with different calibration on each image. Ultrasound Multi frame Image instances may be used to store 3D and 4D ultrasound volume information. The presence of the Presentation Intent Type (0008,0068) value “FOR PRESENTATION” indicates that Pixel Data (7FE0,0010) contains spatially related frames of a 3D volume rather than the normal temporally related frames. 3D Temporal (a.k.a. 4D) volume information is stored using private attributes within a normal Ultrasound Multi frame Image object. The private attributes that are sent via network or media are depend ent on settings for the specific destination, as found in “Print/Network>Device Selection and either t he “Media” tab or “Archive” tab and the “Advanced” button for the highlighted archive device. Table 9.31 SOURCE IMAGE SEQUENCE ITEM EXTENDED ATTRIBUTES Attribute Name Tag Type VR Description Value Source Image Sequence ( 0008,2112) 3 SQ The set of Image SOP Class/Instance pairs of the Images that were used to derive this image. >Purpose of Reference Code ( 0040,A170) 3 SQ Describes the purpose for which the reference is made. E7 and E5/A70 and A50 1.5.x.x DICOM Conformance Statement 000490000000111 Rev A Page 91
Sequence >> Code Value (0008,0100) 1 SH Code Value 121324 >>Code Meaning (0008,0104) 1 LO Code Meaning Source Image 9.5.2 3D Presentation State Private SOP Class EPIQ 7 only may create instances of the following Private SOP Class: Table 9.32 PRIVATE SOP CLASS FOR 3D PRESENTATION STATE SOP Class Name SOP Class UID SCU SCP Private 3D Presentation State 1.3.46.670589.2.5.1.1 Yes No 9.6 PRIVATE TRANSFER SYN T AXES There are no Private Transfer Syntaxes. E7 and E5/A70 and A50 1.5.x.x DICOM Conformance Statement 000490000000111 Rev A Page 92
APPENDIX A – Structured Reports A.1 STRUCTURED REPORTS A.1.1 Introduction EPIQ/Affiniti implement DICOM Structured Report Templates TID 5000 ( OB-GYN Ultrasound Procedure Report ), TID 5100 ( Vascular Ultrasound Report ), TID 5200 (Echocardiography Procedure Report ) , TID 5220 ( Pediatric, Fetal and Congenital Cardiac Ultrasound Reports) , p rivate TID 995300 (Ped Echo) based on s upplement 78 version 2, May 8 2004 , and private TID 995400 (Small Parts ). See section A.2 for the DICOM Structured Report export capabilities for each EPIQ/Affiniti model. DICOM PS 3.17 of the DICOM Standard includes tree diagrams showing graphic examples of the structure of each template. Briefly, a Structured Report (SR) document will contain only the measurements, calculations and observations made during the exam. Its exact structure is therefore determined by two main components, the measurements that are available within the context of the template and its referenced Templates and Context groups, and those measurements and calculations that are implemented on the system creating the report. Supplementing these constraints are private and user defined measurements and calculations, which may be added if the Root Container Template is extensible. In EPIQ/Affiniti , the calculation packages contain a combi nation of template and user defined measurements and calcul ations. A.1.1.1 Measurements Linked to Images Spatial Coordinates (as defined by TID 320 row 3) are included for measurement sub- results whose measurement tools have been captured in an image (as defined by TID 300 row 13). There are three special cases in which the Value Multiplicity for DTID 320 shall be greater than one: a. If more than one image is captured of the same measurement tool, the sub- result will reference each image. b. For a complex measurement tool (the Simpson’s Method of Disk Volume measurement) which with one user action creates 22 unique spatial coordinates, the first 20 spatial coordinate shall be the Chord values, the 21st ROOT Document Title CONTAINER CONTAINS Patient Characteristics CONTAINER 0-1 CONTAINS Observations NUM, CODE, TEXT, DATE 1-N CONTAINS Procedure Summary CONTAINER 0-1 CONTAINS Observations NUM, CODE, TEXT, DATE 1-N CONTAINS Image LIbrary CONTAINER 0-1 CONTAINS Image Entry IMAGE 1-N CONTAINS Observations NUM, CODE, TEXT 0-N CONTAINS Heading CONTAINER 0-N CONTAINS Measurement Group CONTAINER, CODE 0-N CONTAINS, HAS PROP Observations NUM, CODE, TEXT 1-N CONTAINS Fetus Summary CONTAINER 0-n E7 and E5/A70 and A50 1.5.x.x DICOM Conformance Statement 000490000000111 Rev A Page 93
spatial coordinate shall be the ventricular length measurement, and the 22nd spatial coordinat e shall be the v entricular area measurement. c. For Doppler velocity measurements made with a two- point or trace tool, there shall be two spatial coordinates per measurement. The first shall be the single point associated with the velocity measurement, and t he second shall consist of all the points of the measurement tool . A.1.2 Clinical Scope The supported measurements are located in Calcs packages accessed with the “Measure” hard key on EPIQ/Affiniti . Measurements for a given SR section may come from several calc s sidebars/touch screens. A.2 APPLICATIONS THAT EXPORT STRUCTURED REPORTS F ROM EPIQ AND AFFINITI The following table describes the DICOM Structured R eport export capabilities by a pplication for the EPIQ 7 and 5 ultrasound systems . A following table shows those for Affiniti 70 and 50. • “G” is optimized for General Imaging, Shared Service, and Radiology use • “C” is optimized for Cardiology use • “W” is optimized for Women’s Healthcare use OB/Gyn TID 5000 Vascular TID 5100 Abdominal TID 5100 TCD TID 5100 Adult Echo TID 5200 Pediatric Echo TID 5220 Fetal Echo TID 5220 Pediatric Echo TID 995300 (Private ) Small Parts TID 995400 (Private ) EPIQ 7G Affiniti 70G EPIQ 7C Affiniti 70C EPIQ 7W Affiniti 70W EPIQ 5G Affiniti 50G EPIQ 5C Affiniti 50C EPIQ 5W Affiniti 50W A.3 DICOM STRUCTURED REP ORT EXPORT SPECIFICATIONS A.3.1 Philips Healthcare Ultrasound Data Portal Website Rather than list the thousands of DICOM export encodings in this document, t he Philips Healthcare Ultrasound Data Portal website (see link below) provides detailed DICOM export specifications for all Philips ultrasound products in a convenient downloadable spreadsheet form at. These spreadsheets allow users to quickly filter for specific areas of interests such as application, template, label, or DICOM attribute. In addition to DICOM Structured Report specifications, the link also provides many DICOM Structured Report exampl e files and images for import and display testing. E7 and E5/A70 and A50 1.5.x.x DICOM Conformance Statement 000490000000111 Rev A Page 94
To access the Philips Healthcare Ultrasound Data Portal, • Using a browser, to n avigate to the link below . • Select the ult rasound product of interest. • Provide an email address and press the submit button. • Download “ Philips_EPIQAffiniti_v1.5_DICOM_SR_Specification.xlsx ” Philip Ultrasound Data Portal Website Link: http://129.33.58.186/portal/index.asp?pid=us A.4 PRIVATE TEMPLATE EXT ENSIONS The PDE and Study Info data that is not already part of the DICOM templates is included using the following template extensions, per the appropriate application. A.4.1 TID5001: OB -GYN Patient Characteristics NL Rel with Parent VT Concept Name VM Req Type Condition Value Set Constraint … … … … … … … … … 9 > CONTAINS TEXT or DATE or NUM DCID (99002) General Study Info 1 U 10 > CONTAINS TEXT or DATE or NUM DCID (99003) OB Study Info 1 U 11 > CONTAINS TEXT or DATE or NUM DCID (99004) Gyn Study Info 1 U Line 4.1 is used only with a user -defined table or equation. A.4.2 TID5008: Fetal Biometry Group NL Rel with Parent VT Concept Name VM Req Type Condition Value Set Constraint … … … … … … … … … 3 > CONTAINS NUM EV (18185-9 , LN, 1 MC At least one of row 2 and 3 Units = EV (d, UCUM, days) E7 and E5/A70 and A50 1.5.x.x DICOM Conformance Statement 000490000000111 Rev A Page 95
“Gestational Age”) shall be present 4 >> INFERRED FROM CODE DCID (228) Equation or Table 1 U IF row 4.1 is absent DCID (12013) Gestational Age Equations and Tables 4.1 >> INFERRED FROM TEXT DCID (228) Equation or Table 1 U IF row 4 is absent 5 >> R-INFERRED FROM NUM 1-n U A.4.3 TID5101: Vascular Patient Characteristics NL Rel with Parent VT Concept Name VM Req Type Condition Value Set Constraint … … … … … … … … … 7 > CONTAINS TEXT or DATE or NUM DCID (99002) General Study Info 1 U 8 > CONTAINS TEXT or DATE or NUM DCID (99005) Vascular Study Info 1 U 9 > CONTAINS TEXT or DATE or NUM DCID (99006) Abdominal Study Info 1 U E7 and E5/A70 and A50 1.5.x.x DICOM Conformance Statement 000490000000111 Rev A Page 96
A.4.4 TID5202: Echocardiography Patient Characteristics NL Rel with Parent VT Concept Name VM Req Type Condition Value Set Constraint … … … … … … … … … 9 > CONTAINS TEXT or DATE or NUM DCID (99002) General Study Info 1 U 10 > CONTAINS TEXT or DATE or NUM DCID (99007) Adult Echo Study Info 1 U E7 and E5/A70 and A50 1.5.x.x DICOM Conformance Statement 000490000000111 Rev A Page 97
A.5 USER -DEFINED MEASUREMENTS AND CALCULATIONS A.5.1 Description EPIQ/Affiniti provides 2 different DICOM export formats for creating user defined measurements and calculations. The DICOM Encoding format provides a way to create and export user -defined measurements and calculations or user - defined OB authors to PACS using the existing DICOM SR templates. The P hilips Generic export format allows users to create measurements and calculations without specifying complete DICOM encoding attributes and will be exported in a generic DICOM encoding format. A.5.1.1 Philips Generic DICOM Export Format For quickly creating exportable measurement s and calculations without specific DICOM encodings, a generic structure was created that does not assign specific codes to the individual measurements, rather uses the label given by the user. This will allow all user -defined measur ements to be treated in a uniform manner, without needing a per -site dictionary of user -defined codes. CONTAINS“User-defined concepts”CONTAINER HAS CONCEPT MOD“Label” = “MyMeas”TEXT CONTAINS“Measurement”CONTAINER HAS CONCEPT MOD“Label” = “MyCalc”TEXT CONTAINS“Calculation”CONTAINER HAS CONCEPT MOD“Label” = “MyMeas2”TEXT CONTAINS“Measurement”CONTAINER HAS CONCEPT MODSelector = LastCODE CONTAINS“Distance” = 12.3 mmNUM CONTAINS“Value” = 34.5 mm/sNUM CONTAINS“Distance” = 23.4 mmNUM CONTAINS“Distance” = 17.9 mmNUM HAS CONCEPT MODDerivation = MeanCODE CONTAINS“DIstance” = 45.6 mmNUM INFERRED FROMEquation = “MyEqn”TEXT CONTAINS“Gestational Age” =10 days NUM A.5.1.2 DICOM Template Compatible Export Format EPIQ/Affiniti can also export user defined measurements, calculations and OB tables which are compatible with existing DICOM standard templates . These DICOM objects are created with user configurable attributes for the DICOM code scheme designator, code value and code meaning. For more information on the DICOM attributes support ed, see t he Analysis section in the EPIQ and Affiniti Setups menu. E7 and E5/A70 and A50 1.5.x.x DICOM Conformance Statement 000490000000111 Rev A Page 98
A.5.1.3 Private Template and Template Extensions A.5.1.3.1 TID5000: OB -GYN Ultrasound Procedure Report NL Rel with Parent VT Concept Name VM Req Type Condition Value Set Constraint … … … … … … … … … 25 > CONTAINS INCLUDE TID (9902) Fetal Heart Section 1 U 26 > CONTAINS INCLUDE TID (9900) User-defined concepts 1 U A.5.1.3.2 TID5100: Vascular Ultrasound Report NL Rel with Parent VT Concept Name VM Req Type Condition Value Set Constraint … … … … … … … … … 31 > CONTAINS INCLUDE TID (9900) User-defined concepts 1 U A.5.1.3.3 TID5200: Echocardiography Procedure Report NL Rel with Parent VT Concept Name VM Req Type Condition Value Set Constraint … … … … … … … … … 22 > CONTAINS INCLUDE TID (9900) User-defined concepts 1 U A.5.1.3.4 TID995300: Pediatric Echocardiography Procedure Report NL Rel with Parent VT Concept Name VM Req Type Condition Value Set Constraint … … … … … … … … … 22 > CONTAINS INCLUDE TID (9900) User-defined concepts 1 U A.5.1.3.5 TID9900: User -defined concepts NL Rel with Parent VT Concept Name VM Req Type Condition Value Set Constraint 1 CONTAINER DT (T9900-01, 99PMSBLUS, “User-defined concepts”) 1 M 2 > CONTAINS INCLUDE TID (9901) User -defined concept 1-n MC One of row 2 and 3 must be present $Type = DT (T9900- 02, 99PMSBLUS, “Measurement”) 3 > CONTAINS INCLUDE TID (9901) User -defined concept 1-n MC One of row 2 and 3 must be present $Type = DT (T9900- 03, 99PMSBLUS, “Calculation”) A.5.1.3.6 TID9901: User -defined concept E7 and E5/A70 and A50 1.5.x.x DICOM Conformance Statement 000490000000111 Rev A Page 99
NL Rel with Parent VT Concept Name VM Req Type Condition Value Set Constraint 1 CONTAINER $Type 1 M 2 > HAS OBS CONTEXT INCLUDE DTID (1008) Subject Context, Fetus 1 MC IF this template is invoked more than once to describe more than one fetus 3 > HAS CONCEPT MOD TEXT DT (T9900-04, 99PMSBLUS, “Label”) 1 M 4 > HAS CONCEPT MOD CODE DT (G-C171, SRT, “Laterality”) 1 U DCID (244) Laterality 5 > CONTAINS INCLUDE TID (300) 1 1-n IFF $Type = “Measurement” $Measurement = DCID (99008) Results $Derivation = DCID (3627) Measurement Type 6 > CONTAINS INCLUDE TID (300) 1 1 IFF $Type = “Calculation” $Measurement = DT (T9900-05, 99PMSBLUS, “Value”) 7 > CONTAINS NUM EV (18185-9 , LN, “Gestational Age”) 1 U $Units = EV (d, 1.4, UCUM, days) 8 > INFERRED FROM TEXT DCID (228) Equation or Table 1 U A.5.1.3.7 TID9902: Fetal Heart Section NL Rel with Parent VT Concept Name VM Req Type Condition Value Set Constraint 1 CONTAINER DT (99999, 99PMSBLUS, “Fetal Heart”) 1 M … … 2 > HAS OBS CONTEXT INCLUDE DTID (1008) Subject Context. Fetus 1 MC IF this template is invoked more than once to describe more than one fetus 3 > CONTAINS INCLUDE DTID (5008) Fetal Biometry Group 1-n M $Biometery Type=MemberOf (DCID (99001) Fetal Heart) A.5.1.3.8 TID 5009: Fetal Biophysical Profile Section NL Rel with Parent VT Concept Name VM Req Type Condition Value Set Constraint 1 CONTAINER DT 125006,DCM,”Biophysical Profile”) 1 M 2 > HAS OBS CONTEXT INCLUDE DTID (1008) Subject Context. Fetus 1 MC IF this template is invoked more than once to describe more than one E7 and E5/A70 and A50 1.5.x.x DICOM Conformance Statement 000490000000111 Rev A Page 100